Skip to content

Product: STA Liatest D-Di recallZ-0205-2026 · Diagnostica Stago, Inc. · initiated Sep 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Diagnostica Stago, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A potential risk of underestimation of D-Dimer (D-Di) levels with the two listed lots.

FDA root cause category: Under Investigation by firm

What the company is doing

On September 12, 2025 URGENT: MEDICAL DEVICE RECALL (REMOVAL) letters were emailed to customers. Actions for customers: Discontinue use of affected lots and dispose of them in accordance with your local regulations. Complete and return response form. For additional information, please contact Stago's 24/7 Hotline at 1-800-725-0607.

Recall details

Recall number
Z-0205-2026
Status
Open, Classified
Initiated by the company
Sep 12, 2025
Posted by the FDA
Oct 16, 2025
Recalling firm
Diagnostica Stago, Inc.
Quantity
12740 units
Product code
DAP
FDA submissions
K141144, K162227, K964728
Distribution
Worldwide distribution- US Nationwide and the country of Canada.

Product description

Product: STA Liatest D-Di; REF: 00515;

Code information (models, lots, serial numbers)

REF: 00515; UDI-DI: 03607450005158; Lot numbers: 271382, 271411; Expiration 09/30/2025;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates