Skip to content

Amplatzer TorqVue recallZ-1644-2025 · Abbott · initiated Mar 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Abbott or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device may have a small breach in the proximal end of the shaft under the strain relief of the delivery system. Breach is not visibly detectable and may lead to prolonged procedure, blood loss, or air ingress with potential for air embolism.

FDA root cause category: Process control

What the company is doing

On March 17, 2025, the firm notified customers via hand-delivered customer communications titled "Urgent Medical Device Recall." Customers were instructed to share the notice with applicable personnel and to return any remaining unused devices from impacted lots. Abbott representatives will assist with product return and replenishment. If you have any questions, please contact your local Sales Representative or Abbott Support at 1-800-544-1664.

Recall details

Recall number
Z-1644-2025
Status
Open, Classified
Initiated by the company
Mar 17, 2025
Posted by the FDA
Apr 21, 2025
Recalling firm
Abbott
Quantity
2683
Product code
DQY
FDA submissions
K162228
Distribution
Worldwide - US Nationwide distribution.

Product description

Amplatzer TorqVue LP Catheter (TVLPC), Model no. 9-TVLPC4F90/080

Code information (models, lots, serial numbers)

UDI-DI 00811806012926 Lots 9078966 9093895 9134623 9144113 9148110 9148579 9155868 9158302 9241591 9241593 10004059 10011461 10011462 10011463 10011465 10017790 10017792 10196106 10196119 10198795 10198801 10198889 10207472 10207503 10274454 10283818

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 96545).

Sources and updates