uCT recallZ-0218-2026 · UIH Technologies LLC · initiated Sep 12, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from UIH Technologies LLC or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Third-party outer covers can become trapped between patient table and gantry, causing mechanical interference during table movement.
FDA root cause category: Other
What the company is doing
On September 12, 2025, UIH Technologies LLC, issued a recall notification to affected consignees. UIH asked consignees to take the following actions: 1. Ensure patient positioning clearance: When positioning the patient, confirm that arms and legs and clear of all table gaps and moving parts. 2. Avoid use of third-party slickers: We advise against the use of any non-UIH-approved slickers. 3. Service Contact: For technical assistance, contact your United Imaging Service engineer or reach out to our 24/7 Customer Contact Center, Toll-Free: (855) 221-1552/Direct: (832) 699-2799 4.Report adverse events: if any adverse events or quality issue occur, please report them to the FDA via the MedWatch program. 5. Please share this information with all relevant personnel at your facility, including technologist, service staff, and clinical teams.
Recall details
- Recall number
- Z-0218-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 12, 2025
- Posted by the FDA
- Oct 21, 2025
- Recalling firm
- UIH Technologies LLC
- Quantity
- 22 units
- Product code
- JAK
- FDA submissions
- K181414, K200016
- Distribution
- US: OUS:
Product description
Shanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 550 MD, Rx Only
Code information (models, lots, serial numbers)
UDI:(01)06971576831036/ Serial Numbers: 353020 353021 353023 353045 353049 354001 354002 354003 354004 354005 354006 354007 354008 354009 354010 354011 354012 354013 354014 354015 354016 354017
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- uCT uCT 550Shanghai United Imaging Healthcare Co., Ltd. · System, X-Ray, Tomography, Computed2 recalls0 reports
Product lines
Other devices under the same FDA submission
1 device cleared under K181414, K200016 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- uCT uCT 530Shanghai United Imaging Healthcare Co., Ltd. · System, X-Ray, Tomography, Computed3 recalls0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97668).
- Z-0217-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 530 MD, Rx Only
- Z-0219-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 760 MD, Rx Only
- Z-0220-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT 780 MD, Rx Only
- Z-0221-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Computed Tomography X-ray System, Model:uCT ATLAS MD, Rx Only
- Z-0222-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uEXPLORER MD, Rx Only
- Z-0223-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 550 MD, Rx Only
- Z-0224-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI 780 MD, Rx Only
- Z-0225-2026 · Open, ClassifiedShanghai United Imaging Healthcare Co., Ltd. Positron Emission Tomography and Computed Tomography System Model:uMI Panorama 35 MD, Rx Only
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page