LOGIQ recallZ-0156-2026 · GE Medical Systems, LLC · initiated Sep 18, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from GE Medical Systems, LLC or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

The Ultrasound-Guided Attenuation Parameter (UGAP) measurement data may display inaccurate values representing liver steatosis. This could potentially lead to inappropriate clinical decisions impacting overall care.

FDA root cause category: Software design

What the company is doing

An URGENT MEDICAL DEVICE CORRECTION notification letter dated 9/18/25 was sent to customers. Actions to be taken by Customer/User For LOGIQ P9 and LOGIQ P10 series ultrasound systems with system software version R4.5.7, discontinue using the UGAP feature (please see Appendix for product identification). You may continue using other imaging features. After the correction has been implemented, please destroy the installation media for software R4.5.7 at your site. Please ensure all potential users in your facility are made aware of this safety notification and the recommended actions. Please retain this document for your records. Please complete and return the attached acknowledgement form to [email protected]. GE HealthCare will correct all affected products at no cost to you. A GE HealthCare representative will contact you to arrange for the correction. If you have any questions or concerns regarding this notification, please contact GE HealthCare Service at 1-800-437-1171 or your local Service Representative.

Recall details

Recall number
Z-0156-2026
Status
Open, Classified
Initiated by the company
Sep 18, 2025
Posted by the FDA
Oct 15, 2025
Recalling firm
GE Medical Systems, LLC
Quantity
31 units
Product code
IYN
FDA submissions
K214039
Distribution
US Nationwide. Global Distribution.

Product description

LOGIQ P10 R4.5 HD ultrasound system with system software version R4.5.7., Model Number 5877535

Code information (models, lots, serial numbers)

Model Number 5877535 UDI-DI: 00195278464668, Serial Numbers: LPZ490338 LPZ490347 LPZ490339 LPZ490147 LPZ490292 LPZ490019 LPZ490252 LPZ490348 LPZ490303 LPZ490304 LPZ490305 LPZ490306 LPZ490340 LPZ490350 LPZ490341 LPZ440658 LPZ440673 LPZ440674 LPZ490337 LPZ440641 LPZ440192 LPZ440258 LPZ440694 LPZ440695 LPZ440692 LPZ440632 LPZ440185 LPZ440107 LPZ440168 LPZ440046 LPZ440549

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other devices under the same FDA submission

3 devices cleared under K214039 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97726).

Sources and updates