Why it was recalled

Some of the Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator, which could yield erroneous assay results and require the patient to return to the lab for recollection or repeat testing. This may result in delayed diagnosis, cause a delay in clinical decision-making or treatment and in some rare cases, yield incorrect results from less stable parameters.

FDA root cause category: Labeling Change Control

What the company is doing

On 08/22/2025, the firm sent via email an "Urgent Medical Device Correction" letter to customer informing them that some Vacuette CAT Serum Separator Clot Activator tubes were manufactured without gel separator which may affect the stability of the sample. Customers are instructed to: " Stop use immediately: Discontinue use of the affected product (concerned Item/Lot) and isolate it within your facility. " Notify staff and customers: Inform all relevant employees and notify all customers to whom you distributed the affected products. " Assess risk: Evaluate potential risks or consequences in accordance with your internal procedures and take appropriate action. " Confirm product disposition: Complete the attached Product Disposition Site Confirmation Form and return it via fax to 877-217-7927 or email to [email protected]. Confirmation is required that all affected products have been discarded/destroyed. Replacement products will be provided upon receipt of your completed form. For questions, call Greiner Bio-One Technical Service at 800-515-8112.

Recall details

Recall number
Z-0571-2026
Status
Open, Classified
Initiated by the company
Aug 19, 2025
Posted by the FDA
Nov 19, 2025
Recalling firm
Greiner Bio-One North America, Inc.
Quantity
1,980,000
Product code
JKA
FDA submissions
K081929, K983952
Distribution
U.S.: AL, AR, CA, CO, CT, FL, GA, IA, IL, IN, KS, LA, MA, MD, ME, MI, MO, MS, NJ, NY, OH, OK, PA, TN, TX, UT, VA and WY. O.U.S.: None

Product description

Brand Name: Vacuette Product Name: 8 ml CAT Serum Separator Clot Activator Model/Catalog Number: 455071P Product Description: Greiner VACUETTE Blood Collection Tube with Clot Activator and Gel Separator and Standard Stopper (16 x 100 mm, 8ml)

Code information (models, lots, serial numbers)

Lot Code: Lot/Serial Numbers: Item# 455071P Lot# B250437N UDI-DI code: 29120017573061

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

2 devices cleared under K081929, K983952 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.