Skip to content

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number:… recallZ-0327-2026 · Canon Medical System, USA, INC. · initiated Oct 16, 2025

Class II Open, Classified All recalls posted in October 2025

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Canon Medical System, USA, INC. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected system contains components that do not meet the required distance between electrodes for compliance with the IEC 60601-1 standard for medical electrical equipment.

FDA root cause category: Device Design

What the company is doing

Cannon Medical began notifying consignees via email on about 10/16/2025. Consignees were instructed that a representative of Canon Medical Systems USA will contact them to schedule a time to replace the relevant parts. If the message "Gradient power supply error is displayed before the replacement is performed, immediately stop using the system and contact your service representative. Consignees were requested to complete and return the provided response form

Recall details

Recall number
Z-0327-2026
Recall class
Class II
Status
Open, Classified
Initiated by the company
Oct 16, 2025
Posted by the FDA
Oct 24, 2025
Recalling firm
Canon Medical System, USA, INC.
Quantity
2 units
Product code
LNH
FDA submissions
K243335
Distribution
US distribution to Ohio and Nevada

Product description

VANTAGE GALAN 3T. CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7.

Code information (models, lots, serial numbers)

CMSC Model: MRT-3020/U7. Model Number: MEXL-3020/B7. Serial Numbers: B7A2462002, B7A2522014.

Other devices under the same FDA submission

1 device cleared under K243335 — the recall does not name any of them by identifier. One submission often covers many models, so they may or may not be affected.

Sources and updates