Radiometer ABL90 Series - Radiometer ABL90 FLEX Model… recallZ-1659-2025 · Radiometer Medical ApS · initiated Mar 21, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Radiometer Medical ApS or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
ABL90 FLEX and ABL90 FLEX PLUS analyzers running with Radiometer Windows 10 versions RWIN10 1.0 and RWIN10 1.1 may unexpectedly freeze (the analyzer ceases to operate) during a sample measurement. The analyzer software includes a feature (watchdog) that detects a freeze and automatically restarts the software to continue operating the analyzer. However, even if a restart occurs, any sample being processed at that time will be lost, potential to lead to delay in test results.
FDA root cause category: Component design/selection
What the company is doing
On 03/21/2025, the firm sent via email and FedEx an "Urgent: Medical Device Recall" Letter informing customers that, Radiometer has become aware of a potential risk with ABL90 FLEX and ABL90 FLEX PLUS analyzers when operating with Radiometer Windows 10 versions (RWIN10 1.0 and 1.1). Specifically, the analyzer software may unexpectedly "freeze" and restart during a sample measurement, causing the analyzer to cease operation during the startup period. Customers are instructed to: 1. Do not use the affected systems without having an alternative solution readily available. Alternative solutions could include using an additional analyzer or sending the sample to a central laboratory. This also involves ensuring that there is enough patient sample for a second measurement, in case the sample is lost due to the analyzer restarting. Additionally, we ask that you please post a copy of this letter on or near the device(s) until such time that the software has been upgraded. Radiometer will contact customers to schedule the upgrade to analyzers to the available software RWIN10 1.2 or greater. For questions, contact Radiometer at 1-800-736-0600 option 1, Monday - Friday 9 AM to 7 PM (EST).
Recall details
- Recall number
- Z-1659-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 21, 2025
- Posted by the FDA
- Apr 23, 2025
- Recalling firm
- Radiometer Medical ApS
- Quantity
- 6275 analyzers
- Product code
- CHL
- FDA submissions
- K092686, K160153
- Distribution
- U.S.: AL, AR, AZ, BM CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI, WV, and WY. O.U.S.: Canada, Afghanistan, Albania, Algeria, Argentina, Bahrain, Bangladesh, Belarus, Bolivia, Bosnia and Herzegovina, Brazil, Bulgaria, Chile, Colombia, Cuba, Cyprus, Dominican Republic, Ecuador, Egypt, Eritrea, Ethiopia, Greece, Guatemala, Honduras, Iraq, Iran, Israel, Jamaica, Jordan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Malta, Mauritius, Moldova, Nepal, Nicaragua, Nigeria, Oman, Pakistan, Paraguay, Peru, Philippines, Puerto Rico, Qatar, Romania, Saudi Arabia, Serbia, Slovenia, Sri Lanka, Suriname, Syria, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Uganda, Ukraine, United Arab Emirates, Uruguay, Uzbekistan, Venezuela, Vietnam, Yemen, Zimbabwe, Australia, New Zealand, New Caledonia, Tokelau, Austria, Belgium, Netherlands, China, Croatia, Czech Republic, Slovakia, Denmark, Iceland, Finland, Estonia, France, Germany, Hong Kong, Hungary, India, Brunei, Italy, Japan, Kazakhstan, Malaysia, Mexico, Norway, Poland, Russia, Singapore, South Africa, Cameroon, Ivory Coast, Malawi, South Korea, Spain, Portugal, Sweden, Switzerland, Turkey, United Kingdom, Ireland
Product description
Radiometer ABL90 Series - Radiometer ABL90 FLEX Model Number 393-090 Radiometer ABL90 Series - Radiometer ABL90 FLEX PLUS Model Number 393-092 The Radiometer ABL90 Series is a portable, automated analyzer for use in the laboratory environment, near patients or point-of-care settings.
Code information (models, lots, serial numbers)
Show the full code information (23k characters)
Model Number: 393-090 UDI-DI: 05700693930909 Model Number: 393-092 UDI-DI: 05700693930923 Serial Numbers: 090R0643N002 090R0749N001 090R0757N002 090R0757N004 090R0861N018 090R0874N009 090R0874N011 090R0874N012 090R0892N012 090R0897N023 090R0902N025 090R0943N010 090R0965N003 090R0965N007 090R0965N009 092R0150N001 092R0173N047 092R0173N049 092R0177N055 092R0180N011 092R0180N014 092R0180N025 092R0180N027 092R0185N022 092R0185N024 092R0185N025 092R0185N026 092R0187N016 092R0187N019 092R0187N021 092R0187N026 092R0187N027 092R0188N022 092R0188N024 092R0188N025 092R0188N026 092R0188N027 092R0188N028 092R0188N029 092R0188N033 092R0191N002 092R0195N034 092R0197N026 092R0197N028 092R0197N031 092R0198N005 092R0198N006 092R0198N008 092R0198N009 092R0198N014 092R0198N019 092R0198N027 092R0198N031 092R0200N041 092R0201N001 092R0201N003 092R0201N004 092R0201N006 092R0201N008 092R0204N035 092R0204N036 092R0214N011 092R0214N012 092R0214N013 092R0216N003 092R0216N004 092R0216N005 092R0216N006 092R0216N007 092R0216N016 092R0216N017 092R0216N020 092R0216N021 092R0218N014 092R0218N047 092R0218N048 092R0218N050 092R0221N011 092R0223N008 092R0223N011 092R0223N024 092R0223N026 092R0223N048 092R0223N050 092R0225N032 092R0225N039 092R0225N049 092R0225N050 092R0225N051 092R0225N052 092R0225N053 092R0225N054 092R0226N001 092R0226N002 092R0226N003 092R0226N015 092R0226N016 092R0226N018 092R0226N019 092R0227N047 092R0227N050 092R0227N051 092R0227N053 092R0227N054 092R0227N055 092R0227N056 092R0227N062 092R0227N077 092R0227N078 092R0228N007 092R0228N008 092R0228N030 092R0229N040 092R0229N052 092R0230N029 092R0230N030 092R0230N033 092R0230N037 092R0230N059 092R0230N068 092R0230N069 092R0230N070 092R0231N004 092R0231N007 092R0231N012 092R0231N013 092R0231N015 092R0231N016 092R0231N024 092R0231N029 092R0231N030 092R0231N031 092R0231N032 092R0231N033 092R0231N034 092R0231N035 092R0231N037 092R0231N040 092R0232N054 092R0233N024 092R0233N026 092R0234N006 092R0234N008 092R0234N009 092R0234N010 092R0234N017 092R0234N018 092R0234N019 092R0234N020 092R0234N027 092R0234N029 092R0235N002 092R0235N007 092R0235N008 092R0235N009 092R0235N010 092R0236N048 092R0237N022 092R0237N048 092R0237N056 092R0237N060 092R0238N060 092R0239N025 092R0239N026 092R0239N027 092R0239N028 092R0240N007 092R0240N011 092R0240N012 092R0240N013 092R0240N014 092R0240N019 092R0240N020 092R0240N021 092R0241N001 092R0241N003 092R0241N005 092R0241N007 092R0241N008 092R0241N014 092R0241N015 092R0241N016 092R0241N017 092R0241N018 092R0241N022 092R0241N024 092R0241N026 092R0241N027 092R0241N036 092R0241N037 092R0241N038 092R0241N039 092R0241N047 092R0241N048 092R0242N037 092R0244N008 092R0244N009 092R0244N010 092R0244N013 092R0244N014 092R0244N016 092R0244N020 092R0244N023 092R0244N025 092R0244N026 092R0244N031 092R0244N034 092R0244N035 092R0244N038 092R0244N039 092R0244N045 092R0244N049 092R0244N057 092R0244N060 092R0244N063 092R0244N064 092R0244N066 092R0245N045 092R0245N048 092R0245N099 092R0246N002 092R0246N004 092R0246N029 092R0246N063 092R0247N076 092R0247N077 092R0247N078 092R0247N080 092R0247N082 092R0247N083 092R0247N098 092R0248N028 092R0248N035 092R0248N039 092R0248N040 092R0248N045 092R0248N053 092R0248N086 092R0248N088 092R0248N093 092R0249N013 092R0249N024 092R0249N025 092R0250N057 092R0250N067 092R0250N072 092R0250N082 092R0250N083 092R0250N098 092R0251N031 092R0251N033 092R0251N035 092R0251N036 092R0251N037 092R0251N038 092R0251N039 092R0251N044 092R0252N003 092R0252N030 092R0253N001 092R0253N004 092R0253N006 092R0253N008 092R0253N010 092R0253N011 092R0255N003 092R0255N004 092R0255N060 092R0255N072 092R0256N002 092R0256N007 092R0256N008 092R0256N009 092R0256N010 092R0256N012 092R0256N015 092R0256N017 092R0256N021 092R0256N022 092R0256N026 092R0256N028 092R0256N029 092R0256N030 092R0256N031 092R0256N033 092R0256N061 092R0256N064 092R0256N067 092R0256N068 092R0256N069 092R0256N070 092R0256N074 092R0256N077 092R0256N079 092R0257N004 092R0257N010 092R0257N013 092R0257N015 092R0257N018 092R0257N019 092R0257N020 092R0257N022 092R0257N024 092R0257N025 092R0257N031 092R0257N034 092R0257N039 092R0257N040 092R0257N041 092R0257N042 092R0259N012 092R0259N021 092R0259N024 092R0259N025 092R0259N026 092R0259N029 092R0259N062 092R0259N064 092R0259N065 092R0260N039 092R0262N040 092R0262N041 092R0263N053 092R0263N055 092R0263N056 092R0264N001 092R0264N002 092R0266N044 092R0266N045 092R0266N048 092R0267N011 092R0267N024 092R0267N025 092R0271N057 092R0272N024 092R0272N028 092R0272N029 092R0272N032 092R0275N002 092R0275N006 092R0275N024 092R0275N030 092R0275N038 092R0276N042 092R0276N049 092R0276N051 092R0276N052 092R0276N053 092R0276N054 092R0276N057 092R0276N058 092R0276N059 092R0276N064 092R0276N066 092R0276N070 092R0276N071 092R0276N073 092R0276N074 092R0276N079 092R0276N088 092R0276N089 092R0277N007 092R0277N008 092R0277N009 092R0278N001 092R0278N004 092R0278N049 092R0279N052 092R0279N054 092R0279N056 092R0279N057 092R0279N058 092R0280N002 092R0280N003 092R0280N006 092R0280N007 092R0280N008 092R0280N038 092R0280N049 092R0281N027 092R0281N028 092R0281N029 092R0281N039 092R0281N044 092R0283N004 092R0283N006 092R0285N032 092R0285N038 092R0285N039 092R0285N044 092R0285N046 092R0285N053 092R0285N054 092R0285N055 092R0285N056 092R0285N057 092R0285N063 092R0285N064 092R0285N076 092R0285N077 092R0285N078 092R0285N079 092R0285N080 092R0286N037 092R0286N055 092R0287N043 092R0287N049 092R0287N050 092R0287N051 092R0287N052 092R0287N053 092R0287N054 092R0287N056 092R0287N058 092R0287N059 092R0288N036 092R0288N038 092R0288N043 092R0288N044 092R0289N005 092R0289N007 092R0289N015 092R0289N017 092R0289N021 092R0289N023 092R0290N013 092R0290N015 092R0290N016 092R0290N030 092R0290N031 092R0290N034 092R0290N043 092R0291N015 092R0291N029 092R0291N038 092R0291N048 092R0291N050 092R0293N006 092R0293N007 092R0294N001 092R0294N006 092R0294N012 092R0294N014 092R0294N032 092R0294N036 092R0294N038 092R0295N002 092R0295N004 092R0295N006 092R0295N007 092R0295N008 092R0295N009 092R0295N010 092R0295N012 092R0295N014 092R0295N015 092R0295N021 092R0295N023 092R0295N024 092R0295N051 092R0295N052 092R0295N054 092R0295N055 092R0295N056 092R0295N057 092R0295N060 092R0295N061 092R0295N062 092R0295N063 092R0295N064 092R0295N065 092R0295N066 092R0295N067 092R0295N068 092R0295N069 092R0295N070 092R0295N071 092R0295N072 092R0295N073 092R0295N074 092R0295N075 092R0295N076 092R0295N079 092R0295N080 092R0296N001 092R0296N002 092R0296N004 092R0296N005 092R0296N006 092R0296N007 092R0296N008 092R0296N011 092R0296N012 092R0296N013 092R0296N014 092R0296N056 092R0297N005 092R0297N006 092R0297N011 092R0297N012 092R0297N015 092R0297N019 092R0297N020 092R0297N023 092R0297N024 092R0297N025 092R0297N026 092R0297N027 092R0297N028 092R0297N029 092R0297N030 092R0297N031 092R0297N032 092R0297N033 092R0297N034 092R0297N035 092R0298N001 092R0298N002 092R0298N004 092R0298N006 092R0298N007 092R0298N028 092R0298N044 092R0298N053 092R0298N054 092R0298N056 092R0298N057 092R0298N058 092R0298N059 092R0298N060 092R0298N061 092R0298N062 092R0298N063 092R0298N064 092R0298N065 092R0298N066 092R0298N067 092R0298N068 092R0298N069 092R0298N070 092R0298N071 092R0298N072 092R0298N073 092R0298N075 092R0298N076 092R0298N077 092R0298N078 092R0298N080 092R0298N082 092R0299N004 092R0299N009 092R0299N014 092R0299N015 092R0299N016 092R0299N018 092R0299N019 092R0299N022 092R0299N024 092R0299N027 092R0299N030 092R0299N031 092R0299N040 092R0301N005 092R0301N006 092R0301N014 092R0301N015 092R0301N016 092R0301N018 092R0301N020 092R0301N022 092R0301N023 092R0302N010 092R0302N016 092R0302N019 092R0302N021 092R0302N022 092R0302N024 092R0302N025 092R0302N027 092R0302N030 092R0302N031 092R0302N033 092R0302N034 092R0302N035 092R0302N036 092R0303N021 092R0303N032 092R0303N037 092R0303N045 092R0304N018 092R0304N019 092R0304N020 092R0304N021 092R0304N022 092R0304N023 092R0304N024 092R0304N025 092R0304N026 092R0304N032 092R0304N045 092R0304N049 092R0304N050 092R0305N016 092R0305N030 092R0305N033 092R0305N038 092R0306N023 092R0306N024 092R0306N025 092R0306N026 092R0306N027 092R0306N028 092R0306N029 092R0306N030 092R0306N032 092R0306N040 092R0306N041 092R0306N043 092R0306N044 092R0306N045 092R0306N046 092R0306N047 092R0306N048 092R0306N049 092R0306N050 092R0306N051 092R0306N052 092R0306N053 092R0306N054 092R0306N055 092R0306N058 092R0306N065 092R0306N069 092R0306N078 092R0306N079 092R0308N002 092R0308N003 092R0308N004 092R0308N005 092R0308N006 092R0308N007 092R0308N009 092R0308N012 092R0308N013 092R0308N014 092R0308N016 092R0308N018 092R0308N025 092R0308N026 092R0308N027 092R0308N028 092R0308N037 092R0308N039 092R0308N040 092R0308N043 092R0309N003 092R0309N004 092R0309N009 092R0310N013 092R0310N016 092R0310N055 092R0310N057 092R0310N071 092R0311N007 092R0311N008 092R0311N009 092R0311N010 092R0311N011 092R0311N012 092R0311N013 092R0311N017 092R0311N018 092R0311N022 092R0311N023 092R0311N028 092R0311N029 092R0311N031 092R0311N048 092R0311N049 092R0311N050 092R0312N001 092R0312N002 092R0312N004 092R0312N005 092R0312N006 092R0312N007 092R0312N008 092R0312N009 092R0312N010 092R0312N011 092R0312N012 092R0312N013 092R0313N002 092R0313N004 092R0313N005 092R0313N008 092R0313N010 092R0313N014 092R0313N015 092R0314N001 092R0314N006 092R0314N013 092R0314N015 092R0315N002 092R0315N007 092R0315N008 092R0315N009 092R0325N010 092R0326N001 092R0327N001 092R0327N006 092R0327N009 092R0327N012 092R0327N015 092R0328N008 092R0328N015 092R0328N016 092R0329N001 092R0329N005 092R0329N009 092R0329N010 092R0329N012 092R0329N013 092R0329N014 092R0330N001 092R0330N002 092R0330N003 092R0330N005 092R0330N006 092R0330N007 092R0330N011 092R0330N018 092R0330N019 092R0330N020 092R0330N022 092R0330N023 092R0330N025 092R0330N026 092R0330N028 092R0330N029 092R0330N035 092R0330N036 092R0330N037 092R0330N038 092R0330N039 092R0330N040 092R0330N041 092R0330N043 092R0330N045 092R0330N047 092R0330N048 092R0331N004 092R0331N009 092R0331N011 092R0331N012 092R0331N013 092R0331N016 092R0331N018 092R0331N019 092R0336N024 092R0338N031 092R0339N003 092R0339N004 092R0339N005 092R0339N010 092R0339N011 092R0339N012 092R0339N013 092R0339N014 092R0339N015 092R0339N020 092R0339N023 092R0339N025 092R0340N021 092R0340N027 092R0340N029 092R0342N012 092R0342N013 092R0342N018 092R0342N020 092R0343N001 092R0343N006 092R0343N009 092R0343N011 092R0343N014 092R0343N015 092R0343N016 092R0343N020 092R0344N001 092R0344N003 092R0345N005 092R0345N012 092R0345N013 092R0345N014 092R0345N019 092R0347N003 092R0347N007 092R0347N009 092R0348N001 092R0348N004 092R0348N006 092R0348N008 092R0348N011 092R0348N012 092R0348N013 092R0348N015 092R0348N016 092R0348N017 092R0348N018 092R0348N019 092R0348N020 092R0349N002 092R0349N004 092R0349N005 092R0349N007 092R0349N008 092R0349N009 092R0349N010 092R0349N011 092R0350N001 092R0351N001 092R0352N013 092R0352N014 092R0352N016 092R0352N019 092R0352N024 092R0353N004 092R0353N005 092R0353N007 092R0353N010 092R0354N001 092R0354N003 092R0354N004 092R0354N005 092R0354N006 092R0354N007 092R0354N008 092R0354N009 092R0354N010 092R0354N011 092R0354N014 092R0355N002 092R0355N003 092R0355N004 092R0355N005 092R0355N006 092R0355N007 092R0355N008 092R0355N009 092R0355N012 092R0357N010 092R0357N012 092R0357N024 092R0357N030 092R0358N001 092R0358N007 092R0358N011 092R0358N012 092R0358N013 092R0358N014 092R0359N001 092R0359N003 092R0359N005 092R0359N007 092R0359N008 092R0360N009 092R0360N015 092R0361N001 092R0361N002 092R0361N004 092R0361N005 092R0361N006 092R0361N007 092R0361N008 092R0361N009 092R0361N010 092R0361N011 092R0361N012 092R0361N013 092R0361N014 092R0362N003 092R0362N020 092R0362N021 092R0362N022 092R0362N027 092R0362N028 092R0362N029 092R0363N004 092R0363N006 092R0363N008 092R0363N009 092R0363N010 092R0363N013 092R0364N014 092R0365N005 092R0366N002 092R0366N005 092R0366N006 092R0366N007 092R0366N012 092R0366N013 092R0366N014 092R0366N015 092R0366N018 092R0366N019 092R0366N020 092R0367N001 092R0367N002 092R0367N003 092R0367N004 092R0367N006 092R0367N009 092R0367N010 092R0367N013 092R0367N014 092R0367N015 092R0367N018 092R0368N001 092R0368N003 092R0368N004 092R0368N006 092R0368N007 092R0368N010 092R0368N011 092R0368N012 092R0368N013 092R0368N014 092R0369N005 092R0369N008 092R0369N011 092R0369N012 092R0369N013 092R0369N014 092R0370N001 092R0370N008 092R0370N012 092R0371N004 092R0371N005 092R0371N007 092R0371N009 092R0371N010 092R0371N011 092R0371N012 092R0371N013 092R0371N014 092R0371N015 092R0371N017 092R0371N018 092R0371N020 092R0372N004 092R0372N007 092R0372N008 092R0372N010 092R0372N011 092R0372N012 092R0372N013 092R0372N014 092R0372N017 092R0372N018 092R0373N001 092R0373N003 092R0373N004 092R0373N005 092R0373N006 092R0373N007 092R0373N008 092R0373N009 092R0373N010 092R0374N001 092R0374N002 092R0374N003 092R0375N009 092R0375N011 092R0375N012 092R0376N003 092R0376N005 092R0376N010 092R0376N012 092R0376N014 092R0377N002 092R0377N004 092R0377N006 092R0377N011 092R0377N012 092R0377N014 092R0377N015 092R0377N017 092R0377N018 092R0377N019 092R0377N020 092R0378N001 092R0378N002 092R0378N004 092R0378N006 092R0378N007 092R0378N008 092R0378N009 092R0378N010 092R0378N011 092R0378N013 092R0378N014 092R0378N015 092R0378N016 092R0378N017 092R0378N019 092R0378N020 092R0378N021 092R0378N023 092R0378N024 092R0378N026 092R0378N032 092R0378N035 092R0378N036 092R0379N001 092R0379N012 092R0379N015 092R0379N017 092R0379N018 092R0379N020 092R0379N021 092R0379N022 092R0379N025 092R0380N008 092R0380N011 092R0380N013 092R0380N014 092R0380N015 092R0380N026 092R0380N030 092R0381N002 092R0381N010 092R0381N013 092R0381N014 092R0381N016 092R0381N018 092R0381N021 092R0381N027 092R0381N029 092R0381N031 092R0381N034 092R0381N036 092R0381N039 092R0382N001 092R0382N003 092R0382N004 092R0382N005 092R0382N006 092R0382N007 092R0382N008 092R0382N009 092R0382N012 092R0382N026 092R0382N034 092R0382N036 092R0382N037 092R0382N038 092R0382N039 092R0382N040 092R0383N005 092R0383N008 092R0383N009 092R0383N010 092R0383N011 092R0383N017 092R0383N018 092R0383N019 092R0383N022 092R0383N023 092R0383N024 092R0383N025 092R0383N026 092R0383N027 092R0383N028 092R0383N029 092R0383N030 092R0383N031 092R0383N032 092R0383N033 092R0383N034 092R0383N035 092R0383N036 092R0383N037 092R0383N038 092R0383N039 092R0383N040 092R0384N001 092R0384N002 092R0384N003 092R0384N004 092R0384N005 092R0384N006 092R0384N007 092R0384N010 092R0384N011 092R0385N001 092R0385N008 092R0385N011 092R0385N012 092R0385N013 092R0385N014 092R0385N015 092R0385N016 092R0385N017 092R0385N018 092R0385N019 092R0385N020 092R0385N023 092R0386N003 092R0386N005 092R0386N008 092R0386N009 092R0386N018 092R0386N019 092R0386N020 092R0386N023 092R0386N025 092R0386N026 092R0386N027 092R0386N028 092R0387N001 092R0387N003 092R0387N004 092R0387N005 092R0387N006 092R0387N008 092R0387N009 092R0387N010 092R0387N011 092R0387N012 092R0387N013 092R0387N014 092R0387N015 092R0387N018 092R0387N020 092R0387N021 092R0387N022 092R0387N023 092R0387N024 092R0387N027 092R0387N028 092R0387N031 092R0387N035 092R0387N036 092R0387N037 092R0387N038 092R0388N002 092R0388N003 092R0388N005 092R0388N006 092R0388N007 092R0388N008 092R0388N009 092R0388N010 092R0388N011 092R0388N012 092R0388N013 092R0388N015 092R0389N011 092R0389N012 092R0389N016 092R0389N022 092R0389N024 092R0389N025 092R0389N026 092R0390N003 092R0390N017 092R0391N001 092R0391N002 092R0391N027 092R0391N030 092R0392N003 092R0392N004 092R0392N010 092R0392N013 092R0392N017 092R0392N023 092R0392N024 092R0392N026 092R0392N028 092R0392N030 092R0393N003 092R0394N008 092R0395N026 092R0395N028 092R0395N029 092R0395N030 092R0396N001 092R0396N002 092R0396N003 092R0396N005 092R0396N006 092R0396N009 092R0397N004 092R0397N005 092R0397N007 092R0397N011 092R0397N012 092R0397N013 092R0397N014 092R0397N015 092R0397N017 092R0397N018 092R0397N019 092R0397N025 092R0397N026 092R0397N027 092R0397N028 092R0397N029 092R0398N008 092R0399N001 092R0400N002 092R0400N005 092R0400N010 092R0400N011 092R0400N012 092R0400N024 092R0400N029 092R0400N030 092R0401N007 092R0401N012 092R0401N018 092R0401N019 092R0401N020 092R0401N023 092R0401N028 092R0401N030 092R0402N003 092R0402N004 092R0402N007 092R0402N008 092R0402N009 092R0402N012 092R0402N020 092R0403N003 092R0405N023 092R0406N002 092R0406N005 092R0406N007 092R0406N008 092R0406N009 092R0406N011 092R0406N012 092R0406N013 092R0406N015 092R0406N016 092R0406N018 092R0406N022 092R0406N026 092R0406N027 092R0406N029 092R0406N030 092R0407N001 092R0407N002 092R0407N003 092R0407N004 092R0407N005 092R0407N006 092R0414N020 092R0414N022 092R0414N023 092R0414N025 092R0415N006 092R0415N008 092R0415N012 092R0415N014 092R0415N015 092R0415N018 092R0415N019 092R0415N022 092R0415N029 092R0415N030 092R0416N004 092R0416N005 092R0416N006 092R0416N007 092R0416N008 092R0416N009 092R0416N010 092R0416N011 092R0416N012 092R0416N013 092R0416N023 092R0416N024 092R0416N027 092R0416N028 092R0417N001 092R0417N003 092R0417N004 092R0417N006 092R0417N007 092R0417N008 092R0417N013 092R0417N014 092R0417N015 092R0417N016 092R0417N017 092R0417N018 092R0417N020 092R0417N024 092R0417N028 092R0418N006 092R0418N008 092R0418N009 092R0418N010 092R0418N011 092R0418N012 092R0418N013 092R0418N014 092R0418N015 092R0418N016 092R0418N017 092R0418N020 092R0418N021 092R0418N022 092R0418N025 092R0418N026 092R0418N027 092R0418N030 092R0419N002 092R0419N004 092R0419N005 092R0419N008 092R0419N009 092R0419N010 092R0419N011 092R0419N012 092R0419N013 092R0419N014 092R0419N015 092R0419N016 092R0419N018 092R0419N020 092R0419N022 092R0419N025 092R0420N002 092R0420N004 092R0420N005 092R0420N006 092R0420N007 092R0420N009 092R0420N010 092R0420N011 092R0420N012 092R0420N018 092R0420N022 092R0420N024 092R0420N027 092R0420N028 092R0420N029 092R0420N030 092R0421N002 092R0421N007 092R0421N008 092R0421N009 092R0421N011 092R0425N020 092R0425N021 092R0425N023 092R0425N025 092R0425N028 092R0426N001 092R0426N003 092R0426N005 092R0426N006 092R0426N007 092R0426N008 092R0426N009 092R0426N010 092R0426N011 092R0426N012 092R0426N013 092R0426N014 092R0426N015 092R0426N016 092R0426N017 092R0426N018 092R0426N019 092R0426N020 092R0426N021 092R0426N022 092R0426N024 092R0426N025 092R0426N026 092R0426N028 092R0426N029 092R0426N030 092R0426N031 092R0426N032 092R0426N033 092R0426N034 092R0426N035 092R0426N036 092R0426N037 092R0426N038 092R0426N039 092R0426N040 092R0427N001 092R0427N003 092R0427N004 092R0427N005 092R0427N006 092R0427N007 092R0427N009 092R0427N011 092R0427N012 092R0427N013 092R0427N014 092R0427N015 092R0427N016 092R0427N017 092R0427N018 092R0427N019 092R0427N021 092R0427N022 092R0427N023 092R0427N024 092R0427N029 092R0427N030 092R0427N032 092R0427N033 092R0427N036 092R0427N037 092R0427N038 092R0427N039 092R0427N040 092R0428N002 092R0428N003 092R0428N004 092R0428N005 092R0428N006 092R0428N007 092R0428N008 092R0428N009 092R0428N010 092R0428N011 092R0428N012 092R0428N014 092R0428N025 092R0428N026 092R0428N027 092R0429N004 092R0429N007 092R0429N022 092R0429N023 092R0429N024 092R0429N025 092R0429N027 092R0429N028 092R0429N029 092R0429N030 092R0431N008 092R0431N009 092R0431N012 092R0431N013 092R0433N001 092R0434N023 092R0435N015 092R0436N005 092R0436N007 092R0436N008 092R0436N009 092R0436N010 092R0436N012 092R0436N013 092R0436N014 092R0436N017 092R0436N018 092R0436N019 092R0436N020 092R0436N023 092R0438N020 092R0439N002 092R0439N004 092R0439N010 092R0439N015 092R0439N019 092R0441N002 092R0441N005 092R0441N018 092R0442N021 092R0443N006 092R0448N001 092R0448N003 092R0448N004 092R0448N010 092R0448N017 092R0448N024 092R0448N029 092R0450N020 092R0450N021 092R0450N022 092R0450N024 092R0450N028 092R0450N030 092R0451N002 092R0451N004 092R0451N005 092R0451N006 092R0451N007 092R0451N008 092R0451N010 092R0451N028 092R0451N029 092R0452N001 092R0452N003 092R0452N005 092R0452N006 092R0452N007 092R0452N009 092R0452N010 092R0452N016 092R0452N019 092R0452N028 092R0454N001 092R0454N010 092R0454N019 092R0454N023 092R0454N024 092R0455N025 092R0456N004 092R0456N006 092R0456N012 092R0456N013 092R0456N014 092R0456N021 092R0456N025 092R0456N029 092R0459N024 092R0459N029 092R0459N030 092R0460N008 092R0460N009 092R0460N015 092R0460N016 092R0460N017 092R0460N018 092R0460N019 092R0460N020 092R0460N022 092R0460N023 092R0460N024 092R0460N025 092R0460N026 092R0460N027 092R0461N005 092R0461N008 092R0461N009 092R0461N010 092R0461N012 092R0461N020 092R0461N026 092R0461N028 092R0461N029 092R0462N001 092R0462N002 092R0462N005 092R0462N006 092R0462N008 092R0462N009 092R0462N010 092R0462N011 092R0462N015 092R0462N028 092R0462N029 092R0462N030 092R0463N001 092R0463N004 092R0463N006 092R0464N020 092R0464N023 092R0465N014 092R0466N025 092R0469N013 092R0469N014 092R0469N015 092R0469N016 092R0469N017 092R0469N018 092R0469N028 092R0469N029 092R0469N030 092R0469N031 092R0469N032 092R0469N033 092R0469N034 092R0469N035 092R0470N005 092R0470N010 092R0470N011 092R0470N012 092R0470N015 092R0470N016 092R0470N017 092R0470N018 092R0470N019 092R0470N020 092R0470N021 092R0470N022 092R0470N023 092R0470N024 092R0470N025 092R0470N026 092R0470N028 092R0471N001 092R0471N005 092R0471N029 092R0471N031 092R0471N033 092R0471N036 092R0471N038 092R0471N039 092R0476N001 092R0476N002 092R0476N007 092R0476N010 092R0476N011 092R0476N012 092R0476N015 092R0476N017 092R0476N020 092R0476N021 092R0476N022 092R0476N023 092R0476N024 092R0476N025 092R0476N029 092R0477N002 092R0478N029 092R0479N004 092R0479N009 092R0479N022 092R0479N027 092R0480N003 092R0480N004 092R0480N009 092R0480N010 092R0480N011 092R0480N012 092R0480N013 092R0480N017 092R0480N019 092R0480N022 092R0480N024 092R0480N026 092R0480N029 092R0481N003 092R0481N004 092R0481N005 092R0481N006 092R0481N007 092R0481N008 092R0481N010 092R0481N011 092R0482N013 092R0482N016 092R0482N018 092R0484N030 092R0485N019 092R0485N025 092R0485N026 092R0485N029 092R0486N003 092R0486N015 092R0486N016 092R0486N020 092R0486N022 092R0486N024 092R0486N029 092R0487N003 092R0487N006 092R0487N007 092R0487N008 092R0487N009 092R0487N010 092R0487N011 092R0487N019 092R0487N020 092R0487N023 092R0487N024 092R0487N027 092R0488N002 092R0488N004 092R0488N007 092R0488N008 092R0488N009 092R0488N010 092R0488N011 092R0488N014 092R0488N017 092R0489N024 092R0489N025 092R0489N026 092R0489N027 092R0489N029 092R0489N030 092R0490N001 092R0490N003 092R0490N004 092R0490N006 092R0490N007 092R0490N010 092R0490N011 092R0490N012 092R0490N016 092R0490N017 092R0490N018 092R0490N019 092R0490N020 092R0490N021 092R0490N024 092R0490N025 092R0490N029 092R0490N030 092R0491N002 092R0491N004 092R0491N005 092R0491N013 092R0491N017 092R0491N019 092R0491N021 092R0491N022 092R0493N013 092R0493N026 092R0493N029 092R0494N011 092R0494N013 092R0495N004 092R0495N006 092R0495N017 092R0495N020 092R0498N027 092R0499N006 092R0499N007 092R0499N011 092R0499N013 092R0499N014 092R0499N015 092R0499N016 092R0499N017 092R0499N018 092R0500N019 092R0501N013 092R0501N018 092R0504N006 092R0504N007 092R0504N011 092R0504N013 092R0504N014 092R0504N015 092R0504N016 092R0504N018 092R0505N001 092R0505N002 092R0505N003 092R0505N004 092R0505N009 092R0505N015 092R0505N016 092R0505N017 092R0506N015 092R0506N016 092R0506N017 092R0506N019 092R0506N020 092R0506N021 092R0506N023 092R0506N024 092R0506N025 092R0507N001 092R0507N003 092R0507N006 092R0507N007 092R0507N008 092R0507N009 092R0507N011 092R0507N012 092R0507N016 092R0507N017
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- ABL90 FLEX PLUS Analyzer 393-092Radiometer Medical ApS · BILIRUBIN (TOTAL AND UNBOUND) IN THE NEONATE TEST SYSTEM1 recall1 death175 reports
- ABL90 FLEX Analyzer 393-090Radiometer Medical ApS · ELECTRODE MEASUREMENT, BLOOD-GASES (PCO2, PO2) AND BLOOD PH1 recall112 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page