Beckman Coulter DxI 9000 Access Immunoassay Analyzer recallZ-1664-2025 · Beckman Coulter, Inc. · initiated Mar 27, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

When DxI 9000 Access Immunoassay Analyzer is connected to a host system (laboratory information system (LIS) or middleware) and has accumulated canceled QC test results that were not sent by LIS, the analyzer may lose communication with the host system. The lost connection interrupts sample processing, which delays reporting patient test results and may subsequently delay patient treatment.

FDA root cause category: Software design

What the company is doing

Beckman Coulter issued an URGENT MEDICAL DEVICE RECALL notice to its consignees on 03/27/2025 via USPS first class mail. The notice explained the issue with the device, impact to the patient, and provided a workaround until a software upgrade is available. If the consignee further distributed the affected device, they were directed to provide a copy of the notice to the recipient. For questions, contact Customer Support Center, from the website: http://www.beckmancoulter.com

Recall details

Recall number
Z-1664-2025
Status
Open, Classified
Initiated by the company
Mar 27, 2025
Posted by the FDA
Apr 24, 2025
Recalling firm
Beckman Coulter, Inc.
Quantity
370 units
Product code
JJE
FDA submissions
K221225
Distribution
Worldwide distribution.

Product description

Beckman Coulter DxI 9000 Access Immunoassay Analyzer

Code information (models, lots, serial numbers)

UDI/DI 15099590732103, all units run with system software versions SW 1.20.0 and below