MC3 VitalFlow Console, REF 58100 recallZ-0509-2026 · Medtronic Perfusion Systems · initiated Oct 3, 2025
Class II Open, Classified All recalls posted in November 2025
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from Medtronic Perfusion Systems or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
As of August 6, 2025, Medtronic has received eleven reports of VitalFlow Consoles displaying an E70 error code during normal operation. When this occurs, the touch screen may become temporarily unresponsive and go blank for up to two minutes before recovering to full functionality. Importantly, the console continues to maintain set pump speed and function throughout.
FDA root cause category: Software in the Use Environment
What the company is doing
Medtronic Cardiac Surgery issued an Urgent: Medical Device Correction notice to its consignees on 10/03/2025. The notice explained the issue, potential harms, and requested the consignee contact Medtronic Sales Representative to schedule the service event to their affected VitalFlow" Console. Please reference Technical Service Update (TSU) "MOD0209" when scheduling. The notice also indicated that the consignees use unaffected VitalFlow Console devices, if available, until the correction has been completed. For questions regarding the communication, please contact your Medtronic Sales Representative.
Recall details
- Recall number
- Z-0509-2026
- Recall class
- Class II
- Status
- Open, Classified
- Initiated by the company
- Oct 3, 2025
- Posted by the FDA
- Nov 14, 2025
- Recalling firm
- Medtronic Perfusion Systems
- Quantity
- 167 units
- Product code
- QNR
- FDA submissions
- K250199, K230364
- Distribution
- US Nationwide distribution.
Product description
MC3 VitalFlow Console, REF 58100; Blood pump of ecmo
Code information (models, lots, serial numbers)
Show the full code information (1k characters)
UDI/DI 10850007732247, Serial Numbers: 000023, 000047, 000073, 000100, 000126, 000150, 000173, 000197, 000223, 000024, 000048, 000075, 000101, 000127, 000151, 000174, 000202, 000224, 000025, 000051, 000076, 000102, 000128, 000152, 000177, 000203, 000225, 000027, 000052, 000078, 000103, 000129, 000153, 000179, 000204, 000228, 000028, 000054, 000079, 000104, 000130, 000154, 000180, 000205, 000229, 000029, 000055, 000080, 000106, 000131, 000155, 000181, 000206, 000277, 000030, 000056, 000081, 000108, 000132, 000156, 000182, 000207, 000284, 000031, 000058, 000082, 000109, 000133, 000157, 000183, 000208, 000032, 000059, 000083, 000110, 000134, 000159, 000184, 000209, 000033, 000060, 000084, 000112, 000135, 000161, 000185, 000210, 000034, 000061, 000085, 000115, 000136, 000162, 000186, 000211, 000035, 000063, 000087, 000116, 000140, 000163, 000187, 000212, 000037, 000065, 000088, 000117, 000141, 000165, 000188, 000213, 000038, 000066, 000089, 000118, 000142, 000166, 000189, 000214, 000039, 000067, 000090, 000119, 000143, 000167, 000190, 000215, 000040, 000068, 000092, 000120, 000145, 000168, 000191, 000216, 000043, 000069, 000095, 000121, 000146, 000169, 000192, 000217, 000044, 000070, 000096, 000122, 000147, 000170, 000193, 000218, 000045, 000071, 000097, 000124, 000148, 000171, 000194, 000219, 000046, 000072, 000099, 000125, 000149, 000172, 000195, 000221.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- VitalFlow Console 58100Michigan Critical Care Consultants, Inc. · Blood Pump For Ecmo, Long-Term (> 6 Hours) Use1 recall11 reports
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 18, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page