Skip to content

Brand Name: ID NOW RSV 24T recallZ-1576-2025 · Abbott Diagnostics Scarborough, Inc. · initiated Mar 4, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott Diagnostics Scarborough, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

the impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.

FDA root cause category: Process change control

What the company is doing

On March 4, 2025 PRODUCT REMOVAL NOTICE letters were sent to distributors. Necessary Actions to be Taken by Distributor: 1. If you have impacted inventory held in stock, then: "Immediately quarantine impacted product per your local procedure. "Complete and return the Distributor Acknowledgment Form (Form must be completed, signed and returned to receive replacement product). "Contact Customer Service to arrange return of your product by Abbott. 2. If you have forwarded the product listed above to others within your distribution network, then: "Inform them of this Product Removal Notice and provide to them a copy of this notice and request they take the necessary action. 3. If you have forwarded the impacted product to your customers, then: "Provide customer list to Abbott based on the product information in Appendix II as outlined in the distribution agreement within 48 hours of receipt of this Product Removal Notice. "Abbott will notify customers of this Product Removal Notice. 4. If you do not have any inventory of the impacted lot(s), then: "All product lots not identified in Appendix I can continue to be used. If you have questions regarding this information, U.S. Customers please contact Sedgwick at 1 888 671 8858 (available 8am 6pm EST, Monday - Friday). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178). If you have experienced any patient or user injury associated with this Removal Notice, please immediately report the event to your local area Customer Service. ****Update 05/08/2025**** On April 29, 2025 Updated PRODUCT REMOVAL NOTICE letters were sent to distributors notifying of additional lots that were included in the recall.

Recall details

Recall number
Z-1576-2025
Status
Open, Classified
Initiated by the company
Mar 4, 2025
Posted by the FDA
Apr 15, 2025
Recalling firm
Abbott Diagnostics Scarborough, Inc.
Quantity
442,296 eaches
Product code
OCC
FDA submissions
K161375
Distribution
US Nationwide distribution.

Product description

Brand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system Component: No

Code information (models, lots, serial numbers)

Model 435-000; GTIN/DI 10811877010521; Lot 000M906790 Model 435-000; GTIN/DI 10811877010521; Lot 000M907271 Model 435-000; GTIN/DI 10811877010521; Lot 000M923531 Model 435-000; GTIN/DI 10811877010521; Lot 000M923879 Model 435-000; GTIN/DI 10811877010521; Lot 000M924414 Model 435-000; GTIN/DI 10811877010521; Lot 000M924430 Model 435-000; GTIN/DI 10811877010521; Lot 000M926642 Model 435-000; GTIN/DI 10811877010521; Lot 000M926679 Model 435-000; GTIN/DI 10811877010521; Lot 000M927161

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 96500).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page