IMMULITE 2000 Systems IPT Intact PTH recallZ-0366-2026 · Siemens Healthcare Diagnostics, Inc. · initiated Sep 22, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Siemens Healthcare Diagnostics, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The potential for falsely depressed Intact PTH patient results at the low end of the assay range, less than or equal to 50 pg/mL (less than or equal to 5.3 pmol/L), when using specific lots identified by the firm on the IMMULITE 2000/IMMULITE 2000 XPi systems. The bias is observed in both serum and plasma samples.

FDA root cause category: Under Investigation by firm

What the company is doing

On September 22, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Customer Actions: Please review this letter with your Medical Director to determine the appropriate course of action, including for any previously generated results, if applicable. Discontinue use of and discard the kit lot listed in the table above (Products Section). Complete and return the Field Correction Effectiveness Check Form attached to this letter within 30 days. For continued Intact PTH testing in your laboratory during this time, please contact your Siemens representative to discuss alternative Siemens Healthineers solutions. As an alternative, customers who have an ADVIA Centaur system or an Atellica IM/CI may utilize the corresponding iPTH assay on those systems. These platforms are not impacted. Please retain this letter with your laboratory records and forward this letter to those who may have received this product. A follow-up communication will be provided when assay performance has been restored. ***Update 2/18/2026*** On February 10, 2026 Follow-up Urgent Medical Device Correction letters were sent to additional customers to notify of the addition of new lots for the recall. Customer actions remained the same as the original letter.

Recall details

Recall number
Z-0366-2026
Status
Open, Classified
Initiated by the company
Sep 22, 2025
Posted by the FDA
Oct 27, 2025
Recalling firm
Siemens Healthcare Diagnostics, Inc.
Quantity
3318 units (169 US, 3149 OUS)
Product code
CEW
FDA submissions
K013566
Distribution
Worldwide - US Nationwide distribution in the states of AL, CA, DC, FL, ID, IN, LA, MD, NC, NJ, NY, PA, SC and the countries of Argentina, Australia, Bolivia, Bosnia Herzeg, Brazil, Bulgaria, Canada, Chile, Colombia, Cyprus, Czech Republic, Ecuador, Egypt, France, Germany, Greece, Guatemala, Honduras, Hungary, India, Iran, Israel, Italy, Kosovo, Kuwait, Latvia, Lebanon, Macedonia, Mexico, Nepal, Netherlands, P.R. China, Pakistan, Palestinian Ter, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russian Fed., Serbia, Slovakia, Slovenia, South Africa, Spain, Taiwan, Tunisia, Turkey, Turkmenistan, U.A.E., Ukraine, Uruguay, Uzbekistan.

Product description

Assay: IMMULITE 2000 Intact PTH; Test Code: iPT; Siemens Material Number (SMN): (1) 10381441, (2) 10381442; Catalog Number: (1) L2KPP2, (2) L2KPP6;

Code information (models, lots, serial numbers)

1. Siemens Material Number (SMN): 10381441; Catalog Number: L2KPP2; UDI-DI: 00630414961910; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;  2. Siemens Material Number (SMN): 10381442; Catalog Number: L2KPP6; UDI-DI: 00630414961927; Lot Number: 385, ***Expanded 2/18/2026*** 387, 388, 389;

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

2 devices cleared under K013566 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page