Skip to content

EZDilate recallZ-0572-2026 · Olympus Corporation of the Americas · initiated Oct 17, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Olympus Corporation of the Americas or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Device kits contained an incorrectly sized 4mm x 4cm urinary tract balloon catheter (BURS0404) instead of the correct 8mm x 15cm nephrostomy tract balloon catheter (BPCN0815). All other components in the kits were correctly included and packaged in the kit.

FDA root cause category: Process control

What the company is doing

On October 17, 2025, "URGENT: MEDICAL DEVICE REMOVAL" letters were sent to customers. Actions required: 1. 1. Examine your inventory and quarantine any affected devices. 2. Cease usage of the product with immediate effect. 3. Return any remaining affected product, please contact Customer Service at 1-800-848-9024, option 2, to obtain a Return Material Authorization. Olympus will issue a credit to your facility upon return of your affected product. 4. Please forward this notice to other users who may have the affected products if you have further distributed it. Olympus requests you to report any complaints to our Technical Assistance Center (TAC) at 1- 800-848-9024, option 1, and the FDA If you require additional information, please do not hesitate to contact directly by phone at (647) 999-3203 or by e-mail [email protected].

Recall details

Recall number
Z-0572-2026
Status
Open, Classified
Initiated by the company
Oct 17, 2025
Posted by the FDA
Nov 20, 2025
Recalling firm
Olympus Corporation of the Americas
Quantity
60 units
Product code
LJE
FDA submissions
K132383
Distribution
US Domestic: AK, CA, IA, KY, WA, WI; OUS International: Canada and Germany.

Product description

Brand Name: EZDilate Nephrostomy Balloon Catheter and Inflation Device Kit Product Name: Olympus Nephro-EZDilate Nephrostomy Balloon Catheter Model/Catalog Number: BPCN0815K Product Description: The Nephro-EZDilate nephrostomy balloon catheter is a reinforced catheter attached to a distal dilatation balloon. It has a radiopaque tip and two radiopaque markers positioned on the catheter inside the balloon that define the working length

Code information (models, lots, serial numbers)

Model: BPCN0815K; UDI: 00821925033016; Lot Number: 25037006;

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.