Zenition recallZ-0241-2026 · Philips Medical Systems Nederland B.V. · initiated Sep 29, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Philips Medical Systems Nederland B.V. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Over time some systems may experience loss of imaging functionality and/or poor image quality due to potential corrosion of the Image Intensifier Television (IITV) control board.

FDA root cause category: Process control

What the company is doing

On September 29, 2025 URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. Circulate this notice to all users and follow the instructions in the letter. 2. Affected systems may continue to be used in accordance with their intended use and Instructions for Use (IFU). 3. Ensure the letter is in a place likely to be seen/viewed. 4. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction letter to that organization and inform Philips about this transfer through your local Philips representative. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions by Philips: Starting in October 2025, Philips will replace the 11W control board of the affected Zenition 50 systems. Your local Philips representative will contact you to schedule a visit to perform this activity in your system.

Recall details

Recall number
Z-0241-2026
Status
Open, Classified
Initiated by the company
Sep 29, 2025
Posted by the FDA
Oct 23, 2025
Recalling firm
Philips Medical Systems Nederland B.V.
Quantity
29 units
Product code
OWB
FDA submissions
K183101
Distribution
Distribution US states of MO and TX; Belgium, Canada, Fiji, Germany, India, Malaysia, Mexico, Morocco, South Africa, South Korea, Thailand, United Kingdom, and Uzbekistan.

Product description

Zenition 50; Product Number (REF): 718096; Interventional fluoroscopic x-ray system

Code information (models, lots, serial numbers)

Product Number (REF): 718096; UDI-DI: 00884838091535; All serial numbers;