True 3D Viewer, Model: EP-00003 recallZ-1785-2025 · Echopixel, Inc. · initiated Feb 21, 2018
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Echopixel, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
With certain software versions, import of enhanced ultrasound formatted data/ultrasound images into 3D viewer, (for digital image processing/review/analysis/communication/media interchange acquired from CT/MRI/ultrasound, and pre-operative analysis of surgical options), can result in incorrect display of image orientation, which could result in misdiagnosis or incorrect surgical treatment plan.
FDA root cause category: Device Design
What the company is doing
On 2/21/2018, correction notices were emailed to customers who were asked to do the following: a) Affected software versions should not be used with Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM. b) If clinicians are utilizing Enhanced 3D Ultrasound DICOM compliant volumes or Philips 3DDCM volume data sets, verify, retrospectively, any potential risk to health to patients and further determine if patients need to be reevaluated and report this information to the firm. c) Pass a copy of this notice to those staff members who need to be aware of the issue and maintain awareness for a three-month period. d)If devices were transferred to other departments or organizations, provide a copy of this notice to these consignees. Firm will contact customers to provide information about the timing and delivery mechanisms of the software upgrade. If you have questions, contact the firm at: 1-844-273-7766 x2
Recall details
- Recall number
- Z-1785-2025
- Status
- Open, Classified
- Initiated by the company
- Feb 21, 2018
- Posted by the FDA
- May 16, 2025
- Recalling firm
- Echopixel, Inc.
- Quantity
- 18
- Product code
- LLZ
- FDA submissions
- K170167
- Distribution
- Worldwide - US Nationwide distribution in the states of CA, OH, NJ, PA, MA, FL, CT, TX, LA, MI, MN and the country of CA.
Product description
True 3D Viewer, Model: EP-00003
Code information (models, lots, serial numbers)
UDI-DI: 00851325007003. Full release software versions: 1.6.1.3080.428, released 13-Apr-2017 1.6.2.3670.435, released 07-Dec-2017 Develop Versions, for research use: 0.00.2695.422, released 15-Feb-2017 0.00.2927.423, released 15-Feb-2017 0.00.2964.423, released 24-Feb-2017 0.00.3102.428, released 04-Apr-2017 0.00.3245.428, released 05-May-2017 0.00.3302.431, released 02-Jun-2017 0.00.3344.432, released 18-Jul-2017 0.00.3412.433, released 14-Aug-2017 0.00.3656.442, released 09-Nov-2017 0.00.3669.442, released 05-Dec-2017 0.00.3683.445, released 02-Jan-2018 0.00.3702.447, released 11-Jan-2018 0.00.3730.447, released 08-Feb-2018
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- True 3D Viewer EP-00003Echopixel, Inc. · System, Image Processing, Radiological2 recalls1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page