Cardiac Workstation recallZ-0165-2026 · Philips North America · initiated Sep 12, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips North America or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Failure of Environmental Stress Testing from a pinched power module wire within the housing of the Cardiac Workstation device which resulted in a short.

FDA root cause category: Process control

What the company is doing

On September 12, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: Share this communication with all clinical staff to review and undedrstand. Please this letter with the device. Pass the notice along to all those who need to be aware within the organization or to any organization where affected devices were transferred. Ensure technical solution established by Philips is implemented as soon as available within the timeframe communicated by Philips. In case any device cannot be corrected, inform Philips of the reason why the Technical Solution could not be implemented. Action By Philips: A Philips representative will contact customers to arrange a replacement of the device's hardware on affected devices with the pinched power module wire within the housing of the Cardiac Workstation.

Recall details

Recall number
Z-0165-2026
Status
Open, Classified
Initiated by the company
Sep 12, 2025
Posted by the FDA
Oct 16, 2025
Recalling firm
Philips North America
Quantity
28 units
Product code
DPS
FDA submissions
K241556
Distribution
US Nationwide distribution in the states of CA, IA, TN, TX, VA.

Product description

Cardiac Workstation 5000; Model Number: 860439;

Code information (models, lots, serial numbers)

Model Number: 860439; UDI-DI: 00884838094826; Serial Numbers:

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97612).