Salem Sump PVC Tubes recallZ-0208-2026 · Cardinal Health 200, LLC · initiated Sep 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Cardinal Health 200, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Firm has received complaints regarding Anti-Reflux Valve (ARV) breakage due to excessive force during use. Firm is providing additional guidance for use. Expansion of RES 95214.
FDA root cause category: Use error
What the company is doing
On September 18, 2025, the firm notified customers via Urgent Medical Device Product Correction Update letters. This is an update to the notice sent in July 2024 for RES 95214. The present recall includes all lots of additional product codes listed. The letter provides customers with additional guidance on how to properly use the Anti-Reflux Valve (ARV): Important Instructions: After placement of the Salem Sump tube, always inject 10ml-20ml of air before seating the Anti Reflux Valve (ARV) into the blue sump vent lumen. Instructions: 1. Seat the BLUE END of the ARV into the blue sump vent lumen of the Salem Sump tube. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 2. To cap tube, seat the WHITE END of the ARV into the suction lumen. Insert the ARV gently into the Salem Sump tube to avoid difficult removal. 3. To remove, pull valve in the same direction of assembly. Do not pull at an angle to avoid breaking the valve. Please visit https://www.mycardinalmsds.com for updated electronic IFUs.
Recall details
- Recall number
- Z-0208-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 18, 2025
- Posted by the FDA
- Oct 17, 2025
- Recalling firm
- Cardinal Health 200, LLC
- Quantity
- all lots
- Product code
- KNT
- FDA submissions
- K830710
- Distribution
- Worldwide - US Nationwide distribution including in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY, PR, GU and the countries of Australia, Austria, Bahrain, Belgium, Bulgaria, Canada, Chile, Colombia, Croatia, Czech Republic, Denmark, Ecuador, Estonia, Finland, France, French Polynesia, Germany, Greece, Hong Kong, Hungary, Ireland, Israel, Italy, Jordan, Africa, Kuwait, Luxembourg, Mexico Monaco, Netherlands, New Zealand, Northern Ireland, Norway, Pakistan, Panama, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Taiwan, UAE, United Kingdom, Vietnam.
Product description
Salem Sump PVC Tubes: Product Code/Product Name: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48 (122 cm) 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36 (91 cm) 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48 (122 cm) 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48 (122 cm) 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48 (122 cm) 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48 (122 cm)
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
Product Code/Product Name/UDI-DI: 8888265108 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 10 Fr/Ch (3.3 mm), 36 (91 cm) 10884521000766 8888265116 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 12 Fr/Ch (4.0 mm), 48 (122 cm) 10884521020078 8888265124 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 14 Fr/Ch (4.7 mm), 48 (122 cm) 10884521020085 8888265132 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 16 Fr/Ch (5.3 mm), 48 (122 cm) 10884521020092 8888265140 Salem Sump Silicone Dual Lumen Stomach Tube, Radiopaque Line, 18 Fr/Ch (6.0 mm), 48 (122 cm) 10884521020108 8888266213 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 10 Fr/Ch (3.3 mm) x 36 (91 cm) 10192253012781 8888266221 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 12 Fr/Ch (4.0 mm) x 48 (122 cm) 10884521000803 8888266247 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 14 Fr/Ch (4.7 mm) x 48 (122 cm) 10884521000827 8888266262 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 16 Fr/Ch (5.3 mm) x 48 (122 cm) 10884521000841 8888266288 Salem Sump TM Silicone Dual Lumen Stomach Tube, Anti-Reflux Valve and Radiopaque Line, 18 Fr/Ch (6.0 mm) x 48 (122 cm) 10884521000858 All Lots
Devices named in this recall
10 devices in the FDA device registry whose identifier appears in the recall's code information.
- Salem Sump 8888266114Cardinal Health 200, LLC · TUBE, DOUBLE LUMEN FOR INTESTINAL DECOMPRESSION AND/OR INTUBATION2 recalls0 reports
- Salem Sump 8888265132Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall2 reports
- Salem Sump 8888265108Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888266221Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888266247Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888266262Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888266288Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888265116Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888265124Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
- Salem Sump 8888265140Cardinal Health, Inc. · TUBES, GASTROINTESTINAL (AND ACCESSORIES)1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97539).
- Z-0206-2026 · Open, ClassifiedSalem Sump PVC Tubes: Product Code/Product Name: 8888264911E ENFit" SALEM SUMP" PVC TUBE 10FR 36IN 8888264929E ENFit" SALEM SUMP" PVC TUBE 12FR 48IN 8888264945E ENFit" SALEM SUMP…
- Z-0207-2026 · Open, ClassifiedSalem Sump PVC Tubes: Product Code/Product Name: 264929 Salem Sump" Stomach Tube, Dual Lumen, 12 Fr/Ch (4.0 mm) 264945 Salem Sump" Stomach Tube, Dual Lumen, 14 Fr/Ch (4.7 mm) 2649…
- Z-0209-2026 · Open, ClassifiedSalem Sump PVC Tubes: Product Code/Product Name: 266197 Salem Sump" Anti-Reflux Valve CE Marked 8888266197 Salem Sump" Anti-Reflux Valve
- Z-0210-2026 · Open, ClassifiedSalem Sump PVC Tubes: Product Code/Product Name: 266122 Salem Sump" Stomach Tube, Dual Lumen with Anti-Reflux Valve, 14 Fr/Ch (4.7 mm) 266130 Salem Sump" Stomach Tube, Dual Lumen …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page