Skip to content

Dexcom G6 recallZ-0318-2026 · Dexcom, Inc. · initiated Aug 28, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Dexcom, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

A software defect in version v1.15.0 of the G6 Android app can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, alerts or notifications. This could result in the missed detection of a hyperglycemic or hypoglycemic event, protentional resulting in severe hyperglycemia, diabetic ketoacidosis (DKA), or hyperosmolar hyperglycemic state (HHS).

FDA root cause category: Software Design Change

What the company is doing

On 09/18/2025, the firm sent via in-app and email an "URGENT: MEDICAL DEVICE CORRECTION NOTIFICATION" communication informing customer that a bug has been identified within the G6 Android app version 1.15.0 that can cause the app to terminate unexpectedly, which may result in the user not receiving estimated glucose values, alarms, or alerts. Customers are instructed to: Upgrading from app version 1.15.0 is mandatory and they will not be able to use app version 1.15.0 after October 6th, 2025. Update your app now to continue using the app: 1. Open the App Store and search for the Dexcom G6 app. 2. Install the latest Dexcom G6 app version. 3. Open the Dexcom G6 app. For questions or further assistance - contact Technical Support at 1-844-478-1600. Technical Support is available 24 hours a day/7 days a week

Recall details

Recall number
Z-0318-2026
Status
Open, Classified
Initiated by the company
Aug 28, 2025
Posted by the FDA
Oct 30, 2025
Recalling firm
Dexcom, Inc.
Quantity
333
Product code
QBJ
FDA submissions
K223931
Distribution
This product was distributed throughout the US. G6 Android US v1.15.0 SW11678 G6 and G6 Pro Android app release has been halted <1% rollout after this issue was identified. Please see the attached consignee data for additional details.

Product description

Brand Name: Dexcom G6 Continuous Glucose Monitoring System Product Name: Dexcom G6 and G6 Pro Android US CGM App Model/Catalog Number: SW11678 Software Version: version 1.15.0 Product Description: Dexcom G6 and G6 Pro Android US CGM App SW11678 version 1.15.0 Component: Dexcom G6 Continuous Glucose Monitoring System

Code information (models, lots, serial numbers)

Lot Code: UDI: SW11678 - 00386270000811

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page