BD Pyxis recallZ-1028-2026 · CareFusion 303, Inc. · initiated Dec 3, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from CareFusion 303, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due a software issue that may result in equipment not receiving timely updates of patient, medication, or related information from enterprise systems.
FDA root cause category: Under Investigation by firm
What the company is doing
On 12/03/2025, the firm sent U.S. Postal and email an "URGENT: MEDICAL DEVICE CORRECTION" Letter to customers informing them that BD received customer complaints of BD Pyxis" equipment not receiving timely updates of patient, medication, or related information from enterprise systems. This resulted in new patient and/or order information to be delayed from displaying on connected Pyxis equipment, and a delay in any changes to current patient orders being applied to the patient profile within Pyxis. Customers are instructed to: 1.Refer to the BD Pyxis" ES User Guides warning: Instruct end users to utilize the BD Pyxis" ES System patient/order information in conjunction with the medical record when making clinical decisions. Making clinical decisions solely from BD Pyxis" ES System information may result in errors in care resulting in serious injury or death. 2.BD Pyxis" MedStation" devices can be utilized in both profile mode (specific medications orders are associated with specific patients) and non-profiled mode. 3.Facilities should establish comprehensive policies and procedures for situations when an automated dispensing system is not receiving updated information and medications may not be loaded in device inventory. 4.Device keys can be used to manually access unavailable drawers. Keys must be readily available and retrievable to constitute risk mitigation. 5.Devices can also be placed into critical override mode, where all contents of the device will be available to users. (For additional detailed instruction - refer to Urgent Medical Device Correction letter. BD recommends that customers who are preparing to implement or upgrade to ES 1.11 continue as planned in order to receive the bug fixes it contains and prepare to receive the ES 1.11.1 patch when it is available. For Questions: BD Remediation Support: Email: [email protected] BD Technical Support: Website: bd.com/self-service, Phone: 1-800-727-6102, Hours: 24 hours, 7 days a week
Recall details
- Recall number
- Z-1028-2026
- Status
- Open, Classified
- Initiated by the company
- Dec 3, 2025
- Posted by the FDA
- Jan 8, 2026
- Recalling firm
- CareFusion 303, Inc.
- Quantity
- 2,079 systems
- Product code
- BRY
- Distribution
- Worldwide - U.S. Nationwide distribution including in the states of AK,AL,AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, and WY. The countries of Bermuda, and Canada.
Product description
BD Pyxis ES Enterprise Server Catalog Numbers 1115-00, 137672-02,137672-03 137671-02, 137671-03, 137670-02, 137670-03, 139676-01, 139421-01, 139420-01, 139419-01, 139419-02, 139673-01, 137081-01, 136451-01, 136451-02, 136450-01, 136450-02, 136449-01, 136449-02, 134148-01, 133760-01, 134149-01, 137669-02, 137669-03, 139675-01, 139672-01 139418-01, 139418-02, 136448-01, 136448-02, 137673-02, 137673-03 139422-01, 133744-01, 136452-01, 136452-02, 138551-01, 134157-01 Software Versions: System Versions 1.6.1 - 1.11.0 Server Application Versions ES 5.4X - 5.8.X The BD Pyxis Enterprise Server is a web application intended to configure and manage BD Pyxis ES automated dispensing cabinets (ADC) and associated BD PyxisTM ES products.
Code information (models, lots, serial numbers)
Catalog Number: 1115-00 Software Versions: UDI-DI code for Hardware: 10885403518348 UDI-DI code for Software: 10885403520341 System Versions 1.6.1 - 1.11.0 Server Application Versions ES 4.13 - 5.8.X
Devices named in this recall
2 devices in the FDA device registry whose identifier appears in the recall's code information.
- BD Pyxis™ SW-SVR-ES-55Carefusion 303, Inc. · Cabinet, table and tray, anesthesia3 recalls460 reports
- BD Pyxis™ SW-SVR-ES-57Carefusion 303, Inc. · Cabinet, table and tray, anesthesia3 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.