BEAR Implant (Bridge-Enhanced ACL Restoration) recallZ-0479-2026 · Miach Orthopaedics · initiated Feb 2, 2023
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Miach Orthopaedics or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
The field action was initiated in February of 2023 and was limited to the affected lot from sales force trunk stock only.
FDA root cause category: Under Investigation by firm
What the company is doing
On January 30, 2023 RE: BEAR Implant Hydration memo emails were sent to sales staff. If you observe an implant that hydrates too quickly (e.g., less than 30 seconds) and appears that it may not remain intact upon insertion into the knee, do not attempt to implant the device. We would recommend you always have a backup device available to address such circumstances. We plan to revise the Tips and Pearls document, ML-1027, to provide further instructions/details on how to properly hydrate the BEAR device. In the meantime, one way to avoid shortened hydration time is to consider hydrating the dense side of the implant first and continue toward the porous side . To date, all reports of this observation have involved Lot # 7008409. As a precaution and to ensure customer satisfaction, we are asking that you return all devices from this lot that are currently in your trunk stock (in your possession). Chris Brown will work with you to coordinate returns as well as replacing these trunk stock devices. Please contact me ([email protected]) or Arjun ([email protected]) if you have questions.
Recall details
- Recall number
- Z-0479-2026
- Status
- Open, Classified
- Initiated by the company
- Feb 2, 2023
- Posted by the FDA
- Nov 7, 2025
- Recalling firm
- Miach Orthopaedics
- Quantity
- 167 units
- Product code
- QNI
- FDA submissions
- K243578
- Distribution
- US Nationwide distribution.
Product description
BEAR Implant (Bridge-Enhanced ACL Restoration); Model/Catalog: 1000;
Code information (models, lots, serial numbers)
Model/Catalog: 1000; UDI-DI: 00860002987804; Lot number: 7008409;
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- BEAR Implant (Bridge-Enhanced ACL Restoration) 1000Miach Orthopaedics, Inc. · Resorbable Implant For Anterior Cruciate Ligament (Acl) Repair2 recalls29 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page