EVIS EXERA III BRONCHOVIDEOSCOPE recallZ-0049-2026 · Olympus Corporation of the Americas · initiated Sep 11, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Additional IFU updates to provide further clarification on safe and effective use of bronchoscopes when used in conjunction with laser, argon plasma coagulation, and high-frequency therapy equipment.
FDA root cause category: Device Design
What the company is doing
An URGENT Medical Device Correction notification dated 9/11/25 was mailed to consignees to notify them of this recall action. Consignees are asked to carefully read the recall notification and identify any affected devices in their inventory. Consignees are to ensure all personnel are knowledgeable and aware of the updated IFU. Consignees are to acknowledge receipt of the recall notification through Olympus' web portal at https://olympusamerica.com/recall. Forward the provided recall notification if product was further distributed. Consignees with any questions can contact Cynthia Ow by phone at 647-999-3203 or by email at [email protected].
Recall details
- Recall number
- Z-0049-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 11, 2025
- Posted by the FDA
- Oct 16, 2025
- Recalling firm
- Olympus Corporation of the Americas
- Quantity
- 4,297 units US; 4,290 units OUS
- Product code
- EOQ
- FDA submissions
- K121959
- Distribution
- US Nationwide Distribution.
Product description
EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190
Code information (models, lots, serial numbers)
Model No. BF-1TH190; UDI: 4953170434778; All Serial No.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)3 recalls333 reports
Product lines
Other devices under the same FDA submission
10 devices cleared under K121959 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- EVIS EXERA III VIDEO SYSTEM CENTER OLYMPUS CV-190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum1 recall7,688 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope and accessories, flexible/rigid6,617 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-1TH190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)3 recalls5 deaths1,557 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · BRONCHOSCOPE (FLEXIBLE OR RIGID)4 recalls1 death1,214 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum5 recalls280 reports
- BRONCHOVIDEOSCOPE BF-Q170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum5 recalls151 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)2 recalls240 reports
- BRONCHOVIDEOSCOPE BF-1TQ170Olympus Medical Systems Corp. · Endoscope, accessories, narrow band spectrum2 recalls160 reports
- EVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190Olympus Medical Systems Corp. · Bronchoscope (Flexible Or Rigid)2 recalls69 reports
- EVIS EXERA III XENON LIGHT SOURCE OLYMPUS CLV-190Olympus Medical Systems Corp. · Colonoscope And Accessories, Flexible/Rigid77 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97441).
- Z-0045-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T150
- Z-0046-2026 · Open, ClassifiedEVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1T180
- Z-0047-2026 · Open, ClassifiedOES BRONCHOFIBERSCOPE OLYMPUS BF TYPE 1T60
- Z-0048-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF-1TH1100
- Z-0050-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF-1TQ170
- Z-0051-2026 · Open, ClassifiedEVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE 1TQ180
- Z-0052-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF-H1100
- Z-0053-2026 · Open, ClassifiedEVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-H190
- Z-0055-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF TYPE P150
- Z-0056-2026 · Open, ClassifiedEVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE P180
- Z-0057-2026 · Open, ClassifiedEVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-P190
- Z-0058-2026 · Open, ClassifiedOES BRONCHOFIBERSCOPE OLYMPUS BF TYPE P60
- Z-0059-2026 · Open, ClassifiedBRONCHOFIBERSCOPE OLYMPUS BF TYPE PE2
- Z-0060-2026 · Open, ClassifiedBRONCHOVIDEOSCOPE OLYMPUS BF-Q170
- Z-0061-2026 · Open, ClassifiedEVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180
- Z-0062-2026 · Open, ClassifiedEVIS EXERA II BRONCHOVIDEOSCOPE OLYMPUS BF TYPE Q180-AC
- Z-0063-2026 · Open, ClassifiedEVIS EXERA III BRONCHOVIDEOSCOPE OLYMPUS BF-Q190
- Z-0064-2026 · Open, ClassifiedBRONCHOFIBERSCOPE OLYMPUS BF TYPE TE2
- Z-0065-2026 · Open, ClassifiedEVIS EXERA BRONCHOVIDEOSCOPE OLYMPUS BF TYPE XT160
- Z-0066-2026 · Open, ClassifiedEVIS EXERA b BRONCHOVIDEOSCOPE OLYMPUS BF-XT190
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page