Design Options recallZ-0124-2026 · B Braun Medical Inc · initiated Sep 18, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B Braun Medical Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Potential for the lid of the catheter connector to be in the incorrect position.
FDA root cause category: Process control
What the company is doing
B Braun notified consignees on about 09/18/2025 via USPS certified mailed URGENT MEDICAL DEVICE CORRECTION NOTIFICATION. Consignees were instructed to, where possible, have a second kit or additional catheter connector available at the point of use and upon opening the kit, verify the correct catheter connector is present. If the affected component is detected, discard and do not use. Consignees were also instructed to review the Medical Device Correction Notification and ensure all users in your organization and other concerned persons are informed, if affected units have been further distributed, notify customers, post the notification where affected units are stored, and complete and return the "Urgent Medical Device Correction Acknowledgement Form." Additionally, if consignees have identified affected units in inventory, they may order one case (50 pieces) of affected sterilized single use catheter connectors to have on hand.
Recall details
- Recall number
- Z-0124-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 18, 2025
- Posted by the FDA
- Oct 9, 2025
- Recalling firm
- B Braun Medical Inc
- Quantity
- 2,450 units
- Product code
- OGE
- Distribution
- Nationwide distribution to AK, AL, AR, AZ, CA, CO, FL, GA, HI, IA, IL, IN, KS, LA, MA, MD, ME, MI, MN, MO, MS, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, WV.
Product description
Design Options, Pain Management Tray. Product Code: SESK. Product Descriptions: 1. SESK PARKLAND MEMORIAL FX CSE TRAY. 2. SESK OWENSBORO DAVIESS HOSP 20058814. 3. SESK MEMORIAL HEALTH SYSTEM SPINAL/EPID. 4. SESK ST. JOHN'S MERCY MED CTR SPINAL/EPID. 5. SESK MAIMONIDES HOSPITAL SPINAL/EPIDURAL. 6. SESK OU MEDICAL CENTER - SPINAL/EPIDURAL. 7. SESK ELLIS HOSPITAL - SPINAL EPIDURAL. 8. STONY BROOK HSP SPINAL EPIDRAL TRAY SESK. 9. SESK UAB SPINAL EPIDURAL. 10. SESK CREIGHTON UNIV BMH SOINAL EPIDURAL. 11. SESK, BSW SPINAL EPIDURAL TRAY. 12. LYNDON B JOHNSON HOSPITAL SPINAL EPID.
Code information (models, lots, serial numbers)
Show the full code information (2k characters)
1. REF: 530029; Product Code: SESK; UDI-DI (Primary): 04046964305245; UDI-DI (Unit of Use): 04046964305238; Lot Number: 0062012031. 2. REF: 530048; Product Code: SESK; UDI-DI (Primary): 04046964305313; UDI-DI (Unit of Use): 04046964305306; Lot Number: 0062006298. 3. REF: 530155; Product Code: SESK; UDI-DI (Primary): 04046964306211; UDI-DI (Unit of Use): 04046964306204; Lot Number: 0062012054. 4. REF: 530159; Product Code: SESK; UDI-DI (Primary): 04046964669286; UDI-DI (Unit of Use): 04046964669279; Lot Number: 0062012025. 5. REF: 530169; Product Code: SESK; UDI-DI (Primary): 04046964808920; UDI-DI (Unit of Use): 04046964808913; Lot Number: 0062012035. 6. REF: 530171; Product Code: SESK; UDI-DI (Primary): 04046964950391; UDI-DI (Unit of Use): 04046964950384; Lot Number: 0062006303. 7. REF: 530180; Product Code: SESK; UDI-DI (Primary): 04046955156443; UDI-DI (Unit of Use): 04046955156436; Lot Number: 0062000738. 8. REF: 530197; Product Code: SESK; UDI-DI (Primary): 04046955619337; UDI-DI (Unit of Use): 04046955619320; Lot Number: 0062003710. 9. REF: 530194; Product Code: SESK; UDI-DI (Primary): 4046955589388; UDI-DI (Unit of Use): 4046955589371; Lot Number: 0062014360. 10. REF: 530199; Product Code: SESK; UDI-DI (Primary): 04046955619375; UDI-DI (Unit of Use): 04046955619368; Lot Number: 0062001876. 11. REF: 530160; Product Code: SESK; UDI-DI (Primary): 04046964813993; UDI-DI (Unit of Use): 04046964813986; Lot Number: 0062011011. 12. REF: 530184; Product Code: SESK; UDI-DI (Primary): 04046955156528; UDI-DI (Unit of Use): 04046955156511; Lot Number: 0062003228. Expiration Date: 07/31/2026.
Devices named in this recall
12 devices in the FDA device registry whose identifier appears in the recall's code information.
- Design Options® 530184B. Braun Medical Inc. · Spinal epidural anesthesia kit2 recalls1 report
- Design Options® 530159B. Braun Medical Inc. · Spinal epidural anesthesia kit2 recalls0 reports
- Design Options® 530029B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall1 report
- Design Options® 530048B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall1 report
- Design Options® 530155B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall1 report
- Design Options® 530180B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530194B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530197B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530199B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530169B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530160B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
- Design Options® 530171B. Braun Medical Inc. · Spinal epidural anesthesia kit1 recall0 reports
Product lines
Other products in the same recall event
The FDA groups related products into one recall event (event 97550).
- Z-0114-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (9cm) Tuohy Epidural Needle - Winged and PERIFIX FX Springwound Catheter - 19 Ga. Closed Tip. Product…
- Z-0115-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Catheter…
- Z-0116-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Cathete…
- Z-0117-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray with PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter.…
- Z-0118-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray, 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Fixed Wings, 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Product C…
- Z-0119-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Closed tip PERIFIX FX Springwound Epidural Catheter. P…
- Z-0120-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle - Winged and 19 Ga. Open tip PERIFIX FX Springwound Epidural Catheter. Pro…
- Z-0121-2026 · Open, ClassifiedPERIFIX FX Continuous Epidural Anesthesia Set, PERIFIX Tuohy Epidural Needle - 17 Ga. x 3-1/2 in. (8.9 cm) - Winged, PERIFIX FX Springwound Epidural Catheter - 19 Ga. Closed tip. P…
- Z-0122-2026 · Open, ClassifiedESPOCAN Combined Spinal and Epidural Anesthesia Tray, ESPOCAN 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 25 Ga. x 5 in. (12.7 cm) Spinal Needle with Marking and Cent…
- Z-0123-2026 · Open, ClassifiedCombined Spinal and Epidural Anesthesia Tray, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 27 Ga. x 5 in. (12.7 cm) Spinal Needle and 19 Ga. Open Tip PERIFIX F…
- Z-0125-2026 · Open, ClassifiedCombined Spinal and Epidural Anesthesia Tray 2, PERIFIX 17 Ga. x 3-1/2 in. (8.9 cm) Tuohy Epidural Needle, PENCAN 27 Ga. x 5 in. (12.7 cm) Spinal Needle, and 19 Ga. Closed Tip PERI…
- Z-0126-2026 · Open, ClassifiedDesign Options, PERIFIX FX Epidural Anesthesia Tray. Product Code: CESK. Product Descriptions: 1. CESK ROBERT WOOD JOHNSON UNIV HOSP. 2. CESK EAST JEFFERSON GEN HOSP 20071769. …
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page