HistoCore PELORIS 3, Model/Catalog Number:… recallZ-1683-2025 · Leica Biosystems Melbourne Pty Ltd · initiated Apr 2, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Leica Biosystems Melbourne Pty Ltd or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
There is a leakage issue associated with the tubing in the manifold of the instrument.
FDA root cause category: Release of Material/Component prior to receiving test results
What the company is doing
An URGENT MEDICAL DEVICE RECALL notification letter dated 3/31/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instruments tubing and reseat the tubes if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email address: lbsme/[email protected] Please contact your local Leica Biosystems representative immediately if you have any questions or concerns. An URGENT MEDICAL DEVICE RECALL notification letter dated 8/13/25 was sent to customers. ACTIONS REQUIRED 1. If leakage from the instrument is observed in the drip tray or surrounding the instrument, please follow your laboratory's leakage containment/safety protocol. If a protocol is running, wait for the run to finish, switch off the instrument and contact your local Leica Biosystems representative immediately. 2. If no leakage has been observed from the instrument, continue using it as normal. Conduct routine monitoring for leakage until the on-site inspection is completed. 3. Your local Leica Biosystems representative will contact you to schedule an onsite inspection of your instrument(s) for replacing the tubing and/or density meter manifold if needed. 4. Confirm Receipt of Recall Notice Please confirm receipt of this notice by signing and dating the attached Recall Acknowledgement Form by email to the following email ad
Recall details
- Recall number
- Z-1683-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 2, 2025
- Posted by the FDA
- Apr 28, 2025
- Recalling firm
- Leica Biosystems Melbourne Pty Ltd
- Quantity
- 77 units
- Product code
- IEO
- Distribution
- US States: AZ, CA, FL, MN, NJ, NY, NC, OH, PA, TN, TX, UT, VA, and WI.
Product description
HistoCore PELORIS 3, Model/Catalog Number: 45.0005/45.7512.501 A11, Software Version: 3.4.0. The HistoCore PELORIS 3 Rapid Tissue Processor is a dual retort rapid tissue processor used to prepare tissue samples.
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI-DI: 09349458004811; Serial Numbers: 45111117, 45111154, 45111155, 45111156, 45111157, 45111158, 45111159, 45111160, 45111161, 45111162, 45111164, 45111166, 45111167, 45111168, 45111169, 45111170, 45111171, 45111174, 45111175, 45111176, 45111177, 45111178, 45111179, 45111180, 45111181, 45111182, 45111183. 08/12/2025 UPDATE: Additional Serial Numbers: 45111064, 45111065, 45111067, 45111068, 45111069, 45111070, 45111071, 45111073, 45111074, 45111075, 45111076, 45111077, 45111078, 45111079, 45111080, 45111081, 45111082, 45111084, 45111086, 45111087, 45111088, 45111089, 45111090, 45111091, 45111092, 45111093, 45111094, 45111095, 45111097, 45111098, 45111099, 45111100, 45111101, 45111102, 45111103, 45111104, 45111108, 45111109, 45111110, 45111111, 45111112, 45111113, 45111114, 45111115, 45111116, 45111120, 45111122, 45111123, 45111124, 45111125, 45111126, 45111127, 45111128, 45111129, 45111130, 45111131, 45111132, 45111133, 45111134, 45111135, 45111136, 45111137, 45111139, 45111140, 45111141, 45111142, 45111143, 45111144, 45111145, 45111146, 45111147, 45111148, 45111149, 45111150, 45111151, 45111152, 45111163.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- HistoCore PELORIS HistoCore PELORIS 3 (100-120V)Leica Biosystems Melbourne Pty Ltd · Processor, Tissue, Automated1 recall28 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.