GM Helix Acqua Implant, Article Number: 140.985 recallZ-1863-2025 · Straumann USA LLC · initiated May 1, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Straumann USA LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It is possible that a package labelled as an 11.5 mm implant may contain a 13 mm implant instead.
FDA root cause category: Under Investigation by firm
What the company is doing
On May 1, 2025 URGENT FIELD SAFETY ACTION letters were sent to customers. Action to be taken: 1. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, then stop use/distribution of the product immediately and quarantine/segregate physically. 2. If the product was used and no complications or problems were identified, of the patient did not present symptoms there is no need for any additional action with the patient. 3. If article 140.984 Lot JPZ21 or 109.985 Lot JPZ75 are still in your inventory, return it for a credit or replacement as indicated on the Customer Confirmation Form. 4. In the product is not found in your stock, indicate this on the enclosed Customer Confirmation Form. 5 In all cases, complete and return the enclosed Customer Confirmation Form with the affected product using the return UPS label provided to: Straumann USA, LLC; 60 Minuteman Road, Andover, MA 01810. Transmission of the Field Safety Notice: This notice needs to be passed on to all those who need to be aware within your organization or to any organization where the affected devices have been transferred. The FDA has been informed about this Field Safety Corrective Action. We apologize for any inconvenience that this may cause.
Recall details
- Recall number
- Z-1863-2025
- Status
- Open, Classified
- Initiated by the company
- May 1, 2025
- Posted by the FDA
- May 28, 2025
- Recalling firm
- Straumann USA LLC
- Quantity
- 207 units (54 US, 153 OUS)
- Product code
- DZE
- FDA submissions
- K163194
- Distribution
- Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, Il KS, LA, MA, MD, MI, NE, NH, NJ, NY, OH, OK, RI, SC, TN, TX, UT, WI and the countries of Australia, Croatia, France, Hungary, India, Italy, Lebanon, Lithuania, New Zealand, Portugal, Qatar, Serbia, Spain, Tunisia, UAE, United Kingdom, Vietnam.
Product description
GM Helix Acqua Implant, Article Number: 140.985
Code information (models, lots, serial numbers)
Article Number: 140.985; UDI-DI: (01)7899878024569(11)220125(17)270124(10)JPZ75; Lot number: JPZ75;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Neodent 109.985JJGC Indústria e Comércio de Materiais Dentários S/A. · IMPLANT, ENDOSSEOUS, ROOT-FORM1 recall4,934 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96806).
- Z-1862-2025 · Open, ClassifiedGM Helix Acqua Implant, Article Number: 140.984
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page