Ultrasound System 1300, bkSpecto w/battery recallZ-0436-2026 · B-K Medical A/S · initiated Oct 9, 2025
Open, Classified
Is my device affected?
- Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
- If yours is listed, follow the instructions from B-K Medical A/S or your healthcare provider. The company's action is summarized under what the company is doing.
- Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.
Where to find your model and lot numbers · What recall classes and statuses mean
Why it was recalled
GE HealthCare has become aware that the Instructions for Use within the user guide and service manual do not adequately state the required battery replacement intervals, end-of-life handling procedures, or general battery safety precautions.
FDA root cause category: Device Design
What the company is doing
GE Healthcare issued an "URGENT MEDICAL DEVICE CORRECTION NOTICE" to its consignees on 10/09/2025 via traceable means. The notice explained the issue, potential risk, and requested: "Actions to be taken by Customer / User You can continue to use your device by following the instructions below: 1. Review and follow the applicable addendums to the Instructions for Use and the updated Service Manuals. The Addendums and Service Manuals are available for download at: https://www.gehealthcare.com/support/manuals. 2. Follow the conditions defined within the addendum and Service Manual applicable to your product for further instructions. 3. Place the applicable addendum and Service Manual with your existing product Instructions for Use. 4. If you have difficulty accessing the documentation or require a printed copy of the applicable addendum or Service Manual, please contact a GE HealthCare Service Representative for assistance. Please ensure all potential users in your facility are made aware of this safety notification and the required actions. Please retain this document for your records." For questions or concerns regarding this notification, please contact GE HealthCare at 1-800-645-4442 or [email protected] or your local Service Representative.
Recall details
- Recall number
- Z-0436-2026
- Status
- Open, Classified
- Initiated by the company
- Oct 9, 2025
- Posted by the FDA
- Nov 6, 2025
- Recalling firm
- B-K Medical A/S
- Quantity
- 1697 units
- Product code
- IYN
- FDA submissions
- K231764
- Distribution
- Worldwide distribution: US (nationwide) and OUS (foreign) to countries of: Algeria, Argentina, Australia, Austria, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Denmark, Dominican Rep, Ecuador, Estonia, Finland, France, Georgia, Germany, Greece, Hong Kong, Hungary, Iceland, India, Indonesia, Ireland, Israel, Italy, Japan, Korea (Republic Of), Kuwait, Latvia, Lebanon, Luxembourg, Malaysia, Mexico, Mongolia, Morocco, Nepal, Netherlands, New Zealand, Norway, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sweden, Switzerland, Taiwan, Thailand, Turkiye, Ukraine, United Arab Emirates, and United Kingdom.
Product description
Ultrasound System 1300, bkSpecto w/battery; Model No. 1300-21 or 1300-S1; System, Imaging, Pulsed Doppler, Ultrasonic
Code information (models, lots, serial numbers)
Show the full code information (15k characters)
1) 1300-21: GTIN 05704916001285 and pending, Serial Numbers: 2002690, 2008466, 2008495, 2006774, 5012683, 5013881, 5012684, 2008725, 2008280, 2004419, 5012980, 2007101, 2009029, 5014333, 2003852, 2002851, 2002142, 2005860, 5013937, 2007188, 2002950, 2003495, 2008470, 5013934, 5013935, 5016211, 2005193, 5014141, 5014514, 5015892, 5012104, 2009012, 2009153, 5012976, 2008185, 2008137, 5012307, 5014139, 2000774, 2006237, 2008144, 5012319, 5014906, 5012789, 5015458, 2006602, 2004379, 2007123, 2008242, 5011353, 2007736, 5014803, 2006235, 5011354, 2002049, 2005281, 5015083, 5015087, 2008527, 5016292, 5016477, 5012055, 5012326, 5015201, 2003658, 2008735, 5011807, 2005450, 2007082, 2009247, 2009142, 2003588, 5016079, 5016301, 2009108, 2008244, 2003721, 2008663, 2008277, 2006691, 5016299, 2008383, 2006881, 5016300, 2006131, 2001918, 2002172, 2005832, 5015218, 2008792, 5011799, 2003137, 5016193, 5016194, 5012310, 5016082, 2008597, 5011479, 2006483, 2001865, 5012954, 5011802, 2002891, 2008356, 2006359, 2004290, 2002824, 2002177, 5011817, 2003597, 2004373, 5013278, 5013126, 5011800, 5011482, 2004264, 2004295, 5015904, 5014266, 2004372, 2005640, 2001433, 2005658, 2006084, 5012545, 2008386, 2002443, 2008578, 2003686, 2002325, 2001425, 2003150, 2003470, 2008080, 2006885, 2004466, 2008586, 2008472, 2000782, 2000821, 2000862, 2000902, 2000906, 2000914, 2000917, 2003502, 2003594, 2003682, 2003854, 2004519, 2008461, 5010293, 5010925, 5010926, 5010927, 5010929, 5010930, 5010933, 5010934, 5010935, 5010936, 5010938, 5010939, 5010940, 5011142, 5011145, 5011152, 5011154, 5011156, 5011157, 5011160, 5011346, 5011348, 5011351, 5011352, 5011355, 5011461, 5011471, 5012558, 5012566, 5012568, 5012618, 5012682, 5012688, 5012788, 5012964, 5012968, 5012979, 5013125, 5013407, 5013479, 5013656, 5013806, 5013888, 5013889, 5013933, 5014272, 5014432, 5014530, 5015215, 5015217, 5015717, 5016212, 5016294, 2004972, 2006142, 2002845, 2004149, 2007679, 5014334, 2006557, 2006563, 2002052, 2001021, 5013880, 2002289, 5016200, 2008100, 2008685, 2008111, 5012410, 5012547, 2003851, 5012103, 2003684, 5014267, 5012691, 5012790, 2006995, 2008150, 2009155, 2007803, 5013936, 5015456, 5012784, 2008690, 2003710, 2008055, 5013134, 5016204, 2003012, 2002596, 5014819, 2006797, 5013822, 5014904, 2002170, 5015919, 5011465, 5016305, 2002522, 2001145, 2005261, 2008059, 2006635, 5012792, 2002915, 5016096, 2006500, 2006239, 2008501, 2004431, 2003685, 2008902, 2003878, 2005336, 2007801, 2007804, 2006957, 2008287, 2001857, 5011818, 5016085, 2004210, 2001650, 2003676, 2005826, 2005834, 5016467, 2001655, 2008499, 5014283, 2002844, 2001888, 5014204, 5014271, 5012327, 2006271, 5011347, 2000514, 2001148, 2005906, 5015893, 5016007, 2005851, 2008257, 2007728, 2002710, 2001826, 2008249, 2008800, 2003860, 2002016, 2002907, 2002917, 2003056, 2007562, 2003447, 2004296, 2005612, 2002882, 2003144, 2007824, 2008098, 2002708, 2002531, 5014901, 5015998, 5016319, 5016081, 2004395, 5014823, 2001417, 2007781, 2005306, 5011148, 5012052, 2003183, 2006502, 2002904, 5016461, 2002664, 2006274, 2006678, 5011821, 2003499, 2003560, 2003583, 2003617, 2008485, 2003821, 2001458, 2001481, 2001485, 2001487, 2008741, 2006113, 2002589, 5016317, 2008140, 2001645, 2006309, 2006369, 2006458, 2006475, 2006561, 2006566, 2006595, 2006620, 2006778, 2006934, 2006997, 2007167, 2007526, 2007558, 2006987, 2007036, 2008039, 2008161, 2008235, 2009248, 2005330, 2008713, 2003486, 2004191, 2002184, 5015355, 2005731, 5013279, 2005500, 2005825, 2006261, 2005414, 2003492, 2008050, 5014818, 5016302, 2007268, 5015712, 5015711, 5015917, 2008694, 2000822, 2008536, 5011820, 5012317, 2002789, 2008679, 5015411, 5015441, 5015918, 2001695, 2006249, 2001171, 2001936, 2002714, 2007281, 2004128, 5016458, 5012783, 2009187, 5012312, 2001861, 2007305, 2009077, 2008502, 2008743, 2003677, 2003667, 2003747, 5012957, 2005322, 2004982, 2005009, 2007270, 2002420, 2008045, 5012970, 5016090, 2002825, 2006147, 5014510, 5015910, 2008379, 2007808, 5013655, 2008143, 2008201, 5012565, 2002135, 2002150, 2002151, 2002152, 2001426, 2008375, 5013887, 2003630, 2008145, 5011467, 2008279, 2003797, 2006525, 2002575, 2004249, 2003624, 2007032, 2006318, 5013459, 5013886, 2003626, 2003399, 2005277, 2005248, 2008389, 2007627, 5015205, 5015208, 2008708, 2008977, 2008160, 5012960, 5014277, 2008726, 2001843, 2006313, 2007313, 5013463, 2008681, 5016201, 5014919, 5014335, 5016080, 2003355, 2003510, 5013890, 5013132, 5012955, 2001758, 2008246, 2008265, 2007061, 5012051, 2007614, 5012785, 2005638, 2008717, 2008518, 2006672, 2001031, 5014921, 2002458, 5015024, 5011485, 2008899, 2006496, 2006521, 2008061, 5013871, 2006107, 2002813, 2004293, 5015216, 2008387, 2008154, 5012546, 5012548, 5014332, 2008200, 2006230, 2003006, 5016086, 2002143, 2006247, 2002341, 2002889, 2009107, 5012539, 2005517, 2007004, 5015715, 2004375, 2002862, 5011458, 2008253, 2008459, 2008487, 2008493, 2008512, 2008516, 2008524, 5015913, 2005531, 2005537, 2005332, 2003754, 2004647, 2005417, 2004374, 2003612, 2003794, 2007178, 2004203, 2006495, 2007018, 2006998, 2006787, 2007173, 5015716, 2004243, 2002846, 2006862, 2002394, 2002414, 2002885, 2003410, 2003449, 2006215, 2003446, 2003467, 2006860, 5015025, 5013283, 2002332, 2003445, 2003057, 2005528, 2005534, 2001766, 2006132, 2004386, 2003448, 2003179, 2002140, 2004435, 2006661, 2001651, 2006278, 2000712, 5012791, 2003581, 5014801, 5014802, 2001137, 2002051, 2000780, 2002047, 5014136, 5014269, 5014822, 2001701, 2001713, 2001783, 2002338, 5014216, 2008083, 2008125, 2008156, 2008511, 2008734, 2008779, 2009157, 5011483, 2007748, 5011484, 2006988, 2003468, 2000403, 2001084, 2001086, 5014138, 5015446, 2000771, 2006411, 2000281, 2000479, 2000785, 2000248, 5014527, 2000322, 2004287, 2004260, 2002783, 5015444, 5013130, 5012402, 2000164, 5014215, 2005751, 2000577, 2003614, 2007273, 5016077, 5016078, 2002048, 2003586, 2005364, 2001297, 2005755, 2007680, 2004455, 2000467, 5014814, 5014816, 5011161, 2005396, 5014912, 5015699, 2000828, 2001688, 5011812, 5011824, 5015657, 5014140, 5014808, 5014807, 2000504, 2003816, 2001073, 2001478, 2007678, 2007785, 2007792, 2008521, 2008738, 2008765, 5011466, 5013829, 5014331, 5016014, 2006391, 2000109, 2002337, 5011361, 5011363, 2002444, 2005653, 2006242, 5016089, 5014222, 2004412, 5014214, 5014526, 5014536, 2008490, 5014824, 2007524, 5016293, 2004135, 2002782, 2002017, 2005786, 2002709, 2002400, 2001641, 2005001, 2005765, 2007067, 5014221, 2000993, 5010923, 5012315, 5011809, 2002878, 5013135, 2009048, 2009140, 2003496, 2003546, 2005490, 5011462, 5011463, 5012786, 5014820, 2005723, 2006530, 2002881, 2000202, 2002178, 2005462, 2008748, 5010937, 5011149, 5012556, 5012557, 2000245, 2000507, 5014761, 5011151, 2006667, 2002456, 2002459, 2002499, 2002523, 2002524, 2002526, 2002527, 2002528, 2002533, 2002544, 2000908, 2004951, 2001200, 2001264, 2006385, 2000713, 2002280, 2002281, 2002284, 2002811, 2006167, 5012974, 5014223, 2000275, 2001026, 2001685, 2001694, 2002344, 2002801, 2003060, 2003551, 2003659, 2003683, 2003757, 2003763, 2003843, 2004131, 2005449, 2006731, 2008232, 2008546, 2008570, 5011365, 5012563, 5014133, 5015006, 5015914, 5015915, 2005456, 5015358, 5015359, 5012917, 5012314, 5011819, 5015453, 5015455, 5013291, 2001535, 2005952, 5013952, 2000207, 2001796, 5013296, 5014421, 2006877, 5013474, 2000556, 2000558, 2005551, 2000569, 2000907, 2003853, 2005738, 5014815, 2001935, 2003009, 5011806, 2003650, 2005965, 5014798, 5015079, 2000172, 2004984, 2002787, 2006352, 2000168, 2002839, 2000179, 2002210, 2007125, 2001785, 2003838, 2000196, 2006611, 2005259, 2003469, 2006749, 2009137, 2009123, 2003471, 2003475, 5014446, 5015459, 2001938, 5013120, 2002549, 2002598, 2007117, 2005603, 2001090, 5013114, 5015647, 2005455, 5013953, 5015649, 5015650, 2003013, 2006705, 2006537, 2003443, 2007010, 2000578, 5014923, 2001893, 5011147, 2006280, 2006449, 2002841, 2002842, 5014817, 2003348, 2003708, 5013111, 5015448, 2006588, 5011804, 2006714, 2002335, 2005694, 5011810, 2000545, 2001088, 2001172, 2005683, 2005684, 2000321, 5013124, 2000474, 2007286, 2005326, 5011805, 2002806, 2007726, 5015351, 2003479, 2000108, 2000784, 2007729, 2003837, 2001677, 2001764, 5014909, 5014911, 5014130, 2000732, 5012983, 2003709, 2006932, 2006386, 2005812, 5011803, 5014203, 5013404, 2002502, 5014805, 2000861, 5011459, 5011478, 2000638, 5016017, 5016018, 2000816, 2005474, 2005478, 2006606, 2006609, 2007012, 2007523, 2007554, 5015447, 2007513, 5015865, 5016076, 5011153, 2000775, 2001079, 5013112, 5010931, 5014821, 2005240, 5015648, 2002780, 2000478, 2003755, 2007011, 2007052, 2007063, 2007083, 2007601, 5014917, 5016008, 5016009, 2006694, 2006832, 2003547, 5011487, 2002884, 2002399, 2001385, 2003180, 2006599, 2006621, 5012562, 5012685, 2002578, 2002546, 2004639, 2000230, 2006328, 2000395, 2004610, 2006654, 5015646, 2002776, 2005318, 2004282, 5012302, 2002020, 2003351, 2004163, 5013299, 5013473, 5014135, 5014420, 2000269, 2003756, 2003820, 2004280, 2004298, 2004376, 2004385, 2004403, 2004537, 2004545, 2004547, 2004592, 2004601, 2004617, 2004777, 2004794, 2004948, 2005010, 2007527, 2000773, 2006789, 2002214, 2005588, 5013467, 2000703, 2003481, 5013475, 2001461, 2000263, 2001895, 2005360, 2005368, 2006338, 5014903, 5013469, 2002659, 2000415, 5014137, 2000411, 5013636, 2001432, 5013637, 2003472, 2005476, 2006337, 2000352, 2004439, 2002583, 2002847, 2006719, 2005705, 5011159, 5013274, 2001649, 2001930, 5014908, 2000739, 2007119, 2000278, 2007553, 2007737, 2000829, 2003742, 2006849, 2003672, 2001467, 2001793, 2007093, 5011475, 5014343, 5011460, 2002614, 5014210, 2007669, 2007790, 2008057, 2008082, 2008148, 2008205, 2008240, 2008332, 2008336, 2008337, 2008347, 2008352, 2008358, 2008376, 2008477, 2008548, 2008665, 2008712, 2008758, 2008761, 2009035, 2009062, 2009066, 2009082, 2009120, 5010924, 2000469, 2000356, 2003408, 2006613, 5012795, 5012916, 2006233, 2002288, 2000396, 5016189, 2005610, 2007130, 2007331, 2007332, 2007551, 5011801, 2000353, 5015070, 2000813, 5015639, 2000778, 5011144, 2006098, 2002144, 2001816, 2002019, 2003504, 5015638, 2006251, 2001135, 2000553, 2002859, 2001501, 2004638, 2003836, 2001896, 2007103, 2006717, 2007089, 2004202, 2005778, 2002520, 2003849, 2002138, 2001163, 2001648, 5013821, 2000818, 2002850, 2004954, 2001001, 2004953, 2006818, 2006880, 2007145, 2008060, 5011464, 2002922, 2005267, 2005572, 2007516, 2008673, 2002886, 2002948, 2002952, 2006582, 2007006, 2007565, 5015010, 5015011, 5016306, 2002342, 2002909, 2005457, 2005805, 2006878, 2008131, 2008191, 2005743, 2008773, 5011364, 2002495, 2003509, 2002573, 5011474, 2000547, 5011477, 2005571, 2003065, 2008262, 2005472, 2000576, 2001465, 2002874, 2000995, 5011480, 5011481, 5011808, 5012971, 2008492, 2006067, 5015198, 2007007, 5013128, 2000517, 2002828, 2005836, 5014764, 5014806, 2002840, 2007031, 5013131, 5015193, 5015194, 5015195, 5015196, 2004281, 2009185, 2009172, 2009196, 2003629, 5015634, 5013954, 2000825, 2000276, 5014910, 5015353, 5015875, 5011350, 5014433, 5014442, 5014673, 5014678, 2006119, 2003545, 2003669, 5014330, 5014760, 5014762, 5014763, 2005523, 2003675, 2004014, 2003824, 5012921, 2005590, 2005841, 5013119, 2001004, 2000830, 2001369, 2001372, 2001375, 2001376, 2001378, 2003396, 2004443, 2005507, 2006639, 2000268, 2000270, 2000272, 2002333, 2003409, 2003613, 2003705, 2005320, 2005334, 2006117, 2006306, 2006660, 2006770, 2007295, 2007806, 2008728, 5011811, 5013282, 5013882, 5014795, 2006882, 2007005, 2008784, 5015698, 2002416, 2007013, 2008058, 5011146, 2000198, 2001205, 2001260, 2003736, 2004158, 2005335, 2005569, 2005581, 2005595, 2005599, 2005602, 2005605, 2005988, 2006535, 2006568, 2007127, 2007338, 2007520, 2007603, 2007670, 2007681, 2007690, 2007691, 5014211, 2008245, 2000472, 2000483, 2001518, 2005333, 2005941, 2006248, 2007528, 2007557, 2008341, 2008355, 2008483, 5012050, 2007166, 5015903, 2002314, 2003117, 2003557, 2000200, 5011143, 5012108, 2002391, 5011357, 2002918, 2008118, 2008289, 2008790, 5011150, 2009000, 2008825, 5013805, 2002211, 2000854, 2002182, 2004550, 5011472, 5013292, 5013295, 5013300, 5013406, 5013413, 5013464, 5014426, 2001019, 5013281, 5013287, 5015713, 5015714, 2000110, 2000232, 5011359, 5014448, 2000919, 2001656, 2001708, 2002294, 2001810, 5014796, 2002802, 5014905, 2000920, 2006647, 2006655, 2007185, 5012693, 2004978, 2001824, 2001923, 2000389, 2003788, 5012973, 5011356, 5013277, 2002698, 5014134, 2001206, 2001204, 5015021, 5016213, 5016214, 5011827, 5013899, 5015197, 2003850, 2007170, 5011358, 5011360, 5011362, 2003478, 2003759, 2001209, 2006149, 5012961, 2001164, 2004508, 2004521, 5012159, 5012303, 5012553, 5014419, 5016010, 2002791, 2002704, 5013129, 5015853, 2003740, 2007102, 2007143, 2007560, 2007567, 2007671, 2007706, 2007718, 2008737, 2008752, 2008762, 2000737, 2001201, 2001213, 2003487, 2004453, 2005894, 2006254, 5015891, 5013654, 5014804, 5014913, 2001027, 5012411, 5014427, 2001852, 2005771, 2006073, 2008076, 5015331, 2004292, 2003205, 2003519, 2000387, 2002043, 2004446, 5011349, 5014922, 2003236, 2006839, 2003653, 5012787, 5013950, 2003610, 5011158, 2005981, 2003587, 2003402, 5012978, 2003826, 2005676, 2006166, 2005811, 2006172, 2006825, 2009159, 2006814, 2004418, 2004255, 2003628, 2003632, 2005504, 2005511, 2006192, 2000918, 2006811, 2003181, 5013951, 2006424, 2006615, 2008064, 2008065, 2008075, 2008081, 2008110, 2008196, 2008203, 2008711, 2008736, 2008781, 2008901, 2008908; 2) 1300-S1: UDI/DI pending, Serial Numbers: 2007274, 2001754, 2003771, 2003611, 2005842, 2007514, 2004986, 2003773, 2006652, 2006072, 2005329, 2006408, 2005728, 2003239, 2002206, 2006085, 2005444, 2002946, 2005987, 2006453, 5014797, 2006768, 2006876, 2007053, 5016019, 2003453, 2004206, 2004197, 2001387, 2006265, 2003746, 5015897, 2002715, 5016202, 2003656, 2002551, 2007115, 5015916, 2007564, 2001024, 2004362, 2004577, 2002911, 2001922, 2005628, 2005856, 2002860, 2003750, 2006212, 2002945, 2005670, 2002340, 2007676, 2004148, 2003116, 2001822, 5015641, 5016006, 2007141, 2005339, 2003398, 2006066, 2006580, 2003330, 2006228, 2003198, 2001025, 2003579, 2001690, 5015332, 2004299, 2004144, 2007049, 2005190, 2001788, 2002809, 2000921, 2000997, 2001010, 5015643, 5016471, 5016475, 2005532, 2004157, 2006367, 2005840, 2003829, 2007556, 2003412, 2006875, 2004168, 2001300, 2004985, 2005608, 2005837, 2006933, 2002045, 2006589, 2003718, 2007179, 2007687, 2005803, 2005421, 2002803, 2005729, 2002893, 2004271, 2003815, 5016020, 2006666, 2007317, 2004156, 5015340, 2002287, 5015345, 2003118, 2002018, 2002452, 5015719, 5016455, 2002455, 2003689, 2005584, 2002812, 2001937, 2001931, 2003549, 2004142, 2006451, 5015718, 2003680, 2007682, 2001867, 2007068, 2006473, 2003688, 2001170, 2001304, 2005469, 2002613, 2005394, 2006523, 2002879, 5014902, 5016190, 2006204, 2003244, 2001679, 2002148, 5016000, 2004252, 2004460, 5016084, 2001866, 2003762, 2003822, 5015460, 2001752, 5015854, 5015999, 2001827, 2001855, 2005747, 2003678, 2001149, 2001151, 5015334, 5015335, 5015336, 5015706, 5015707, 5016304, 5016210, 2007090, 5013127, 2001784, 2001707, 2001851, 2003748, 2006255, 2005630, 2006201, 2006226, 2005219, 2005485, 2003485, 2006202, 2002147, 2001524, 2002149, 2005617, 2005667, 2005718, 2003484, 2006809, 2002453, 2004253, 5016191, 5016203, 2004279, 2001308, 5015997, 2006643, 2004383, 2004457, 2004536, 2004608, 2004634, 2004635, 2004637, 2004785, 2004796, 2004797, 2004931, 2004946, 2004952, 5015874.
Devices named in this recall
1 device in the FDA device registry whose identifier appears in the recall's code information.
- bkSpecto Ultrasound System 1300Bk Medical Aps · Transducer, Ultrasonic, Diagnostic1 recall0 reports
Other products in the same recall event
The FDA groups related products into one recall event (event 97830).
- Z-0435-2026 · Open, Classified1202 Flex Focus Ultrasound System, cart models UA1214 and UA1814, used with: Flex Focus 200, Flex Focus 300, Flex Focus 400, Flex Focus 500, Flex Focus 700, Flex Focus 800; Syste…
- Z-0437-2026 · Open, ClassifiedUUltrasound System 2300, bkActiv w/battery; Model No. 2300-11; System, Imaging, Pulsed Doppler, Ultrasonic
- Z-0438-2026 · Open, Classifiedbk3000 Ultrasound System w/battery; Model No. 2300-66; System, Imaging, Pulsed Doppler, Ultrasonic
- Z-0439-2026 · Open, Classifiedbk3500 Ultrasound System w/battery; Model No. 2300-41; System, Imaging, Pulsed Doppler, Ultrasonic
- Z-0440-2026 · Open, Classifiedbk5000 Ultrasound System w/battery; Model No. 2300-61; System, Imaging, Pulsed Doppler, Ultrasonic
Sources and updates
- Recall record: FDA Medical Device Recalls, file dated Oct 16, 2026 · View the FDA recall notice ↗
- Devices: FDA GUDID, file dated Oct 13, 2026. Devices are linked when their identifier (with a valid check digit) appears in the recall's code information.
- Page last rebuilt: Oct 16, 2026. Text above is shown as published by the FDA.
How recalls are matched to devices · Editorial policy · Report an error on this page