Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS recallZ-1767-2025 · Baxter Healthcare Corporation · initiated Apr 24, 2025

Class I Open, Classified All recalls posted in May 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Baxter Healthcare Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

There is a potential for the Novum IQ LVP due to the potential for underinfusion following use of the "standby mode" feature or if the device is powered off with the set loaded.

FDA root cause category: Device Design

What the company is doing

Baxter issued an URGENT MEDICAL DEVICE CORRECTION notice to its consignees on 04/24/2025 via USPS First Class Mail. The notice explained the issue, hazard, and requested the following: Actions to be Taken by Customers 1. For flow rates greater than 50 mL/hour, do not exceed a programmed standby time of 2 hours and 30 minutes. Monitor patients frequently to ensure that the appropriate infusion is being delivered. 2. Please remove the set upon powering off the device. 3. Post the enclosed informational poster with Novum IQ LVPs in your facility. 4. Acknowledge receipt of this notification by following the instructions on the enclosed reply instruction sheet, even if you have no remaining inventory. Acknowledging receipt will prevent you from receiving repeat notices. If you do not complete the acknowledgement, you will receive a phone call from OnProcess Technology on behalf of Baxter to confirm your receipt of this notification. 5. Please forward a copy of this communication to the Chief Medical Officer, Medical Director, Director of Nursing, Director of Pharmacy, Facility Risk Manager, Director of Purchasing/Central Supply, and any other departments within your institution who use the affected product. For general questions regarding this communication, or if you experience quality problems, please contact your Baxter sales representative, or Baxter Global Technical Services at 800-843-7867 (select option 2, then option 2 again) Monday through Friday, between 7:00 am and 7:00 pm Eastern Time.

Recall details

Recall number
Z-1767-2025
Recall class
Class I (most serious)
Status
Open, Classified
Initiated by the company
Apr 24, 2025
Posted by the FDA
May 20, 2025
Recalling firm
Baxter Healthcare Corporation
Quantity
34,524 units
Product code
FRN
FDA submissions
K211122
Distribution
US and Canada

Product description

Baxter Novum IQ LVP INFUSION SYSTEM, REF 40700BAXUS

Code information (models, lots, serial numbers)

All serial numbers

Other devices under the same FDA submission

1 device cleared under K211122 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.