Sharesource Connectivity Platform for Use with Homechoice… recallZ-0409-2026 · Vantive US Healthcare LLC · initiated Aug 28, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Vantive US Healthcare LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Vantive has identified a software defect within the Sharesource Claria software, which may cause a patient s updated prescription program settings to not be properly saved on the actual cycler that the patient is using to perform their therapy. As a result, the clinician may think that the patient is receiving the correct prescribed therapy because of what is erroneously displayed on the clinician's remote Sharesource application, when in fact the cycler that is with the patient is actually delivering an outdated and incorrect therapy.

FDA root cause category: Software design

What the company is doing

Clinics whose patients' programs were identified as being impacted by this issue were contacted by phone beginning 8/29/2025. Follow-up written notifications were sent out to all affected customers starting September 11, 2025. The Sharesource Claria software fix has been deployed and no further actions need to be taken by facilities who did not have directly affected patients. Note: If you have received the previous direct notification regarding a patient at your clinic, please note that Vantive will continue to be in touch with you regarding the next steps to correct those specific patient programs.

Recall details

Recall number
Z-0409-2026
Status
Open, Classified
Initiated by the company
Aug 28, 2025
Posted by the FDA
Oct 29, 2025
Recalling firm
Vantive US Healthcare LLC
Quantity
1
Product code
FKX
FDA submissions
K201867
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Hong Kong, Korea, New Zealand, Taiwan, Thailand, France, Germany, Ireland, Netherlands, Italy, Spain, Turkey, United Kingdom, Chile, and Colombia.

Product description

Sharesource Connectivity Platform for Use with Homechoice Claria Product Code 5CGM01 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Code information (models, lots, serial numbers)

UDI-DI 15413765545723 software versions 7.13.3 (China), 8.13.2 (APAC and Korea), and 8.13.4 (US, EMEA, and LATAM)

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

1 device cleared under K201867 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page