Skip to content

Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND… recallZ-0573-2026 · Sterilmed, Inc. · initiated Oct 8, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Sterilmed, Inc. or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Affected catheters were reprocessed beyond validated cycles. This may compromise device sterility or mechanical integrity of the device and can lead to severe adverse events such as infection, vascular injury or embolism.

FDA root cause category: Reprocessing Controls

What the company is doing

Initial phone call notification was performed 10/10/2025 to affected customers. Customers were instructed to immediately stop use of identified devices. Affected products should be segregated. Follow-up written notification was sent to customers on 10/17/25 which included instructions on how to return affected products.

Recall details

Recall number
Z-0573-2026
Status
Open, Classified
Initiated by the company
Oct 8, 2025
Posted by the FDA
Nov 20, 2025
Recalling firm
Sterilmed, Inc.
Quantity
6
Product code
OWQ
FDA submissions
K043453
Distribution
United States Only: Kentucky, Maryland, Massachusetts, Mississippi, New Jersey, Pennsylvania, Texas, Utah

Product description

Brand Name: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Product Name: CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Model/Catalog Number: ACU10135910 Software Version: N/A Product Description: ACUSON, CATHETER ACUNAV DIAGNOSTIC ULTRASOUND 8FR NOT FOR USE WITH SIEMENS SYSTEM Component: N/A

Code information (models, lots, serial numbers)

Lot Code: GTIN/UDI: 10888551004725 LOT# SERIAL #: 2196112 999253 2200877 937157 2219499 1182771 2219926 1048031 2220881 1126584 2234882 2304200759

Devices named in this recall

1 device in the FDA device registry whose identifier appears in the recall's code information.

Other devices under the same FDA submission

3 devices cleared under K043453 are not named by identifier in this recall. One submission often covers many models, so they may or may not be affected.

Other products in the same recall event

The FDA groups related products into one recall event (event 97803).

Sources and updates

How recalls are matched to devices · Editorial policy · Report an error on this page