FDR Visionary Suite recallZ-0044-2026 · FUJIFILM Healthcare Americas Corporation · initiated Sep 15, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from FUJIFILM Healthcare Americas Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
It was found that the shaft for mounting the X-ray tube unit on the support may break. In addition, it was found that the protective parts might not work effectively only when the shaft was broken while being operated toward the front while lifting.
FDA root cause category: Device Design
What the company is doing
On September, 15, 2025 URGENT: MEDICAL DEVICE RECALL letters were sent to customers. Actions to be taken: Perform the daily inspection described in the operation manual to check for wobbles or shakes, misalignment of the light irradiation position, etc. If any abnormality is found, discontinue use and contact Technical Assistance Center (TAC). When rotating the X-ray tube unit, do not hit the stopper at the end. Do not rotate the X-ray tube over or near the patient. Please post this notice on or near the device. Complete and return the response form provided. Fujifilm Service Engineers will install countermeasure parts that will prevent the X-ray tube support from coming off even if the shaft is damaged. Please contact Technical Assistance Center (TAC) at 1-888-385-4633 or 1-888-FUJIMED for any functional/correction questions as well as to schedule your correction.
Recall details
- Recall number
- Z-0044-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 15, 2025
- Posted by the FDA
- Oct 7, 2025
- Recalling firm
- FUJIFILM Healthcare Americas Corporation
- Quantity
- 16 units
- Product code
- KPR
- FDA submissions
- K152294
- Distribution
- Domestic: AK, CA, FL, IL, MA, NH, OH, WI.
Product description
FDR Visionary Suite, CH-200 Model/Version Numbers: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33
Code information (models, lots, serial numbers)
Model/Version Number: (1) 566-16130-23, (2) 566-16130-31, (3) 566-16130-33; UDI-DI: (1) 04540217052226, (2) 04540217057436, (3) 04540217057450; Serial Numbers: (1) MP95A8F5A001, MP95A9F6A001, MP95A9F6B001, MP95A9F74001, MP95A9F74002, MP95A9F76001, MP95A9F78001, (2) MP95AA783001, (3) MP95AA983001, MP95AA983002, MP95AA983003, MP95AA985001, MP95AA985003, MP95AA985002, MP95AA988001, MP95AA988002;
Devices named in this recall
3 devices in the FDA device registry whose identifier appears in the recall's code information.
- FDR Visionary Suite 566-16130-23Shimadzu Corporation · System, X-Ray, Stationary4 recalls0 reports
- FDR Visionary Suite 566-16130-31Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
- FDR Visionary Suite 566-16130-33Shimadzu Corporation · System, X-Ray, Stationary2 recalls0 reports
Product lines
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page