SAFE-T-FILL Micro Capillary Blood Collection, 125 L… recallZ-0372-2026 · ASP Global, LLC. dba Anatomy Supply Partners, LLC. · initiated May 5, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from ASP Global, LLC. dba Anatomy Supply Partners, LLC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

All RAM SAFE-T-FILL Micro Capillary Blood Collection tubes manufactured between July 31, 2023 to February 28, 2025 cause false positive results when used with Magellan Diagnostics LeadCare Testing Systems. False positive lead results may cause delayed results and additional unnecessary testing.

FDA root cause category: Component design/selection

What the company is doing

Starting on 5//5/2025 customers were informed that there had been reports of falsely elevated blood lead levels associated with the use of affected micro capillary tubes for lead testing when used with Magellan LeadCare Testing Systems. On 9/8/2025, recall notices were emailed to customers who were asked to do the following: Discontinue use of the recalled products immediately. b) Quarantine any remaining inventory to prevent further use. c) Dispose remaining inventory in accordance with local regulations or notify firm at [email protected] for return authorization and disposal assistance. d) Complete and return the enclosed Recall Response Form to confirm receipt and actions to [email protected] Distributors are asked to notify their customers of the recall or to forward their customer information to the recalling firm so they might notify these customers.

Recall details

Recall number
Z-0372-2026
Status
Open, Classified
Initiated by the company
May 5, 2025
Posted by the FDA
Oct 27, 2025
Recalling firm
ASP Global, LLC. dba Anatomy Supply Partners, LLC.
Quantity
4,272,858
Product code
JKA
FDA submissions
K981973
Distribution
Worldwide - US Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, PA, SC, SD, TN, TX, UT, VA, WA, WI and the country of CA.

Product description

SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Purple, Model: 07 6011; SAFE-T-FILL Micro Capillary Blood Collection, 125 L Prepared with Dipotassium EDTA Self-sealing Cap, Purple, Model: 07 6013; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7051; SAFE-T-FILL Micro Capillary Blood Collection, 150 L Prepared with Dipotassium EDTA Purple, Model: 07 7052; SAFE-T-FILL Micro Capillary Blood Collection, 300 L Prepared with Dipotassium EDTA Purple, Model: 07 7053; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Purple, Model: 07 7056; SAFE-T-FILL Micro Capillary Blood Collection, 200 L Prepared with Dipotassium EDTA Flat Bottom Microtube, Purple, Model: 07 7058 SAFE-T-FILL End-to-End Capillary Tubes, 30 L; Plastic, Prepared with EDTA Dipotassium Model: 06 0910

Code information (models, lots, serial numbers)

Show the full code information (1k characters)

Model/UDI-DI/Lots: 07 6011/10643351000014/23H4110, 23I4017, 23I4051, 23J4031, 23K4058, 23K4059, 23L4082, 24A4103, 24B4186, 24B4187, 24D4015, 24E4138, 24E4139, 24H4121, 24I4317, 24I4328, 24J4041, 24J4137, 24J4138, 24J4233, 24J4316, 24J4348; 07 6013/10643351000021/23H4111, 23I4018, 24B4188, 24H4125, 25B4099; 07 7051/10643351000038/23H4157, 23I4020, 23I4374, 23J4033, 23K4060, 23L4083, 24A4104, 24B4190, 24D4016, 24E4140, 24F4053, 24I4321, 24J4043, 24J4139, 24J4237, 24L4001, 25B4107; 07 7052/10643351000045/23H4113, 23I4021, 23J4034, 24A4105, 24D4017, 24F4052, 24H4126, 24I4323, 24J4042, 24J4140, 24J4338, 24J4355, 25B4108; 07 7053/10643351000052/23I4022, 23J4035, 23L4084, 24D4018, 24E4132, 24F4192, 25B4106; 07 7056/10643351000069/24A4106, 24B4191, 24B4192, 24E4133, 24I4315, 24J4136, 24J4347; 07 7058/10643351000076/23G4430, 23G4168. Manufactured between July 31, 2023 and February 28, 2025.