Apo B Reagent, REF: OSR6143 recallZ-1815-2025 · Beckman Coulter Ireland, Inc. · initiated Apr 23, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Beckman Coulter Ireland, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The product Value Assignment Sheet (VAS) is used in calibration of the device. An erroneous value was included in the VAS which has the potential to generate an incorrect APO B patient result which could exceed the total allowable error for the APO B assay (11.6%). The error results in results with a positive bias up to 42%. However, based on the mitigating factors the risk to health is considered to be negligible or limited.

FDA root cause category: Error in labeling

What the company is doing

On April 23, 2025, Beckman Coulter issued a "Urgent Medical Device Recall" Notification to affected consignees via mail and E-Mail. Beckman Coulter asked consignees to take the following actions: 1. If your laboratory is reporting Apo B results generated using the combination of reagent OSR6143, Lot 2631 and calibrator ODR3022, Lot 3000 in mg/dL units, a retrospective review of patient results greater than 77 mg/dL and less than 176 mg/dL is recommended. 2. Discard the value assignment sheet located within each kit of Apo B OSR6143 Lot 2631. 3 Download the updated VAS kit insert from the Beckman Coulter website by searching LOSR6143_2631_ML.02. Beckman Coulter Diagnostics | Beckman Coulter. 4. Ensure that the level 4 calibrator ODR3022, Lot 3000 target value used for calibration of Apo B OSR6143 Lot 2631 is 116 mg/dL and aligns with the corrected VAS. 5. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product(s) listed above to another laboratory, please provide them a copy of this letter.

Recall details

Recall number
Z-1815-2025
Status
Open, Classified
Initiated by the company
Apr 23, 2025
Posted by the FDA
May 23, 2025
Recalling firm
Beckman Coulter Ireland, Inc.
Quantity
2747 units
Product code
JHN
FDA submissions
K952466
Distribution
Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, IL, IN, MD, MI, NJ, NY, OH, OK, OR, PA, TX, VA, WA, WV and the countries of Argentina, Australia, Belarus, Belgium, Bulgaria, Canada, China, Czechia, Egypt, Estonia, France, Germany, Ghana, Greece, Hong Kong, India, Ireland, Israel, Italy, Korea, Republic of, Malaysia, Netherlands, New Zealand, Russian Federation, Saudi Arabia, Slovakia, Spain, Switzerland, Taiwan, Province of China, United Kingdom of Great Britain and Northern Ireland, United States of America, Viet Nam.

Product description

Apo B Reagent, REF: OSR6143

Code information (models, lots, serial numbers)

Lot # 2631/UDI: 15099590010409

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.