Brand Name: Vacuette recallZ-0571-2026 · Greiner Bio-One North America, Inc. · initiated Aug 19, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Greiner Bio-One North America, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Some of the Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.
FDA root cause category: Labeling Change Control
What the company is doing
On 08/22/2025, the firm sent via email an "Urgent Medical Device Correction" letter to customer informing them that some Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator which may affect the stability of the sample. Customers are instructed to: " Stop use immediately: Discontinue use of the affected product (concerned Item/Lot) and isolate it within your facility. " Notify staff and customers: Inform all relevant employees and notify all customers to whom you distributed the affected products. " Assess risk: Evaluate potential risks or consequences in accordance with your internal procedures and take appropriate action. " Confirm product disposition: Complete the attached Product Disposition Site Confirmation Form and return it via fax to 877-217-7927 or email to [email protected]. Confirmation is required that all affected products have been discarded/destroyed. Replacement products will be provided upon receipt of your completed form. For questions, call Greiner Bio-One Technical Service at 800-515-8112.
Recall details
- Recall number
- Z-0571-2026
- Status
- Open, Classified
- Initiated by the company
- Aug 19, 2025
- Posted by the FDA
- Nov 19, 2025
- Recalling firm
- Greiner Bio-One North America, Inc.
- Quantity
- 1,980,000
- Product code
- JKA
- FDA submissions
- K081929, K983952
- Distribution
- U.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, VA and WY. O.U.S.: None
Product description
Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)
Code information (models, lots, serial numbers)
Lot Code: Lot/Serial Numbers: Item# 455071P Lot# B250437N UDI-DI code: 29120017573061
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vacuette 455071PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection2 recalls0 reports
Other devices under the same FDA submission
2 devices cleared under K081929, K983952 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vacuette 454067PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection1 recall0 reports
- Vacuette 456073PGreiner Bio-One GmbH · Tubes, vials, systems, serum separators, blood collection1 report
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page