Brand Name: AirLife recallZ-1698-2025 · Vyaire Medical · initiated Apr 10, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vyaire Medical or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Adapters may disconnect during setup or while in use, potentially interrupting ventilation. Any disconnection can result in severe consequences for neonates, including hypoxia, hypercapnia, and organ failure.

FDA root cause category: Component design/selection

What the company is doing

On April 10, 2025, the firm notified customers via emailed letters titled URGENT: MEDICAL DEVICE RECALL. Customers are instructed to examine their inventory for the affected lots of product. Product in inventory should be quarantined. For in-use product - 1) For affected product in use that is utilizing the adaptor connection, immediately stop use of the product. 2) For affected product in use that is not utilizing the adaptor connection, ventilator circuit may continue to be utilized. Customers were provided with additional precautions to take to ensure proper function. Customers can choose between returning or destroying affected product. Firm will provide replacement product. If you have any questions regarding this field action, please call AirLife at 1-800-433-2797, or e-mail at [email protected].

Recall details

Recall number
Z-1698-2025
Status
Open, Classified
Initiated by the company
Apr 10, 2025
Posted by the FDA
May 9, 2025
Recalling firm
Vyaire Medical
Quantity
3053
Product code
BZE
FDA submissions
K151303
Distribution
Worldwide distribution - US Nationwide and the countries of Germany, Poland, Belgium, Spain, Romania, UK, Italy, Netherlands, India, Slovenia, Qatar, and Saudi Arabia.

Product description

Brand Name: AirLife Product Name: AirLife Infant Heated Wire Circuit Model/Catalog Number: AH265 Software Version: N/A Product Description: AirLife Infant Heated Wire Circuit Kit Dual-limb, Dual-heat, high-flow circuit (>4 L/min) Contains AH165 Circuit, AH290 Chamber Component: No. a conduit for respiratory gas between the patient and a ventilator.

Code information (models, lots, serial numbers)

Lot Code: Lot/Serial Number(s): 0004247924 0004247925 0004252940 0004252941 0004256292 0004256293 0004256294 0004256295 0004272325 0004272868 0004278978 0004279156 0004280036 0004285304 0004288036 0004289514 UDI-DI: AirLife Label Each: 10889483595909 Case: 30889483595903 Vyaire Label: Each: 10190752145160 Case: 50190752145168

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K151303 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 96664).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page