FlexLab (FLX) System. Potassium Test System. in vitro… recallZ-0106-2026 · Inpeco S.A. · initiated Sep 3, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Inpeco S.A. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The FlexLab System Sample Integrity Module (SIM) includes function that determines the hemolysis, icterus, and lipemia (HIL) in blood samples. This function has not been evaluated for safety and effectiveness by the FDA, so its use may cause an increase in erroneous results or a delay in patient results.

FDA root cause category: No Marketing Application

What the company is doing

Consignees received an Urgent Medical Device Recall notification via email dated 9/3/25. The notification instructs consignees to not rely on HIL functionality for clinical decision making or to flag potentially erroneous samples. HIL functionality must be discontinued and cannot be disabled by users. Inspeco must send an Inspeco Servicer to consignee locations to disable HIL functionality. Inspeco will contact consingnees to schedule this service. Consignees are to forward the provided notification to those whom it might concern. Consignees with questions can contact Regulatory Affairs via email at [email protected] or Inspeco Service Team via email at [email protected].

Recall details

Recall number
Z-0106-2026
Status
Open, Classified
Initiated by the company
Sep 3, 2025
Posted by the FDA
Oct 9, 2025
Recalling firm
Inpeco S.A.
Quantity
8 units
Product code
CEM
FDA submissions
K121012
Distribution
US distribution to CA & NY.

Product description

FlexLab (FLX) System. Potassium Test System. in vitro diagnostic

Code information (models, lots, serial numbers)

PN: FLX-056-01, FLX-056-11; UDI: 07640172340004: Serial No. FLX.0204, FLX.0205, FLX.0209, FLX.0210, FLX.0216, FLX.0220, FLX.0221, FLX.0222.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K121012 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97565).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page