Brand Name: ID NOW COVID-19 2.0 24T recallZ-1571-2025 · Abbott Diagnostics Scarborough, Inc. · initiated Mar 4, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Abbott Diagnostics Scarborough, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The impacted lots have a higher occurrence of invalid rates when compared to the product Instructions for Use.
FDA root cause category: Process change control
What the company is doing
On March 4, 2025 FIELD CORRECTION NOTICE letters were sent to end user customers. Please complete the following actions, as applicable. 1. If you have impacted inventory in stock, then: " Discontinue use of and destroy any remaining inventory of the impacted lots according to your procedures. " Complete and return the Customer Reply Form (Form must be completed, signed and returned to receive a replacement lot). " Please retain this letter for your records. 2. If you have forwarded the product listed above to others in your network, then: - Inform them of this Field Correction Notice, provide to them a copy of this notice and request they take the necessary action. 3. If you do not have impacted inventory in stock, then: - All product lots not identified in Appendix I can continue to be used. If you have questions regarding this information, please contact Sedgwick by phone at 888 671 8858 (available 8am 5pm EST, Monday - Friday). Adverse reactions or quality problems experienced with the use of this product may be reported to the FDA s MedWatch Adverse Event Reporting program either online (http://www.fda.gov/MedWatch/report.htm), by mail (http://www.fda.gov/MedWatch/getforms.htm), by phone (1-800-332-1088), or by fax (1-800-FDA-0178). If you have experienced any patient or user injury associated with this Field Correction Notice, please immediately report the event to your local area Customer Service. ****Update added 5/8/2025**** On April 29, 2025 FIELD CORRECTION NOTICE letters were sent to customers informing them of additional lots included in this recall.
Recall details
- Recall number
- Z-1571-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 4, 2025
- Posted by the FDA
- Apr 15, 2025
- Recalling firm
- Abbott Diagnostics Scarborough, Inc.
- Quantity
- 339456 eaches (337560 US, 1896 OUS)
- Product code
- QWR
- FDA submissions
- K221925
- Distribution
- Worldwide distribution - US Nationwide and the countries of Australia; Austria; Belgium; Bulgaria; Canada; Costa Rica; France; French Polynesia; Germany; Italy; Luxembourg; Malaysia; Netherlands; Paraguay; Poland; Qatar; Saudi Arabia; Slovenia; South Korea; Spain; Switzerland; Taiwan; United Kingdom; Uruguay.
Product description
Brand Name: ID NOW COVID-19 2.0 24T Product Name: ID NOW COVID-19 2.0 24T Model/Catalog Number: 192-000 Software Version: N/A Product Description: Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets Component: No
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
Model No 192-000; GTIN/DI 10811877011351; Lot 000M884910 Model No 192-000; GTIN/DI 10811877011351; Lot 000M885121 Model No 192-000; GTIN/DI 10811877011351; Lot 000M911157 Model No 192-000; GTIN/DI 10811877011351; Lot 000M913227 Model No 192-000; GTIN/DI 10811877011351; Lot 000M913286 Model No 192-000; GTIN/DI 10811877011351; Lot 000M913303 Model No 192-000; GTIN/DI 10811877011351; Lot 000M913641 Model No 192-000; GTIN/DI 10811877011351; Lot 000M914188 Model No 192-000; GTIN/DI 10811877011351; Lot 000M914221 Model No 192-000; GTIN/DI 10811877011351; Lot 000M914852 Model No 192-000; GTIN/DI 10811877011351; Lot 000M915373 Model No 192-000; GTIN/DI 10811877011351; Lot 000M916267 Model No 192-000; GTIN/DI 10811877011351; Lot 000M921404 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922204 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922578 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922601 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922634 Model No 192-000; GTIN/DI 10811877011351; Lot 000M924048 Model No 192-000; GTIN/DI 10811877011351; Lot 000M925615 Model No 192-000; GTIN/DI 10811877011351; Lot 000M938692 Model No 192-000; GTIN/DI 10811877011351; Lot 00M922634A Model No 192-000; GTIN/DI 10811877011351; Lot 00M924048A ****Updated 5/8/2025**** 04/29/2025 (Adding Additional Lots): Model No 192-000; GTIN/DI 10811877011351; Lot 000M913147 Model No 192-000; GTIN/DI 10811877011351; Lot 000M913159 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922950 Model No 192-000; GTIN/DI 10811877011351; Lot 000M922988 Model No 192-000; GTIN/DI 10811877011351; Lot 000M926188
Other devices under the same FDA submission
1 device cleared under K221925 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ID NOW COVID-19 2.0 192-000Alere Scarborough, Inc. · Simple Point-Of-Care Device To Detect Sar-Cov-2 Nucleic Acid Targets From Clinical Specimens In Near-Patient Settings147 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96519).
- Z-1572-2025 · Open, ClassifiedBrand Name: ID NOW Influenza A/B 2 24T Product Name: ID NOW Influenza A/B 2 24T Model/Catalog Number: 427-000 Software Version: N/A Product Description: Influenza A And Influenza B…
- Z-1573-2025 · Open, ClassifiedBrand Name: ID NOW RSV 24T Product Name: ID NOW RSV 24T Model/Catalog Number: 435-000 Software Version: N/A Product Description: Respiratory virus panel nucleic acid assay system C…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page