Skip to content

Catalyst+ recallZ-1858-2025 · C-Rad Positioning AB · initiated Apr 22, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from C-Rad Positioning AB or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Software issue with scanning equipment that can results in the filed rotation not being applied correctly.

FDA root cause category: Software design

What the company is doing

On April 22, 2025, C-RAD Positioning AB issued a Field Safety Notice to affected consignees via E-Mail. C-RAD asked consignees to take the following actions: 1. As an immediate action, the affected users are recommended to: - Do not use Site Groups for treatment plans containing non-coplanar fields. - Ungroup all existing Site Groups. 2. Ensure that all the staff operating the Catalyst family systems and the c4D software application are aware of this information. 3. All customers are requested to reply to C-RAD using the attached FSN Customer Reply Form .

Recall details

Recall number
Z-1858-2025
Status
Open, Classified
Initiated by the company
Apr 22, 2025
Posted by the FDA
May 28, 2025
Recalling firm
C-Rad Positioning AB
Quantity
44 units
Product code
IYE
FDA submissions
K200435
Distribution
Worldwide - US Nationwide distribution in the states of AK, FL, WA, VA, WA, NC, IL, AR, NC and the countries of Germany, Italy, Uzbekistan, South Korea, Vietnam, Austria, Australia, China, Estonia, Belgium, Uzbekistan.

Product description

Brand Name: Catalyst+ Product Name: C4D software used in conjunction with Catalyst (SP001-0025), Catalyst HD (SP001-0026), Catalyst PT (SP001-0027), Catalyst X4 (SP002-0035), Catalyst+ (SP003-0001), Catalyst+ HD (SP003-0002), Catalyst+ PT (SP003-0008), Catalyst+ X4, (SPOO3-0009) Product Description: Patient positioning device, diagnostic imaging/radiotherapy, laser Component: Not a component

Code information (models, lots, serial numbers)

All Catalyst products using software Version: c4D 6.5.1 SP2 Catalyst -SP001-0025/UDI: 07350005280024; Catalyst HD- SP001-0026/UDI: 07350005280024; Catalyst PT- SP001-0027/UDI: 07350005280024; Catalyst X4- SP002-0035/UDI: 07350005280024; Catalyst+ - REF: SP003-0001/UDI: 17350005280762 Catalyst+ HD- SP003-0002/UDI: 17350005280779 Catalyst+ PT- SP003-0008/UDI: 17350005280793 Catalyst+ X4 - SP003-0009/UDI: 17350005280809

Devices named in this recall

5 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page