Skip to content

ClariTEE Miniaturized TEE probe. Model Number: CLT-010 recallZ-0038-2026 · ImaCor Inc. · initiated Aug 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ImaCor Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Production assembly error resulting in the potential for reversed articulation, where moving the lever forward caused the probe tip to anteflex rather than retroflex, and vice versa.

FDA root cause category: Process control

What the company is doing

ImaCor, Inc. notified consignees on 08/29/2025 via email. Consignees were instructed to Immediately discontinue use of the affected ClariTEE probes, quarantine all affected units in inventory, confirm receipt of this notice, return all affected units to ImaCor, and complete and return the response form.

Recall details

Recall number
Z-0038-2026
Status
Open, Classified
Initiated by the company
Aug 29, 2025
Posted by the FDA
Oct 3, 2025
Recalling firm
ImaCor Inc.
Quantity
45 units
Product code
ITX
FDA submissions
K220490
Distribution
US Nationwide distribution in the states of South Carolina, New Jersey, Pennsylvania, Florida, and Georgia.

Product description

ClariTEE Miniaturized TEE probe. Model Number: CLT-010. Ultrasound miniaturized TEE probe for use with Mobile Ultrasound, ImaCor model ZHH.

Code information (models, lots, serial numbers)

Model CLT-010; UDI-DI 00861589000108; Lot number: 20261209.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page