ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E recallZ-1857-2025 · Onkos Surgical, Inc. · initiated Sep 22, 2024

Class II Open, Classified All recalls posted in May 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Onkos Surgical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Potential challenge with intraoperative assembly of the external taper of a single midsection with the internal taper of a mating proximal femur.

FDA root cause category: Nonconforming Material/Component

What the company is doing

Onkos Medical notified consignees via email on about 09/22/2024. Consignees were instructed to segregate and quarantine any product on hand immediately, complete and return the response form, and arrange for existing implants to be returned.

Recall details

Recall number
Z-1857-2025
Recall class
Class II
Status
Open, Classified
Initiated by the company
Sep 22, 2024
Posted by the FDA
May 28, 2025
Recalling firm
Onkos Surgical, Inc.
Quantity
24 units
Product code
KRO
FDA submissions
K161520
Distribution
US Nationwide distribution in the states of AZ, CA, FL, FL, MD, MO, NC, TN.

Product description

ELEOS MALE-FEMALE MIDSECTION 40MM. Part Number: 25001040E. Component of ELEOS Limb Salvage System

Code information (models, lots, serial numbers)

Part Number: 25001040E. UDI-DI B27825001040E0. Lot Numbers: 1989873, 1989874