Change Healthcare Radiology Solutions recallZ-0030-2026 · Change Healthcare Canada Company · initiated Aug 13, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Change Healthcare Canada Company or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to software issue, radiology reports may not be fully displayed when viewing.

FDA root cause category: Software Design Change

What the company is doing

On August 13, 2025, Change Healthcare Canada Company issued a Urgent Medical Device Recall notification to affected consignees via Email. Chang Healthcare asked consignees to take the following actions: 1. Generate an audit report to list all studies accessed since CHRS 14.2.2 client was first installed. 2. Re-evaluate the identified cases and take any necessary follow-up measures to confirm that each patient s diagnosis and management remain appropriate. 4. This notice must be distributed to all personnel within your organization who need to be aware 5. To ensure effectiveness of any required corrective actions, please maintain awareness of this RECALL until the issue has been fully resolved. 6. Please notify our Customer Support department at: 1-800-663-2533 that you have read and understood the recall notice and schedule your product update to be installed on your system.

Recall details

Recall number
Z-0030-2026
Status
Open, Classified
Initiated by the company
Aug 13, 2025
Posted by the FDA
Oct 2, 2025
Recalling firm
Change Healthcare Canada Company
Quantity
12 programs
Product code
LLZ
FDA submissions
K140909
Distribution
Worldwide - US Nationwide distribution in the states of AR, WI, NJ, IN, AL, CA, WA, FL, MA, LA and the country of Canada.

Product description

Change Healthcare Radiology Solutions software version 14.2.2

Code information (models, lots, serial numbers)

Version 14.2.2/UDI: (01)17540262030068

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K140909 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.