Skip to content

HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm… recallZ-0490-2026 · Stryker Sustainability Solutions · initiated Sep 24, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Sustainability Solutions or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Due to manufacturing issue, reprocessed shears, may become stuck in the "on position" resulting in continuous activation.

FDA root cause category: Process control

What the company is doing

On September 23, 2025, Stryker issued an "Urgent: Medical Device Recall" notification dated September 22, 2025, to affected consignees. Styker asked consignees to take the following actions: 1. Inform individuals within your organization who need to be aware of this action. 2. Check your internal inventory to locate the affected product (Attachment B). Segregate the affected products, quarantine and discontinue use. 3. Please complete and return the Business Reply Form to [email protected] or use the QR Code on Attachment A. Response is required. a. If you have indicated that product remains in your inventory, a prepaid shipping label will be issued for the return of the product. Credit will be issued to your facility for products returned upon receipt. 4. Maintain awareness of this communication internally until all required actions have been completed within your facility. For any additional information or questions please reach out to [email protected] or Stryker Customer Service at 800-624-4422.

Recall details

Recall number
Z-0490-2026
Status
Open, Classified
Initiated by the company
Sep 24, 2025
Posted by the FDA
Nov 12, 2025
Recalling firm
Stryker Sustainability Solutions
Quantity
22, 476 units
Product code
NLQ
FDA submissions
K170456
Distribution
US Nationwide distribution and OUS (foreign) to country of: Canada.

Product description

HARMONIC FOCUS Shears + Adaptive Tissue Technology, 9 cm Length REF: HAR9F, Reprocessed Device for Single Use, STERILEEO, RX ONLY

Code information (models, lots, serial numbers)

Show the full code information (2k characters)

UDI: (01)37613327348249/ 162xxxxx Lot Numbers: 16275164 16284043 16288055 16294327 16296918 163xxxxx Lot Numbers: 16301496 16305751 16306347 16306436 16316531 16316651 16342389 16343642 16346325 16346612 16347602 16348678 16349187 16353884 16355545 16357189 16360679 16363788 16366396 16367903 16368245 16375951 16376461 16378500 16380079 16382574 16383957 16386402 16399914 164xxxxx Lot Numbers: 16405308 16407386 16408732 16410690 16413617 16413975 16416261 16418939 16420520 16422830 16432199 16432342 16436871 16436987 16442725 16444477 16446841 16448857 16449208 16449296 16453140 16459248 16463302 16464895 16467199 16471023 16475746 16482558 16487182 16488252 16491206 16492941 16494926 16497950 16498718 165xxxxx Lot Numbers: 16502233 16503856 16506269 16507481 16511866 16511995 16514496 16514579 16520053 16521262 16521892 16522603 16527022 16527858 16530602 16531805 16534581 16535595 16538257 16539232 16540322 16540636 16549592 16551658 16554621 16559625 16567068 16567902 16569680 16573872 16575403 16576366 16576418 16579676 16583068 16585378 16587095 16587292 16592009 16592273 16595099 16595611 16598700 166xxxxx Lot Numbers: 16602578 16604258 16607822 16612246 16614330 16617406 16625703 16630079 16630328 16632353 16636528 16640175 16641167 16644299 16645144 16645768 16649786 16652064 16655193 16656536 16659901 16661415 16664176 16681913 16683266 16690151 16693679 167xxxxx Lot Numbers: 16705748 16759491 16762280 16763927 16766282 16767078 16768138 16768724 16769146 16775982 16778745 16779900 16795717 16796578 16797254 16797381 16797381 16799710 168xxxxx Lot Numbers: 16801351 16801564 16804498 16806005 16821036 16828177 16843649 16844103 16848493 16852781 16864607 16867268 16881353 16888650 16895321 16898535 169xxxxx Lot Numbers: 16910107 16913454 16920690 16928682 16931180 16935100 16945644 16951372 16962742 16967419 16969395 16977884 16979276 16981365 16988861 16991698 16993581 16996795 170xxxxx Lot Numbers: 17001825 17009610 17012825 17016204 17021294

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page