3.0 Liter Disposable Set (4 sets/case). Model Number:… recallZ-1678-2025 · Belmont Instrument LLC · initiated Mar 31, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Belmont Instrument LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential crack in the female connector located on the Heat Exchanger/patient line set, which will lead to a fluid leak through the crack during priming of the Rapid Infuser.
FDA root cause category: Process design
What the company is doing
Belmont Medical notified consignees on about 03/31/2025 via "URGENT MEDICAL DEVICE CORECTION" FedEx or emailed letter. Consignees were instructed to locate affected product and to return the customer Response Form indicating if any affected units were on hand. Consignees were also instructed to follow instructions provided in the letter to inspect affected lots and if any units are identified as having cracks or leaks during priming, they should be discarded. Lot number of the set should be noted and reported to Belmont Medical Technologies. Belmont has indicated that they will replace any product found defective or at the request of the user any affected product that has not been used.
Recall details
- Recall number
- Z-1678-2025
- Status
- Open, Classified
- Initiated by the company
- Mar 31, 2025
- Posted by the FDA
- Apr 25, 2025
- Recalling firm
- Belmont Instrument LLC
- Quantity
- 603 cases (2,412 sets)
- Product code
- LGZ
- FDA submissions
- K242735
- Distribution
- Nationwide distribution to AL, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, LA, MA, MD, ME, MI, MN, MO, NC, ND, NE, NJ, NV, NY, OH, OR, PA, SC, TX, UT, VA, WA, WI. International distribution to Czechia, Italy, Norway.
Product description
3.0 Liter Disposable Set (4 sets/case). Model Number: 903-00032. The P/N 903-00032 consists of 4 Large Volume Reservoirs (P/N 903-00018) and 4 Heat Exchanger/Patient Line set (P/N 403-00382). The two components are packaged separately in individual pouches and placed in the same case for shipment. The female quick connector that may be cracked is located on the Heat Exchanger/Patient Line set.
Code information (models, lots, serial numbers)
Model Number: 903-00032. UDI-DI: 00896128002534. Lot Numbers: 20240911, 20241004, 20241011, 20241105
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- 3.0 Liter Disposable Set 903-00032Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 report
Other devices under the same FDA submission
7 devices cleared under K242735 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- 3-Spike Disposable Set (1 Piece) 903-00006PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 recall1 death97 reports
- Rapid Infuser RI-2, 120V, 750 ml/min 903-00039Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid132 reports
- Rapid Infuser RI-2, 120V, 1000ml/min 903-00037Belmont Instrument Corporation · Warmer, Thermal, Infusion Fluid1 death76 reports
- 3.0 Liter Reservoir (1 Piece) 903-00018PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid14 reports
- 3-Spike Disposable Set 903-00006Belmont Instrument Corporation · Pump, Infusion8 reports
- Dual Patient Line (1 Piece) 903-00004PBelmont Instrument Corporation · Warmer, Thermal, Infusion Fluid5 reports
- 3.0 Liter Reservoir (6 Count) 903-00018Belmont Instrument Corporation · Pump, Infusion1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page