Skip to content

NICO Myriad-LX recallZ-0600-2026 · Stryker Corporation · initiated Oct 31, 2025

Open, Classified

Is my device affected?

  1. Compare the model, catalog and lot or serial numbers on your device or its packaging with the code information below. A recall usually covers only specific models or lots.
  2. If yours is listed, follow the instructions from Stryker Corporation or your healthcare provider. The company's action is summarized under what the company is doing.
  3. Do not stop using a prescribed or implanted device on your own — talk to your healthcare provider first.

Where to find your model and lot numbers · What recall classes and statuses mean

Why it was recalled

Tape used to secure components within packaging contains latex, despite products being labeled as latex-free. Latex could potentially cause an allergic reaction.

FDA root cause category: Under Investigation by firm

What the company is doing

On October 31, 2025, the firm began notifying impacted customers through Urgent Medical Device Recall letters. Customers were instructed to mark affected product with "CONTAINS LATEX" and place a copy of the notice at the product's storage location before use. Product returns are not required. However, customers who would like to receive a replacement were given options to destroy or return product for a replacement.

Recall details

Recall number
Z-0600-2026
Status
Open, Classified
Initiated by the company
Oct 31, 2025
Posted by the FDA
Nov 26, 2025
Recalling firm
Stryker Corporation
Quantity
39,148
Product code
GEI
FDA submissions
K191599, K232567
Distribution
Worldwide distribution - US Nationwide and the countries of Australia, Belgium, Canada, England, Ireland, Italy, Kuwait, Singapore, Slovenia, Spain, and the Netherlands.

Product description

NICO Myriad Illumination Pack Product No. NN-1000 11g Illumination Sleeve/Fiber; Product No. NN-1001 13g Illumination Sleeve/Fiber

Code information (models, lots, serial numbers)

Product Number/UDI-DI (GTIN)/Lot Numbers: NN-1000 00816744021009 37AC5163 37AE9811 37AJ1438 37BJ0513 37CB3410 37CB3416 37CD6003 37CE3030 37EB5720 37FA4260 37GD2216 37GJ1603 37IC7023 37IJ0208 37JJ0122 37KI1300 37LA0158 37LI0320 37LJ1522; NN-1001 00816744021016 37AC5177 37AE9942 37AE9955 37AJ1439 37BJ0516 37CB3380 37CB3419 37CD6005 37CJ0204 37DE4864 37EB5721 37FA4261 37GD2219 37GJ1607 37IC7027 37IJ0209 37JD5753 37JJ0123 37KI1301 37LA0159 37LI0322 37LJ1525

Devices named in this recall

2 devices in the FDA device registry whose identifier appears in the recall's code information.

Product lines

Other products in the same recall event

The FDA groups related products into one recall event (event 97833).