DxFLEX Flow Cytometer Catalog Number C78500 recallZ-0541-2026 · Beckman Coulter Inc. · initiated Sep 22, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Beckman Coulter Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Due to a software error code that crashes during acquisition on patient sample when running customer defined acquisition protocol. This may lead to delayed patient results.
FDA root cause category: Under Investigation by firm
What the company is doing
On 09/22/2025. an Urgent Medical Device Recall letter was sent via postal mail and/or email informing customers that Beckman Coulter has determined that the DxFLEX software version 2.3.4.37 may crash due to a code error issue.While using the DxFLEX Flow Cytometer instrument for data acquisition using user defined acquisition protocol, the CytExpert for DxFLEX software (version 2.3.4.37) may unexpectedly crash. Due to that, the data acquisition may not be completed according to the preset conditions (i.e., number of events and/or acquisition time). When the software crash is not detected promptly, the sample may be completely consumed. Customers are instructed to: Follow the steps below when running experiments in the instrument until the software is upgraded to the version that will fix the issue: a. Prepare the experiment file (create sample, set location, etc.) b. Switch to first Tube sheet for each task to be acquired c. Save and close the experiment file d. Open the experiment again (DO NOT switch to REPORT sheet after opening experiment) This step will prevent the software from crashing due to this code error issue. e. Perform acquisition f. Analyze data. " Beckman Coulter recommends posting this letter on or near the affected analyzers until the update to the software that addresses this issue is completed. For questions: contact our Customer Support Center " From our website: http://www.beckman.com " By phone: call (800) 369-0333 in the United States.
Recall details
- Recall number
- Z-0541-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 22, 2025
- Posted by the FDA
- Nov 18, 2025
- Recalling firm
- Beckman Coulter Inc.
- Quantity
- 25 units
- Product code
- OYE
- FDA submissions
- K232600
- Distribution
- Worldwide - U.S. Nationwide distribution in the states of CA, CT, FL, IL, NY, OH, OR, TN, and TX. The country of India.
Product description
DxFLEX Flow Cytometer Catalog Number C78500 UDI-DI code: 15099590754013 The DxFLEX Flow Cytometer is intended for use as an in vitro diagnostic device for immunophenotyping up to ten fluorescent detection channels using three lasers (488 nm, 638 nm and 405 nm) and two light scatter detection channels. It is intended for use with in vitro diagnostic (IVD) assays and software that are indicated for use with the instrument.
Code information (models, lots, serial numbers)
Catalog Number: C78500 UDI-DI code: 15099590754013 Serial Numbers: BH44060, BJ09004, BJ14015, BJ09010, BJ13012, BJ14014, BJ13011, BJ14017, BJ13013, BJ09003, BJ09008, BJ09009, BH39057, BH44059, BJ14016, BJ05002, BJ17020, BJ17021, BH26031, BJ19024, BJ09005, BJ09006, BJ09007, BJ19022, BJ19023
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- DxFLEX C78500Beckman Coulter, Inc. · Flow Cytometric Reagents And Accessories.1 recall0 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page