Skip to content

Halyard CHRIST SUTURE KIT. General Procedure Tray… recallZ-1741-2025 · AVID Medical, Inc. · initiated Mar 31, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from AVID Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

An issue with a raw material used to coat some surgical needles that may result in diminished performance and/or removal of some silicone needle coating during use.

FDA root cause category: Nonconforming Material/Component

What the company is doing

An URGENT: MEDICAL DEVICE RECALL NOTIFICATION dated 3/31/25 was emailed to consignees. The notification indicates that this recall is due to Ethicon having recalled PDS II, PDS Plus, PROLENE, and ATHIBOND EXCEL sutures, which are components in the convenience kits being recalled by the recalling firm. Consignees are being instructed to discontinue use of the Johnson & Johnson Ethicon Sutures contained inside the recalled AVID kits identified in the recall notification and to add the provided labels to affected kits to ensure recalled sutures are discarded. Distributors are asked to forward the provided recall notification to their customers to which product was further distributed. Consignees are asked to return the provided response form via email. Questions regarding this recall event are to be directed to your distributor or sales representative.

Recall details

Recall number
Z-1741-2025
Status
Open, Classified
Initiated by the company
Mar 31, 2025
Posted by the FDA
May 9, 2025
Recalling firm
AVID Medical, Inc.
Quantity
72 units
Product code
LRO
Distribution
US Nationwide distribution in the states of IL & LA.

Product description

Halyard CHRIST SUTURE KIT. General Procedure Tray Convenience Kit

Code information (models, lots, serial numbers)

Model No. ADVC800-03; UDI: 10809160450703; Lot Code: 1640852; Exp: 2/4/2028.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other products in the same recall event

The FDA groups related products into one recall event (event 96731).