Endo Surgical Navigation Advanced Platform Patient… recallZ-0323-2026 · Surgical Theater Inc · initiated Sep 16, 2025
Completed
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Surgical Theater Inc or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
A software anomaly, under specific conditions when large rotational alignment values are applied in 2D Image Fusion, may lead to misalignment of any secondary image layer(s) to the primary image layer.
FDA root cause category: Under Investigation by firm
What the company is doing
Surgical Theater issued an URGENT: Medical Device Software correction notice dated September 17, 2025 to its consignees on 09/22/2025 via email. The notice explained the issue, risk to health, and provided the following: Customers can continue to use the software version(s) noted above until a software patch is made available by Surgical Theater. Until the software patch is installed, please follow these steps: 1. Read and share this notice with all relevant staff. 2. Continue to use the 3D Fusion functions to review the fusion results for all case files to verify alignment of secondary layer(s) to primary layer and to adjust fusion as necessary, and 3. Acknowledge your receipt of this notification by completing and returning the included acknowledgement form via email to [email protected]. A software patch is currently undergoing verification and should be available in the next 10 days. Surgical Theater has revised the 2D Image Fusion function to ensure the x, y and z axis rotation values are consistently applied when generating the aligned Axial, Sagittal and Coronal images. In addition, the rotate left/right buttons in all three planes were modified so the volume rotation stays consistent with the respective plane orientation. A Surgical Theater representative will contact your facility to schedule installation once the patch is available. For questions regarding this notification contact: Kevin Murrock, Sr. Director of Quality & Regulatory, Tel. 330-472-6520 (Monday through Friday, 9:00 AM to 5:00 PM, Eastern Time), email: [email protected]
Recall details
- Recall number
- Z-0323-2026
- Status
- Completed
- Initiated by the company
- Sep 16, 2025
- Posted by the FDA
- Oct 24, 2025
- Recalling firm
- Surgical Theater Inc
- Quantity
- 19 units
- Product code
- LLZ
- FDA submissions
- K160584
- Distribution
- Worldwide - US Nationwide distribution in the states of MI, FL, DC, MN, CA, TX, NJ, PA, AZ, NY, KY, IN, WV and the countries of Italy, Germany, Israel, UK.
Product description
Endo Surgical Navigation Advanced Platform Patient Engagement (EndoSNAP), software version SNP.1.12.0, imaging processing radiological system
Code information (models, lots, serial numbers)
UDI/DI 00863157000329
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Endo SNAPSurgical Theater LLC · System, Image Processing, Radiological1 recall0 reports
Other devices under the same FDA submission
3 devices cleared under K160584 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- SyncAR SNAPSurgical Theater LLC · System, Image Processing, Radiological1 recall0 reports
- SNAP ORSurgical Theater LLC · System, Image Processing, Radiological1 recall0 reports
- SNAP Patient Engagement PE SNAPSurgical Theater LLC · System, Image Processing, Radiological1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97715).
- Z-0319-2026 · CompletedSuRgical Planner (SRP), software version SRP.7.12.0, imaging processing radiological system
- Z-0320-2026 · CompletedSurgical Navigation Advanced Platform (SNAP), software version SNP.3.12.0, imaging processing radiological system
- Z-0321-2026 · CompletedSurgical Navigation Advanced Platform Patient Engagement (PE SNAP), software version SNP.3.12.0, imaging processing radiological system
- Z-0322-2026 · CompletedSyncAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SNP.1.12.0, imaging processing radiological system
- Z-0324-2026 · CompletedSpineAR Surgical Navigation Advanced Platform Patient Engagement (SyncAR SNAP), software version SPR.1.2.0, orthopedic augmented reality
- Z-0325-2026 · CompletedSurgical Planner, BrainSTORM, software version BST.9.3.0, orthopedic augmented reality
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.