Skip to content

Adult Manual Resuscitator with Medium Adult Mask, Bag… recallZ-1814-2025 · SunMed Holdings, LLC · initiated May 1, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from SunMed Holdings, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected lots were manufactured with B/V Filter incorrectly attached to the wrong port (patient port instead of the exhalation port). If not noticed prior to patient use, there would be interruption or delay in patient resuscitation, which may lead to life threatening consequences, including hypoxia, hypercapnia, organ failure, and death.

FDA root cause category: Process control

What the company is doing

Recall initiated May 1, 2025. Customers instructed to immediately cease use of any affected product and quarantine all affected product. Customers were instructed to return or destroy affected product. AirLife will send new replacement product to customers once disposition is confirmed. If you need replacement products to be sent to you urgently, please call AirLife directly on 1-800-433-2797 and the firm will make every effort to accommodate your needs.

Recall details

Recall number
Z-1814-2025
Status
Open, Classified
Initiated by the company
May 1, 2025
Posted by the FDA
May 30, 2025
Recalling firm
SunMed Holdings, LLC
Quantity
11,358
Product code
BTM
FDA submissions
K953546
Distribution
US Nationwide distribution via Medline.

Product description

Adult Manual Resuscitator with Medium Adult Mask, Bag REservoir, Filter, Manometer and 7 ft Oxygen Tubing, REF CPRM1116F

Code information (models, lots, serial numbers)

UDI-DI: EA: 10884389164822 CS: 40884389164823 Lots: 526782 526790 526796 526797 526798 526800 526802 526804 526805 526806 526807 526808 526809 526810 526814 526815 526816 526811 526817 526818

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page