TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX recallZ-1807-2025 · Echopixel, Inc. · initiated Apr 11, 2016
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Echopixel, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Imaging system data may be displayed with an incorrect orientation (sidedness) where the patient side is not correct (left, right, superior, inferior, anterior or posterior). This may lead to an inaccurate measurements.
FDA root cause category: Software design
What the company is doing
On 03/25/2026, the firm sent via email and telephone an Urgent Field Safety Notice, informing customers of a potential misdiagnosis or incorrect surgical treatment planning approach can result when utilizing Enhanced Multi-frame DICOM image data sets. Customers were instructed to: verify if Enhanced Multi-Frame CT DICOM files have been used with the software. If users have utilized this file type, we request that users verify retrospectively any potential risk to health of patients and further determine if this patient needs to be reevaluated and report it to EchoPixel, Inc. We request that the True 3D Viewer versions 1.0, 1.25, 1.35 NOT be used with this file type. For any questions or concerns - contact EchoPixel, Inc Monday thru Friday 9 AM - 6 PM Pacific at: Mailing address: 2490 Hospital Dr. Suite 310, Mountain View CA 94040 Email: [email protected] Phone: +1 (844) 273-7766
Recall details
- Recall number
- Z-1807-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 11, 2016
- Posted by the FDA
- May 21, 2025
- Recalling firm
- Echopixel, Inc.
- Quantity
- 9 systems
- Product code
- LLZ
- FDA submissions
- K142107
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL and the country of Canada.
Product description
TRUE 3D VIEWER SYSTEM Model Number EP-00001-XX Software versions 1.0, 1.25, and 1.35 The t3D-Viewer is designed for use by health care professionals HCPs and is intended to assist the clinician who is responsible for making all patient management decisions. The True 3D Viewer (t3D-Viewer) system is comprised of a commercial off the shelf hardware platform and a proprietary software application that enables a health care professional (HCP) to visualize and interact with CT and MRI DICOM image data to assist in clinical decision making. The t3D-Viewer system hardware platform is comprised of an off the shelf stereoscopic display, an optical or electromagnetic motion tracking system and a computer system.
Code information (models, lots, serial numbers)
Model Number EP-00001-XX with software versions 1.0, 1.25, and 1.35 UDI-DI Codes: 00851325007003 Serial Numbers: Field Products: C2015-09-0001 C2015-09-0003 D2015-05-0002 C2015-06-0002 D2015-12-0004 C2015-08-0001 D2015-12-0005 C2015-09-0002 D2015-12-0001 C2016-01-0001 D2015-12-0002 C2015-12-0002 D2015-12-0003 D2016-01-0002
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- True 3D Viewer EP-00003Echopixel, Inc. · System, Image Processing, Radiological2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page