Vivacit-E recallZ-0419-2026 · Zimmer, Inc. · initiated Sep 26, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Zimmer, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
Zimmer Biomet is conducting a medical device recall for two (2) lots of the Dual Mobility Vivacit-E Bearing due to a commingle. The outer package is labeled as a Size F 44 mm, however, the implant inside the package is a Size G 46 mm, and vice versa.
FDA root cause category: Labeling mix-ups
What the company is doing
Zimmer Biomet issued an URGENT MEDICAL DEVICE recall notice to its consignees on 9/25/2025 via FedEx, 2nd Day Delivery. The notice explained the issue, potential risks, and requested the return of all affected devices in inventory. Risk Managers were notified that the sales representatives would be removing the affected products. Recommendation for patient monitoring was made to the implanting surgeons. For further questions or concerns after reviewing this notice, please call customer service at 574-371-3071 between 8:00 a.m. and 5:00 p.m. EST, Monday through Friday. Calls received outside of call center operating hours will receive a voicemail prompt or be transferred to an on-call representative in the event of an emergency. Alternatively, questions may be emailed to [email protected].
Recall details
- Recall number
- Z-0419-2026
- Status
- Open, Classified
- Initiated by the company
- Sep 26, 2025
- Posted by the FDA
- Nov 3, 2025
- Recalling firm
- Zimmer, Inc.
- Quantity
- 63 units
- Product code
- LPH
- FDA submissions
- K190656
- Distribution
- US: OH, GA, WI, NE, IA, IL, CA, NY, MD, NC, PA, AL, TN, MI, WA, FL, TX, KS, NM, AZ, and CANADA, AUSTRALIA, NETHERLANDS
Product description
Zimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 46 mm O.D., Size G, Bearing, Model/Catalog Number: 110031013; semi constrained hip prosthesis
Code information (models, lots, serial numbers)
UDI (01)00889024572713(17)300406(10)67160480; Lot Number 67160480
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- Vivacit-E® 110031013Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 recall15 reports
Product lines
Other devices under the same FDA submission
16 devices cleared under K190656 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.
- Vivacit-E® 110031012Zimmer, Inc. · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented1 recall33 reports
- Vivacit-E® 110031009Zimmer, Inc. · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented25 reports
- Vivacit-E® 110031010Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented11 reports
- Vivacit-E® 110031011Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented6 reports
- Longevity® 110031000Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented4 reports
- Vivacit-E® 110031015Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented3 reports
- Vivacit-E® 110031314Zimmer, Inc. · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented3 reports
- Vivacit-E® 110031313Zimmer, Inc. · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented2 reports
- Longevity® 110030997Zimmer, Inc. · Prosthesis, hip, hemi-, femoral, metal/polymer, cemented or uncemented1 report
- Longevity® 110030998Zimmer, Inc. · Prosthesis, hip, semi-constrained, metal/ceramic/polymer, cemented or non-porous, uncemented1 report
- Longevity® 110030999Zimmer, Inc. · Hip Prosthesis, semi-constrained, cemented, metal/polymer, + additive, porous, uncemented1 report
- Longevity® 110031001Zimmer, Inc. · PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, POROUS UNCEMENTED1 report
Other products in the same recall event
The FDA groups related products into one recall event (event 97765).
- Z-0418-2026 · Open, ClassifiedZimmer Biomet Dual Mobility Vivacit-E Highly Crosslinked Polyethylene, 28 mm I.D., 44 mm O.D., Size F, Bearing, Model/Catalog Number: 110031012; semi constrained hip prosthesis
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page