Medline Kits containing Codman Disposable Perforator 14 mm recallZ-1810-2025 · MEDLINE INDUSTRIES, LP - Northfield · initiated Apr 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from MEDLINE INDUSTRIES, LP - Northfield or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Affected kits contain recalled Integra Lifesciences Codman Disposable Perforators due to inadequate welds on specific lots. Perforators with inadequate welds may break or separate.
FDA root cause category: Nonconforming Material/Component
What the company is doing
On April 22, 2025, the firm notified customers of the recall through letters titled Immediate Action Required Voluntary Medical Device Recall. Customers were instructed to identify and quarantine all affected product on hand. Customers are to submit a response form to the recalling firm. Upon receipt of the response firm, Medline will provide the customer with over-labels to place on affected inventory, which will include instructions for staff to remove and discard the affected component prior to using the kit. Distributors should notify downstream customers/accounts of this recall. If you have any questions, contact the Recall Department at 866-359-1704 or [email protected].
Recall details
- Recall number
- Z-1810-2025
- Status
- Open, Classified
- Initiated by the company
- Apr 22, 2025
- Posted by the FDA
- May 31, 2025
- Recalling firm
- MEDLINE INDUSTRIES, LP - Northfield
- Quantity
- 1853 total
- Product code
- OJG
- FDA submissions
- K213481
- Distribution
- US distribution to CA, FL, IL, IN, KY, MA, MD, MN, NY, TX, VT, WA and WI. No OUS distribution.
Product description
Medline Kits containing Codman Disposable Perforator 14 mm: CRANIOTOMY, SKU DYNJ35349C
Code information (models, lots, serial numbers)
UDI-DI 10888277858060 (EA); 40888277858061 (CASE) KIT LOTS 24CBJ925 24DBT405 24FBO318
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Medline Industries, Inc. DYNJ35349CMedline Industries, Inc. · Neurological tray4 recalls0 reports
Other devices under the same FDA submission
321 devices cleared under K213481 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Centurion ART690Centurion Medical Products Corporation · General Surgery Tray4 recalls0 reports
- Centurion CVI4505Centurion Medical Products Corporation · General Surgery Tray3 recalls0 reports
- MEDLINE ECVC6355Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE ECVC1045Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ART255Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- Centurion CVI3880Centurion Medical Products Corporation · General surgery tray3 recalls0 reports
- MEDLINE ECVC1375Medline Industries, Inc. · General surgery tray3 recalls0 reports
- MEDLINE DYNDV2520Medline Industries, Inc. · General surgery tray3 recalls0 reports
- Centurion ST11670ACenturion Medical Products Corporation2 recalls0 reports
- Medline Industries, Inc. ECVC6370Centurion Medical Products Corporation · General Surgery Tray2 recalls0 reports
- MEDLINE ECVC6840Medline Industries, Inc. · General surgery tray2 recalls0 reports
- MEDLINE ECVC6265Medline Industries, Inc. · General Surgery Tray2 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 96821).
- Z-1809-2025 · Open, ClassifiedMedline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ADD A PACK, SKU DYNJ30926O 2. CRANIOTOMY, SKU CDS983611N 3. CRANIOTOMY CDS, SKU CDS983467J 4. CRANIOTOMY PACK,…
- Z-1811-2025 · Open, ClassifiedMedline Kits containing Codman Disposable Perforator 14 mm: 1. CRANI ACCESSORY PACK SKU DYNJ59270 2. CRANI PACK-LF SKU DYNJ44805M 3. CRANIOTOMY SKU CDS982719V CDS982719W DYNJ9…
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.