ICU Medical recallZ-0362-2026 · ICU Medical, Inc. · initiated Sep 8, 2025
Class II Open, Classified All recalls posted in October 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from ICU Medical, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Port weld of drug transfer device may separate or break during use and potentially result in a leak.
FDA root cause category: Under Investigation by firm
What the company is doing
On September 8, 2025, ICU Medical issued an "Urgent Medical Device" Recall notification via E-Mail. ICU Medical ask consignees to take the following actions: 1. Check all inventory locations within your institution for the affected products listed in recall notification and discontinue use. Destroy all affected products following your institution's process for destruction. If destroying is not immediately possible at your facility, then the product should be quarantined until disposal. 2. Share this notification with all potential users of the device, to ensure they are aware of this notification. If the devices are used at another location, please ensure this communication is delivered there. 3. Complete and return the attached Customer Response Form to [email protected] within 10 days of receipt to acknowledge your understanding of this notification. 4. If you have distributed affected products to your customers, please immediately forward this notice to them. Request that they complete the response form and return it to [email protected] For further inquiries, please contact ICU Medical Customer Service via email: [email protected] or call: 1-(800)-258-5361 for additional information or technical assistance. Or contact Field Service Processing via email: [email protected] for Assistance with Customer Notifications and Response Forms.
Recall details
- Recall number
- Z-0362-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Sep 8, 2025
- Posted by the FDA
- Oct 27, 2025
- Recalling firm
- ICU Medical, Inc.
- Quantity
- 200 units
- Product code
- ONB
- FDA submissions
- K131549
- Distribution
- Worldwide distribution: US (Nationwide) including states of: AK AL AR AZ CA CO CT DE FL GA IA ID IL KS KY LA MA MD MI MN MS MT NC NE NJ NM NV NY OH OK OR PA SD TN TX VA WA WY and OUS (foreign) to countries of: Canada, France, and Japan.
Product description
ICU Medical Oncology Kit w/2 ChemoLock", 13mm Closed Vial Spike w/ChemoLock" Port, ChemoLock" Bag Spike, Ext Set w/ChemoLock" Port, Clamp, Graduated Adapter REF: CL4188
Code information (models, lots, serial numbers)
Lot:14369087 ; Distributed in Canada
Other devices under the same FDA submission
91 devices cleared under K131549 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.
- ICU Medical CL3011Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls26 reports
- ICU Medical CL3511Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls6 reports
- ICU Medical CL4136Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls1 report
- ICU Medical CL-80Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system3 recalls0 reports
- ICU Medical CL3900Icu Medical, Inc. · Set, administration, intravascular3 recalls0 reports
- ICU Medical CL3960Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls26 reports
- ICU Medical CL4130Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls6 reports
- ICU Medical CL4131Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls6 reports
- ICU Medical CL-80SIcu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls5 reports
- ICU Medical CL-82Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls3 reports
- ICU Medical CL-80S-5Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls2 reports
- ICU Medical CL-72Icu Medical, Inc. · Closed antineoplastic and hazardous drug reconstitution and transfer system2 recalls1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 97583).
- Z-0328-2026 · Open, ClassifiedChemoLock Bag Spike, 01 units REF: CL-10-10
- Z-0329-2026 · Open, ClassifiedICU Medical Chemo Lock Universal Vented Vial Spike REF: CL-70-10
- Z-0330-2026 · Open, ClassifiedICU Medical ChemoLock" Port, Bulk, Non-Sterile REF: CL2100-NS
- Z-0331-2026 · Open, ClassifiedICU Medical Clave Bag Spike w/ChemoLock Port REF: CL2150
- Z-0332-2026 · Open, ClassifiedICU Medical 30" (76 cm) Appx 3.3 mL, Admin Set w/20 Drop Integrated ChemoLock Drip Chamber, Spiros", Hanger, Drop-In Red Cap REF: CL3011
- Z-0333-2026 · Open, ClassifiedICU Medical 30" (76 cm) Appx 3.6 ml, 20 Drop Admin Set w/Integrated ChemoLock Port Drip Chamber, ChemoLock w/Red Cap, Bag Hanger REF: CL3511
- Z-0334-2026 · Open, ClassifiedICU Medical Oncology Kit w/31" (79 cm) 20 Drop Admin Set w/Integrated ChemoLock" Port Drip Chamber, Spiros"; ChemoLock"; ChemoLock" Universal Vented Vial Spike; Bag Hanger REF: C…
- Z-0335-2026 · Open, ClassifiedICU Medical ChemoLock Bag Spike with Clave Additive Port REF: CL3940
- Z-0336-2026 · Open, ClassifiedICU Medical Oncology Kit w/ChemoLock, ChemoLock Port REF: CL4136
- Z-0337-2026 · Open, ClassifiedICU Medical Oncology Transfer Kit w/ChemoClave and ChemoLock Port; Spiros w/Red Cap; ChemoLock Port REF: CL4138
- Z-0338-2026 · Open, ClassifiedICU Medical 45" (114 cm) Appx 1.7 ml, Smallbore Ext Set w/Spiros w/Red Cap, Anti-Siphon Valve, Clamp, Y-Connector w/ChemoLock Port, Rotating Luer REF: CL4151
- Z-0339-2026 · Open, ClassifiedICU Medical 60" (152 cm) Appx 1.2 ml, Ext Set, Smallbore w/ChemoLock Port, Clamp, Spiros w/Red Cap REF: CL4173
- Z-0340-2026 · Open, ClassifiedICU Medical 73" (185 cm) Appx 2.5 ml, Ext Set, Smallbore w/ChemoLock Port, Clamp, ChemoLock w/Red Cap REF: CL4176
- Z-0341-2026 · Open, ClassifiedICU Medical ChemoLock Bag Spike REF: CL-10
- Z-0342-2026 · Open, ClassifiedICU Medical Double ChemoLock Port Syringe Transfer Set REF: CL-33
- Z-0343-2026 · Open, ClassifiedICU Medical ChemoLock Universal Vented Vial Spike REF: CL-70
- Z-0344-2026 · Open, ClassifiedICU Medical ChemoLock Vented Vial Spike, 20mm REF: CL-70S
- Z-0345-2026 · Open, ClassifiedICU Medical ChemoLock Vented Vial Spike, 13mm REF: CL-72
- Z-0346-2026 · Open, ClassifiedICU Medical ChemoLock Closed Vial Spike REF: CL-80
- Z-0347-2026 · Open, ClassifiedICU Medical ChemoLock Vial Spike, 20mm REF: CL-80S
- Z-0348-2026 · Open, ClassifiedICU Medical ChemoLock Closed Vial Spike w/Skirt, 5 Units REF: CL-80S-5
- Z-0349-2026 · Open, ClassifiedICU Medical ChemoLock 13mm Closed Vial Spike REF: CL-82
- Z-0350-2026 · Open, ClassifiedICU Medical ChemoLock Port REF: CL2100
- Z-0351-2026 · Open, ClassifiedICU Medical 21" Ext Set w/ChemoLock Port, Clamp, Graduated Connector REF: CL3360
- Z-0352-2026 · Open, ClassifiedICU Medical Graduated Connector w/ChemoLock Port REF: CL3900
- Z-0353-2026 · Open, ClassifiedICU Medical 6.5" Bag Spike w/ChemoLock Port Additive Port, ChemoLock Port REF: CL3955
- Z-0354-2026 · Open, ClassifiedICU Medical ChemoLock Vented Vial Spike, 20mm, 20 Units REF: 011-CL-70S-20
- Z-0355-2026 · Open, ClassifiedICU Medical ChemoLock w/Mini Bag Spike REF: C-17
- Z-0356-2026 · Open, ClassifiedICU Medical Syringe Transfer Set w/MicroClave, ChemoLock Port REF: CL-34
- Z-0357-2026 · Open, ClassifiedICU Medical ChemoLock Port w/Spiros, Red Cap REF: CL-35
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page