PROPONENT SR SL (Model L200) recallZ-0085-2026 · Boston Scientific Corporation · initiated Aug 20, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Boston Scientific Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Software to enhance Safety Architecture is available and designed to prevent initiation of Safety Mode in an ambulatory setting due to a high battery impedance state for the ACCOLADE family of devices which includes ACCOLADE, PROPONENT, ESSENTIO, and ALTRUA 2 dual chamber (DR) standard life (SL) and DR extended life (EL) pacemakers; and VISIONIST and VALITUDE cardiac resynchronization therapy pacemakers (CRT-Ps).
FDA root cause category: Other
What the company is doing
Boston Scientific issued an "Urgent Medical Device Correction" notice to its customers on 08/18/2025 via letter. The notice provides an update to the December 2024 customer communication regarding the potential for the ACCOLADE family of pacemakers to initiate Safety Mode due to a high battery impedance state and contains recommendations timely upgrade of pacemaker software to prevent the occurrence of Safety Mode in an ambulatory setting. On 09/05/2025 an update to the 08/18/2025 notice (via letter), superseding the prior notice to address two unintended behaviors associated with the Model 3869 v2.04 software (designed to prevent potential initiation of Safety Mode due to a high battery impedance condition). The behaviors are: 1. Incomplete disablement of wandless automatic telemetry after detecting a high battery impedance condition. If an ACCOLADE family device with an elevated battery impedance induced Code-1003 remains in service, the software s new telemetry activated battery test is designed to continue assessing battery impedance and disable all wandless telemetry operation before high impedance is achieved and device resets can initiate Safety Mode. 2. Potential for the software s new daily battery test to misinterpret measurements in the presence of a magnet, initiating a false positive response that results in disablement of wandless telemetry sessions. For patients monitored via LATITUDE, this will trigger an alert message indicating an erroneous explant indicator (e.g., explant indicator was reached January 1, 2000) if the device had not previously reached replacement indication, stating that LATITUDE is no longer available, and prompting to contact Technical Services. Patients were notified regarding battery performance. Advisory is published at www.BostonScientific.com/advisory. If you have any questions, contact US Technical Services at 1.800.CARDIAC (227.3422) or email to: [email protected].
Recall details
- Recall number
- Z-0085-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 20, 2025
- Posted by the FDA
- Oct 10, 2025
- Recalling firm
- Boston Scientific Corporation
- Quantity
- 1380 units
- Product code
- LWP
- Distribution
- Worldwide
Product description
PROPONENT SR SL (Model L200)
Code information (models, lots, serial numbers)
Show the full code information (11k characters)
GTIN 00802526559020, Serial numbers: 722605, 722614, 722634, 722647. GTIN 00802526576362, Serial numbers: 464359, 464382, 464387, 464444, 464447, 464457, 464462, 464471, 464505, 464511, 464514, 464517, 464521, 464525, 464527, 464541, 464549, 464552, 464554, 464561, 464576, 708351, 708451, 708466, 708471, 708495, 708513, 708514, 708515, 708516, 708517, 708525, 708529, 708532, 708533, 708536, 708618, 708619, 708620, 708622, 708623, 708624, 708625, 708627, 708630, 708651, 708653, 708654, 708655, 708656, 708658, 708660, 708737, 708751, 708764, 708783, 708816, 708824, 708826, 708857, 708862, 708867, 708868, 708879, 708880, 708885, 708887, 708889, 708894, 708898, 708912, 708920, 708922, 708924, 708929, 708939, 708942, 708944, 708945, 708946, 708947, 708948, 708949, 708952, 708959, 708960, 708967, 708968, 708969, 708976, 708980, 708983, 708997, 709003, 709009, 709011, 709015, 709024, 709025, 709037, 709040, 709041, 710248, 710517, 710627, 710705, 710754, 710765, 710771, 710772, 710788, 710789, 710790, 710824, 710825, 710831, 710889, 710905, 710907, 710910, 710916, 710931, 710954, 710980, 711315, 711412, 711640, 711655, 714062, 714644, 714688, 714708, 715328, 717903, 717931, 717934, 717941, 717948, 717950, 717951, 717965, 717976, 717981, 717982, 717992, 718007, 718009, 718138, 718522, 718780, 718783, 718937, 720309, 720504, 720556, 720597, 720650, 721983, 721991, 722012, 722438, 722565, 722689, 722724, 722756, 722967, 723456, 724170, 724319, 724398, 724404, 724485, 724623, 725175, 725182, 725198, 725262, 725272, 725283, 725363, 725365, 725368, 725370, 725371, 725372, 725373, 725383, 725388, 725389, 725390, 725392, 725404, 725409, 725418, 725419, 725425, 725444, 727098, 728855, 728893, 728894, 728917, 728922, 729054, 729410, 729411, 729413, 729421, 729589, 729599, 729602, 731110, 731612, 734834. GTIN 00802526578007, Serial Numbers: 464370, 464388, 464446, 464452, 464458, 464460, 464467, 464468, 464470, 464475, 464477, 464481, 464482, 464483, 464486, 464490, 464491, 464492, 464498, 464500, 464501, 464506, 464508, 464512, 464516, 464519, 464523, 464526, 464528, 464532, 464563, 464568, 464570, 464590, 705011, 705236, 705383, 705409, 705419, 705462, 705468, 705555, 705576, 705584, 705622, 705628, 705702, 705706, 705711, 705713, 705728, 705751, 705753, 705795, 705812, 706007, 706098, 706215, 706321, 706350, 706372, 706698, 706846, 706879, 706918, 706923, 706931, 706947, 706967, 707053, 707069, 707089, 707138, 707173, 708261, 708473, 708494, 708504, 708528, 708534, 708546, 708549, 708585, 708589, 708595, 708629, 708652, 708657, 708721, 708722, 708723, 708749, 708750, 708752, 708770, 708774, 708775, 708776, 708780, 708797, 708798, 708802, 708815, 708819, 708820, 708831, 708832, 708841, 708843, 708844, 708848, 708851, 708853, 708858, 708859, 708874, 708877, 708883, 708886, 708892, 708896, 708905, 708907, 708910, 708913, 708914, 708915, 708916, 708917, 708918, 708923, 708925, 708928, 708930, 708932, 708933, 708935, 708936, 708937, 708938, 708940, 708941, 708943, 708950, 708956, 708958, 708965, 708966, 708971, 708974, 708975, 708977, 708978, 708981, 708982, 708984, 708986, 708987, 708992, 709000, 709002, 709004, 709006, 709007, 709008, 709012, 709013, 709014, 709016, 709023, 709027, 709029, 709033, 709034, 709039, 709042, 709043, 709044, 709045, 710352, 710458, 710499, 710501, 710532, 710562, 710568, 710569, 710577, 710667, 710675, 710676, 710727, 710729, 710735, 710739, 710766, 710769, 710796, 710801, 710802, 710808, 710811, 710820, 710895, 710897, 710903, 710921, 710957, 710982, 711020, 711082, 711102, 711225, 711231, 711235, 711286, 711314, 711328, 711345, 711353, 711354, 711356, 711455, 711473, 711530, 711534, 711546, 711552, 711562, 711569, 711577, 711594, 711596, 711598, 711599, 711604, 711611, 711614, 711616, 711617, 711619, 711641, 711763, 711765, 711768, 711776, 711777, 711780, 711791, 711795, 711804, 711814, 711823, 711839, 711852, 711863, 711865, 711866, 711868, 711870, 711872, 711875, 711887, 711890, 711893, 711894, 711899, 711910, 711914, 711915, 711917, 711924, 711925, 711926, 711928, 711931, 711938, 711942, 711946, 711947, 711954, 711963, 711970, 711971, 711980, 711982, 711987, 711997, 712009, 712017, 712026, 712053, 712056, 712058, 712084, 712086, 712088, 712093, 712098, 712107, 712125, 712129, 712131, 712137, 712145, 712174, 712177, 712178, 712210, 712219, 712232, 712239, 712249, 712255, 712256, 712257, 712264, 712279, 712284, 712286, 712290, 712302, 712312, 712321, 712350, 712352, 712385, 712390, 712396, 712397, 712410, 712414, 712428, 712436, 712442, 712449, 712459, 712470, 712475, 712477, 712489, 712521, 712534, 712537, 712571, 712598, 712631, 712716, 712742, 712750, 712751, 712820, 712829, 712843, 712848, 712860, 712862, 712864, 712868, 712875, 712882, 713006, 713041, 713048, 713052, 713053, 713054, 713055, 713056, 713058, 713060, 713062, 713063, 713073, 713074, 713083, 713125, 713126, 714102, 714105, 714164, 714206, 714221, 714223, 714226, 714246, 714260, 714261, 714307, 714338, 714345, 714358, 714360, 714361, 714374, 714443, 714518, 714546, 714550, 714599, 714622, 714631, 714638, 714646, 714647, 714651, 714653, 714657, 714660, 714683, 714690, 714692, 714694, 714695, 714699, 714707, 714709, 714710, 714719, 714744, 714748, 714754, 714761, 714785, 714791, 714827, 715038, 715040, 715041, 715042, 715058, 715106, 715124, 715205, 715237, 715241, 715243, 715258, 715260, 715273, 715278, 715279, 715284, 715285, 715287, 715289, 715293, 715297, 715300, 715304, 715317, 715325, 715327, 715332, 715333, 716079, 716137, 716179, 716191, 716196, 716203, 716204, 716306, 716353, 716354, 716356, 716370, 716385, 716411, 716420, 716449, 716861, 716864, 716915, 716926, 716927, 716930, 716931, 716933, 716935, 716937, 716938, 716940, 716944, 716947, 716948, 716949, 716950, 716951, 716964, 716965, 716966, 716974, 716978, 716986, 716987, 716989, 716991, 717163, 717181, 717194, 717197, 717199, 717200, 717201, 717202, 717203, 717204, 717205, 717339, 717356, 717357, 717358, 717359, 717360, 717361, 717369, 717374, 717377, 717383, 717385, 717392, 717393, 717395, 717401, 717402, 717423, 717514, 717624, 717755, 717784, 717786, 717795, 717796, 717797, 717798, 717799, 717800, 717805, 717918, 717925, 717930, 717936, 717949, 717953, 717954, 717955, 717968, 717972, 717983, 718008, 718014, 718054, 718060, 718061, 718067, 718072, 718074, 718079, 718080, 718085, 718086, 718091, 718098, 718109, 718113, 718117, 718120, 718128, 718134, 718136, 718169, 718204, 718205, 718217, 718224, 718395, 718420, 718446, 718447, 718448, 718449, 718450, 718467, 718697, 718711, 718719, 718728, 718769, 718774, 718775, 718782, 718784, 718795, 718921, 718923, 718932, 718991, 719000, 719003, 719007, 719009, 719034, 719040, 719041, 719080, 719097, 719099, 719156, 719160, 719182, 719186, 719196, 719198, 719211, 719230, 719236, 719237, 719239, 719248, 719252, 719253, 719255, 719261, 719265, 719270, 719271, 719301, 719357, 719367, 719372, 719373, 719376, 719379, 719421, 719453, 719456, 719460, 719730, 719752, 719753, 719754, 719755, 719756, 719759, 719761, 719894, 719925, 719929, 719943, 719969, 719972, 720012, 720019, 720109, 720110, 720119, 720122, 720124, 720132, 720134, 720135, 720141, 720142, 720144, 720145, 720147, 720150, 720154, 720172, 720173, 720183, 720184, 720191, 720198, 720228, 720234, 720240, 720242, 720250, 720502, 720503, 720538, 720541, 720593, 720600, 720604, 720611, 720618, 720633, 720641, 720660, 720663, 720697, 720729, 720767, 720782, 720792, 720793, 720801, 720832, 720858, 720859, 720874, 720920, 720931, 720943, 720961, 720981, 721004, 721204, 721205, 721213, 721219, 721228, 721230, 721276, 721812, 721872, 721884, 721977, 722030, 722046, 722047, 722178, 722180, 722181, 722333, 722448, 722479, 722539, 722558, 722569, 722570, 722578, 722580, 722612, 722639, 722655, 722668, 722671, 722703, 722705, 722808, 722822, 722843, 722844, 722905, 722906, 722910, 722917, 722921, 722930, 722933, 722944, 722945, 722947, 722948, 722949, 722957, 722958, 722962, 722963, 722964, 722965, 722968, 722969, 722970, 723010, 723298, 723305, 723317, 723580, 723713, 723745, 723749, 723751, 723755, 723766, 723772, 723775, 723777, 723779, 723782, 723784, 723794, 723802, 723830, 724358, 724397, 724400, 724436, 724443, 724448, 724456, 724471, 724496, 724500, 724739, 725060, 725063, 725073, 725074, 725077, 725078, 725083, 725086, 725088, 725100, 725102, 725108, 725112, 725115, 725121, 725132, 725156, 725165, 725179, 725193, 725200, 725202, 725247, 725266, 725270, 725273, 725274, 725275, 725276, 725277, 725279, 725281, 725285, 725288, 725289, 725290, 725291, 725296, 725298, 725300, 725302, 725304, 725305, 725306, 725307, 725308, 725309, 725312, 725322, 725339, 725367, 725386, 725407, 725449, 725459, 725460, 725484, 725659, 725697, 725699, 725703, 725704, 725706, 725710, 725742, 725747, 725750, 725753, 725755, 725756, 725757, 725758, 725759, 725760, 725761, 725765, 725766, 725769, 725772, 725773, 725774, 725776, 725786, 725792, 725888, 725889, 725906, 725909, 725911, 725915, 725917, 725922, 725923, 725957, 726031, 726034, 726061, 726089, 726117, 726118, 726119, 726121, 726125, 726144, 726156, 726160, 726166, 726168, 726169, 726171, 726173, 726175, 726177, 726178, 726179, 726184, 726185, 726187, 726193, 726197, 726199, 726200, 726205, 726209, 726220, 726227, 726228, 726358, 726576, 726591, 726596, 726600, 726619, 726625, 726682, 726772, 726943, 726945, 727060, 727062, 727063, 727064, 727065, 727087, 727092, 727096, 727139, 727140, 727160, 727442, 727559, 727575, 727590, 727595, 727678, 727679, 727729, 727739, 727877, 727879, 727885, 727892, 727906, 727935, 727941, 728077, 728110, 728120, 728121, 728123, 728124, 728125, 728136, 728139, 728357, 728371, 728438, 728445, 728447, 728449, 728454, 728465, 728467, 728468, 728469, 728473, 728474, 728493, 728500, 728505, 728516, 728529, 728530, 728534, 728537, 728557, 728558, 728634, 728666, 728677, 728846, 728849, 728850, 728851, 728854, 728919, 728926, 728944, 729047, 729050, 729057, 729059, 729060, 729062, 729065, 729134, 729136, 729138, 729141, 729155, 729157, 729166, 729170, 729187, 729206, 729211, 729414, 729417, 729436, 729438, 729451, 729456, 729513, 729524, 729568, 729573, 729575, 729576, 729632, 729634, 729635, 729637, 729640, 729655, 729656, 729659, 729661, 729685, 730227, 730349, 731139, 731159, 731183, 731184, 731189, 731208, 731210, 731217, 731224, 731228, 731239, 731241, 731249, 731296, 731331, 731332, 731337, 731342, 731343, 731351, 731364, 731368, 731370, 731408, 731415, 731423, 731439, 731452, 731453, 731455, 731462, 731466, 731468, 731471, 731634, 731669, 731773, 731781, 731800, 731815, 731824, 731830, 731878, 731886, 731888, 731932, 731967, 731970, 731972, 731974, 731980, 731986, 731992, 731999, 732280, 732321, 732430, 732436, 732682, 732694, 732749, 732898, 732903, 732913, 733412, 733461, 733496, 733499, 733501, 733533, 733535, 733536, 733537, 733539, 733540, 733926, 733954, 733961, 734000, 734423, 734499, 734508, 734514, 734535, 734563, 734575, 734676, 734682, 734729, 734842, 734886, 734979, 734986, 734989, 734997, 735022, 735024, 735032, 735037, 735038, 735047, 735051, 735055, 735061, 735062, 735063, 735066, 735070, 735075, 735078, 735084, 735094, 735604.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- PROPONENT™ SR L200Boston Scientific Corporation · Implantable pulse generator, pacemaker (non-CRT)1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97467).
- Z-0079-2026 · Open, ClassifiedACCOLADE SR SL (Model L300)
- Z-0080-2026 · Open, ClassifiedACCOLADE DR SL (Model L301)
- Z-0081-2026 · Open, ClassifiedACCOLADE SR SL MRI (Model L310)
- Z-0082-2026 · Open, ClassifiedACCOLADE DR SL MRI (Model L311)
- Z-0083-2026 · Open, ClassifiedACCOLADE DR EL (Model L321)
- Z-0084-2026 · Open, ClassifiedACCOLADE DR EL MRI (Model L331)
- Z-0086-2026 · Open, ClassifiedPROPONENT DR SL (Model L201)
- Z-0088-2026 · Open, ClassifiedModel Number L210 PROPONENT SR SL Pacemaker
- Z-0089-2026 · Open, ClassifiedModel Number L211 PROPONENT DR SL MRI Pacemaker
- Z-0091-2026 · Open, ClassifiedModel Number L231, PROPONENT DR EL MRI Pacemaker
- Z-0092-2026 · Open, ClassifiedModel Number L100, ESSENTIO SR SL Pacemaker
- Z-0093-2026 · Open, ClassifiedModel Number L101, ESSENTIO DR SL Pacemaker
- Z-0094-2026 · Open, ClassifiedModel Number L110, ESSENTIO SR SL MRI Pacemaker
- Z-0095-2026 · Open, ClassifiedModel Number L111, ESSENTIO DR SL MRI Pacemaker
- Z-0096-2026 · Open, ClassifiedModel Number L121 ESSENTIO DR EL Pacemaker
- Z-0097-2026 · Open, ClassifiedModel Number L131, ESSENTIO DR EL MRI Pacemaker
- Z-0101-2026 · Open, ClassifiedModel Number U125, VALITUDE CRT-P EL
- Z-0102-2026 · Open, ClassifiedModel Number U128, VALITUDE CRT-P EL MRI
- Z-0103-2026 · Open, ClassifiedModel Number U225 VISIONIST CRT-P EL
- Z-0104-2026 · Open, ClassifiedModel Number U226, VISIONIST CRT-P EL
- Z-0105-2026 · Open, ClassifiedModel Number U228, VISIONIST CRT-P EL MRI
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page