Skip to content

Excelsior recallZ-0234-2026 · Medline Industries, LP · initiated Sep 11, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Medline Industries, LP or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Medline has identified a limited quantity of saline IV flush syringes that may not meet the required concentration of sodium chloride.

FDA root cause category: Process control

What the company is doing

Medline issued an Urgent Medical Device Recall notice to its consignees on 9/11/2025 via email and mail. The notice discussed the issue, potential risk, and noted that affected boxes of syringes were also inadvertently packaged within cases of syringe lots that meet all requirements and were still safe for use. Within the item number, EMZ111240, the firm specified which lots were to be destroyed and which lots were safe to use. Cases of certain lots labeled with safe to use codes must be opened and inspected for affected units which must be destroyed. A response is required to receive credit for the destroyed affected units. For questions, please contact the Recall Department at 866-359-1704 or [email protected].

Recall details

Recall number
Z-0234-2026
Status
Open, Classified
Initiated by the company
Sep 11, 2025
Posted by the FDA
Oct 23, 2025
Recalling firm
Medline Industries, LP
Quantity
8,267,150 syringes
Product code
NGT
FDA submissions
K173348
Distribution
Distribution US nationwide and Canada

Product description

Excelsior Medical, 0.9% Sodium Chloride Injection, USP - 10 mL Fill, Item Number EMZ111240. Prefilled saline flush syringes

Code information (models, lots, serial numbers)

UDI (Case): 10363807000039, UDI (Box): 363807000032, UDI (Each): 363807100114, Lot Numbers: 25AJA049, 25BJA001, 25BJA032, 25CJA053, 25CJA105, 25CJA158.

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

1 device cleared under K173348 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page